Deck 11: Instrumentation

Full screen (f)
exit full mode
Question
The gas in a dose calibrator is usually

A) air.
B) helium.
C) argon.
D) quenching.
Use Space or
up arrow
down arrow
to flip the card.
Question
When performing quality control on a well counter or thyroid probe, constancy is tested by performing a(n)

A) chi-square test.
B) FWHM.
C) intrinsic resolution check.
D) count for each radionuclide used.
Question
Geometric efficiency is defined

A) as the ratio of the number of detected events to the number of source emissions.
B) as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) as the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) as the amount of blurring that is produced by an imaging system, characterized by point spread function and line spread function.
Question
Which survey meter is appropriate to use when measuring high levels of radiation?

A) Ionization chamber
B) Dose calibrator
C) GM meter
D) Body badge
Question
Quenching gas is used in a(n)

A) dose calibrator.
B) Geiger-Mueller (GM) detector.
C) ionization chamber.
D) pocket dosimeter.
Question
A PHA includes all of the following EXCEPT

A) lower-level discriminator.
B) upper-level discriminator.
C) anticoincidence circuit.
D) amplifier.
Question
In a scintillation camera, geometric resolution depends on

A) the size of the patient.
B) the position of the source.
C) the configuration of the holes in a collimator.
D) filtering the image.
Question
For a parallel-hole collimator, spatial resolution degrades as

A) the source-to-collimator distance increases.
B) the source-to-collimator distance decreases.
C) the collimator holes decrease in size.
D) the collimator holes increase in number.
Question
When performing quality control on a dose calibrator, it is required that the accuracy test is performed

A) daily.
B) quarterly.
C) annually.
D) at installation.
Question
Extrinsic spatial resolution is monitored using

A) FWHM.
B) quadrant bar phantom.
C) flat field collimator.
D) electronic zoom.
Question
The percent full width at half maximum (FWHM) is often used as a measure of

A) precision.
B) sensitivity.
C) intensity.
D) resolution.
Question
The anticoincidence component on a pulse-height analyzer determines if

A) more than one photon at the same energy is counted.
B) two photons are counted at the same time.
C) the lower discriminator is triggered, but not the upper energy discriminator.
D) the upper and lower energy discriminators are triggered simultaneously.
Question
Spatial resolution is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system characterized by point spread function and line spread function.
Question
Which detector is confined for use in detecting radiation where the exposure levels are below 2000 mR/hr?

A) Ionization chamber
B) Dose calibrator
C) GM meter
D) Body badge
Question
When performing quality control on a well counter or thyroid probe, energy resolution is evaluated

A) weekly.
B) monthly.
C) quarterly.
D) annually.
Question
When performing quality control on a dose calibrator, it is recommended to test geometry:

A) daily.
B) quarterly.
C) annually.
D) at installation.
Question
Detector efficiency is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) the geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system, characterized by point spread function and line spread function.
Question
When performing quality control on a dose calibrator, the testing of the correct activity over the entire range of radioactivity used in the laboratory is the

A) accuracy test.
B) geometry test.
C) linearity test.
D) constancy test.
Question
Intrinsic efficiency is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system characterized by point spread function and line spread function.
Question
When performing quality control on a dose calibrator, it is required that the constancy test is performed

A) daily.
B) quarterly.
C) annually.
D) at installation.
Question
When referring to image quality, resolution

A) refers to how well objects can be separated in space.
B) determines whether two objects have different activity concentrations.
C) refers to the ability to see the edge of a structure.
D) occurs because of the random nature of counting statistics.
Question
In general, the pixel size should be less than:

A) one-quarter the system resolution.
B) one-third the system resolution.
C) one-quarter the system efficiency.
D) one-third the system efficiency.
Question
Increased sensitivity for SPECT imaging can be achieved by using a

A) general-purpose collimator.
B) converging collimator.
C) high-energy collimator.
D) high-resolution collimator.
Question
Ramp-filtered projections are characterized by

A) a star artifact.
B) edge information enhancement.
C) smoothing of edge information.
D) enhancing noise.
Question
In SPECT/CT imaging, it is best if the patient does shallow breathing throughout the scan.
Question
Body contouring or elliptical orbits in SPECT imaging have an advantage over circular orbits because they

A) improve the accuracy of gantry location.
B) reduce the camera-to-patient distance.
C) maximize resolution.
D) minimize noise.
Question
In multihole low-energy collimators, longer and smaller holes produce

A) better uniformity.
B) better sensitivity.
C) better intrinsic resolution.
D) better spatial resolution.
Question
When the same nonuniformity is back-projected at each of the angles of acquisition, the result is

A) blurring.
B) a star artifact.
C) a ring artifact.
D) streak artifact.
Question
The center-of-rotation measurement

A) determines the number of projections.
B) corresponds to the pixel size at the center of the orbit.
C) corresponds to the radius.
D) determines the offset between the camera center and projections.
Question
In the reconstruction of SPECT images, segmenting refers to

A) volume rendering.
B) determining the number of counts per pixel.
C) separating the organ from the background or nearby structures.
D) dividing an organ into different components.
Question
A low-pass filter

A) allows low frequencies to pass and blocks high frequencies.
B) blocks low frequencies and allows high frequencies to pass.
C) smoothes images by reducing low frequencies.
D) reduces noise by blocking low frequencies.
Question
Compared to a low-energy collimator, a high-energy collimator will have all of the following EXCEPT

A) thicker septa.
B) longer holes.
C) lower resolution.
D) higher sensitivity.
Question
When referring to image quality, contrast

