Deck 12: Ct Physics and Instrumentation

Full screen (f)
exit full mode
Question
What is a reconstruction kernel?

A) A unit of data in a reconstructed CT image
B) The entire data set used to produce reconstructed images in CT
C) A measurement used to determine the data set used for CT reconstruction
D) An algorithm that defines the clinical application and amount of image smoothing applied in the reconstruction
Use Space or
up arrow
down arrow
to flip the card.
Question
Important technical developments that define current helical CT systems include all of the following EXCEPT

A) continuous reduction in radiation of the x-ray tube.
B) continuous-motion gantry using slip-ring technology.
C) continuous motion of the imaging table.
D) continuous data transfer from the detector array.
Question
The purpose of using a filter in CT scanning is

A) to absorb the very-low-energy x-rays.
B) to act as an upper level discriminator.
C) to act as a window.
D) To attenuate the high-energy x-rays.
Question
In helical CT, pitch is defined as

A) the ratio of the patient's movement through the gantry during one 360-degree beam rotation relative to the beam collimation.
B) the total time for a patient to move through the gantry during a CT scan.
C) the ratio of the beam collimation relative to the mAs for a CT scan.
D) the coordination of a patient's possible movement,
E)g., breathing, during the CT scan relative to slice thickness.
Question
What is the purpose of administering a contrast agent for a CT procedure?

A) Reduce patient dose for longer scans that require thinner slicers
B) Reduce streaking artifacts when scanning high-density objects
C) Reduce attenuating properties of tissues with wide variances in density
D) Enhance detection of lesions and aid in defining normal structures
Question
In x-ray tubes, the types of reactions that produce x-rays include

A) Bremsstrahlung and scatter.
B) Bremsstrahlung and characteristic.
C) characteristic and scatter.
D) characteristic and refraction.
Question
Using a higher mA

A) results in an increase in image noise.
B) results in an increase in contrast resolution.
C) results in a decrease in dose to the patient.
D) shortens the imaging time for the patient.
Question
Which of the following restricts the width and shapes the beam to the area of interest for a single rotation of the tube and detectors around the patient?

A) Prepatient collimators
B) Postpatient collimators
C) Pitch
D) Rotation speed
Question
All of the following are true about the kVp setting EXCEPT

A) it selects the maximum possible energy of the x-ray beam.
B) it defines the fraction of photons that will reach the detectors.
C) it affects the radiation exposure to the patient.
D) it increases beam hardening.
Question
In an x-ray tube, when the milliamperes (mA) are increased:

A) Bremsstrahlung and characteristic x-rays increase.
B) Bremsstrahlung and characteristic x-rays decrease.
C) Bremsstrahlung x-rays increases and characteristic x-rays decrease.
D) Bremsstrahlung x-rays decrease and characteristic x-rays increase.
Question
In an x-ray tube, of the energy applied to the system, the percentage of x-rays that exit is

A) 1%.
B) 10%.
C) 50%.
D) 99%.
Question
Which of the following may be used to reduce streaking artifacts when scanning anatomy with high-density objects?

A) Bow-tie filter
B) Adaptive detector array
C) Thinner collimation
D) Decreased kVp
Question
Computed tomography (CT) scanning accomplishes all of the following EXCEPT

A) minimal superimposition of tissues.
B) improved contrast resolution.
C) reduction of radiation dose compared to conventional radiography.
D) differentiation of tissue densities.
Question
In an x-ray tube, a low kVp results in

A) high-energy Bremsstrahlung x-rays.
B) low-energy Bremsstrahlung x-rays.
C) fewer Bremsstrahlung x-rays.
D) an increase in tungsten characteristic x-rays.
Question
Energies required for typical PET/CT studies of the body will be in the range of

A) 20 to 30 kVp.
B) 40 to 75 kVp.
C) 80 to 140 kVp.
D) 145 to 160 kVp.
Question
A general rule is that a 15% increase in kVp results in

A) a 15% increase in mAs.
B) doubling the mAs.
C) a 15% reduction in mAs.
D) no affect on mAs.
Question
Collimators in CT scanning are used to do all of the following EXCEPT

A) limit the beam size.
B) reduce scattered x-rays.
C) define slice thickness.
D) offset the alignment of the tube and detector array to optimize the beam location.
Question
In an x-ray tube, a high kVp results in

A) low-energy Bremsstrahlung x-rays.
B) a decrease in tungsten characteristic x-rays.
C) an increase in tungsten characteristic x-rays.
D) fewer Bremsstrahlung x-rays.
Question
In an x-ray tube, the anode, for the most part, is most commonly constructed from

A) tungsten.
B) lead.
C) rhenium.
D) silver.
Question
In a CT scanner, the oil bath that may surround the rotating anode tubes serves to

A) make the tubes rotate more quickly.
B) make the tubes rotate more smoothly.
C) cool the tubes.
D) provide efficient conduction of the current.
Question
The slice thickness that may be reconstructed can be up to half as thin as the collimator width.
Question
Thin collimation has an advantage because it results in faster scanning time.
Question
In high-performance CT scanners, the speed of the data transfer from the detector array determines the maximum resolution.
Question
A disadvantage to having image overlap on CT images is the possibility of artificially creating small lesions.
Question
What is the CT number for water?

A) 0
B) 10
C) 100
D) 1000
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/25
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 12: Ct Physics and Instrumentation
1
What is a reconstruction kernel?

A) A unit of data in a reconstructed CT image
B) The entire data set used to produce reconstructed images in CT
C) A measurement used to determine the data set used for CT reconstruction
D) An algorithm that defines the clinical application and amount of image smoothing applied in the reconstruction
An algorithm that defines the clinical application and amount of image smoothing applied in the reconstruction
2
Important technical developments that define current helical CT systems include all of the following EXCEPT

A) continuous reduction in radiation of the x-ray tube.
B) continuous-motion gantry using slip-ring technology.
C) continuous motion of the imaging table.
D) continuous data transfer from the detector array.
continuous reduction in radiation of the x-ray tube.
3
The purpose of using a filter in CT scanning is

A) to absorb the very-low-energy x-rays.
B) to act as an upper level discriminator.
C) to act as a window.
D) To attenuate the high-energy x-rays.
to absorb the very-low-energy x-rays.
4
In helical CT, pitch is defined as

A) the ratio of the patient's movement through the gantry during one 360-degree beam rotation relative to the beam collimation.
B) the total time for a patient to move through the gantry during a CT scan.
C) the ratio of the beam collimation relative to the mAs for a CT scan.
D) the coordination of a patient's possible movement,
E)g., breathing, during the CT scan relative to slice thickness.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
5
What is the purpose of administering a contrast agent for a CT procedure?

A) Reduce patient dose for longer scans that require thinner slicers
B) Reduce streaking artifacts when scanning high-density objects
C) Reduce attenuating properties of tissues with wide variances in density
D) Enhance detection of lesions and aid in defining normal structures
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
6
In x-ray tubes, the types of reactions that produce x-rays include

A) Bremsstrahlung and scatter.
B) Bremsstrahlung and characteristic.
C) characteristic and scatter.
D) characteristic and refraction.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
7
Using a higher mA

A) results in an increase in image noise.
B) results in an increase in contrast resolution.
C) results in a decrease in dose to the patient.
D) shortens the imaging time for the patient.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
8
Which of the following restricts the width and shapes the beam to the area of interest for a single rotation of the tube and detectors around the patient?

A) Prepatient collimators
B) Postpatient collimators
C) Pitch
D) Rotation speed
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
9
All of the following are true about the kVp setting EXCEPT

A) it selects the maximum possible energy of the x-ray beam.
B) it defines the fraction of photons that will reach the detectors.
C) it affects the radiation exposure to the patient.
D) it increases beam hardening.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
10
In an x-ray tube, when the milliamperes (mA) are increased:

A) Bremsstrahlung and characteristic x-rays increase.
B) Bremsstrahlung and characteristic x-rays decrease.
C) Bremsstrahlung x-rays increases and characteristic x-rays decrease.
D) Bremsstrahlung x-rays decrease and characteristic x-rays increase.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
11
In an x-ray tube, of the energy applied to the system, the percentage of x-rays that exit is

A) 1%.
B) 10%.
C) 50%.
D) 99%.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
12
Which of the following may be used to reduce streaking artifacts when scanning anatomy with high-density objects?

A) Bow-tie filter
B) Adaptive detector array
C) Thinner collimation
D) Decreased kVp
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
13
Computed tomography (CT) scanning accomplishes all of the following EXCEPT

A) minimal superimposition of tissues.
B) improved contrast resolution.
C) reduction of radiation dose compared to conventional radiography.
D) differentiation of tissue densities.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
14
In an x-ray tube, a low kVp results in

A) high-energy Bremsstrahlung x-rays.
B) low-energy Bremsstrahlung x-rays.
C) fewer Bremsstrahlung x-rays.
D) an increase in tungsten characteristic x-rays.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
15
Energies required for typical PET/CT studies of the body will be in the range of

A) 20 to 30 kVp.
B) 40 to 75 kVp.
C) 80 to 140 kVp.
D) 145 to 160 kVp.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
16
A general rule is that a 15% increase in kVp results in

A) a 15% increase in mAs.
B) doubling the mAs.
C) a 15% reduction in mAs.
D) no affect on mAs.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
17
Collimators in CT scanning are used to do all of the following EXCEPT

A) limit the beam size.
B) reduce scattered x-rays.
C) define slice thickness.
D) offset the alignment of the tube and detector array to optimize the beam location.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
18
In an x-ray tube, a high kVp results in

A) low-energy Bremsstrahlung x-rays.
B) a decrease in tungsten characteristic x-rays.
C) an increase in tungsten characteristic x-rays.
D) fewer Bremsstrahlung x-rays.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
19
In an x-ray tube, the anode, for the most part, is most commonly constructed from

A) tungsten.
B) lead.
C) rhenium.
D) silver.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
20
In a CT scanner, the oil bath that may surround the rotating anode tubes serves to

A) make the tubes rotate more quickly.
B) make the tubes rotate more smoothly.
C) cool the tubes.
D) provide efficient conduction of the current.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
21
The slice thickness that may be reconstructed can be up to half as thin as the collimator width.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
22
Thin collimation has an advantage because it results in faster scanning time.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
23
In high-performance CT scanners, the speed of the data transfer from the detector array determines the maximum resolution.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
24
A disadvantage to having image overlap on CT images is the possibility of artificially creating small lesions.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
25
What is the CT number for water?

A) 0
B) 10
C) 100
D) 1000
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 25 flashcards in this deck.