A) refers to how well objects can be separated in space.
B) determines whether two objects have different activity concentrations.
C) refers to the ability to see the edge of a structure.
D) occurs because of the random nature of counting statistics.
Question
According to the Nuclear Regulatory Commission (NRC), survey meters must be calibrated:

A) daily.
B) monthly.
C) quarterly.
D) annually.
Question
Digitization of the signal of each photomultiplier tube (PMT)

A) increases sensitivity.
B) decreases scan time.
C) increases contrast.
D) increases processing flexibility.
Question
The 360-degree orbit is most often used in SPECT studies EXCEPT for scans of

A) the brain.
B) the heart.
C) the liver.
D) bone.
Question
The devices that operate in the saturation range include all of the following EXCEPT

A) dose calibrator.
B) ionization chamber.
C) personnel dosimeters.
D) Geiger-Mueller counter.
Question
If a detector can produce an electronic signal for each individual interaction whose amplitude reflects the amount of energy absorbed, it is said to have spatial resolution.
Question
Quench gas is used in a Geiger-Mueller counter to

A) enhance ionizations.
B) stop the discharge.
C) alter the voltage.
D) improve the counting of alphas.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/39
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 11: Instrumentation
1
The gas in a dose calibrator is usually

A) air.
B) helium.
C) argon.
D) quenching.
argon.
2
When performing quality control on a well counter or thyroid probe, constancy is tested by performing a(n)

A) chi-square test.
B) FWHM.
C) intrinsic resolution check.
D) count for each radionuclide used.
chi-square test.
3
Geometric efficiency is defined

A) as the ratio of the number of detected events to the number of source emissions.
B) as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) as the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) as the amount of blurring that is produced by an imaging system, characterized by point spread function and line spread function.
as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
4
Which survey meter is appropriate to use when measuring high levels of radiation?

A) Ionization chamber
B) Dose calibrator
C) GM meter
D) Body badge
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
5
Quenching gas is used in a(n)

A) dose calibrator.
B) Geiger-Mueller (GM) detector.
C) ionization chamber.
D) pocket dosimeter.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
6
A PHA includes all of the following EXCEPT

A) lower-level discriminator.
B) upper-level discriminator.
C) anticoincidence circuit.
D) amplifier.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
7
In a scintillation camera, geometric resolution depends on

A) the size of the patient.
B) the position of the source.
C) the configuration of the holes in a collimator.
D) filtering the image.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
8
For a parallel-hole collimator, spatial resolution degrades as

A) the source-to-collimator distance increases.
B) the source-to-collimator distance decreases.
C) the collimator holes decrease in size.
D) the collimator holes increase in number.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
9
When performing quality control on a dose calibrator, it is required that the accuracy test is performed

A) daily.
B) quarterly.
C) annually.
D) at installation.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
10
Extrinsic spatial resolution is monitored using

A) FWHM.
B) quadrant bar phantom.
C) flat field collimator.
D) electronic zoom.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
11
The percent full width at half maximum (FWHM) is often used as a measure of

A) precision.
B) sensitivity.
C) intensity.
D) resolution.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
12
The anticoincidence component on a pulse-height analyzer determines if

A) more than one photon at the same energy is counted.
B) two photons are counted at the same time.
C) the lower discriminator is triggered, but not the upper energy discriminator.
D) the upper and lower energy discriminators are triggered simultaneously.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
13
Spatial resolution is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system characterized by point spread function and line spread function.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
14
Which detector is confined for use in detecting radiation where the exposure levels are below 2000 mR/hr?

A) Ionization chamber
B) Dose calibrator
C) GM meter
D) Body badge
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
15
When performing quality control on a well counter or thyroid probe, energy resolution is evaluated

A) weekly.
B) monthly.
C) quarterly.
D) annually.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
16
When performing quality control on a dose calibrator, it is recommended to test geometry:

A) daily.
B) quarterly.
C) annually.
D) at installation.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
17
Detector efficiency is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) the geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system, characterized by point spread function and line spread function.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
18
When performing quality control on a dose calibrator, the testing of the correct activity over the entire range of radioactivity used in the laboratory is the

A) accuracy test.
B) geometry test.
C) linearity test.
D) constancy test.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
19
Intrinsic efficiency is defined as

A) the ratio of the number of detected events to the number of source emissions.
B) geometric efficiency and is defined as the fraction of emitted photons (gamma or x-rays) that hit the detector; it depends on the detector configuration and its distance from the source.
C) the fraction of gamma rays that are detected when they hit the detector; it depends on the thickness (t) and attenuation coefficient ( μ\mu ) of the detecting material.
D) the amount of blurring that is produced by an imaging system characterized by point spread function and line spread function.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
20
When performing quality control on a dose calibrator, it is required that the constancy test is performed

A) daily.
B) quarterly.
C) annually.
D) at installation.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
21
When referring to image quality, resolution

A) refers to how well objects can be separated in space.
B) determines whether two objects have different activity concentrations.
C) refers to the ability to see the edge of a structure.
D) occurs because of the random nature of counting statistics.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
22
In general, the pixel size should be less than:

A) one-quarter the system resolution.
B) one-third the system resolution.
C) one-quarter the system efficiency.
D) one-third the system efficiency.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
23
Increased sensitivity for SPECT imaging can be achieved by using a

A) general-purpose collimator.
B) converging collimator.
C) high-energy collimator.
D) high-resolution collimator.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
24
Ramp-filtered projections are characterized by

A) a star artifact.
B) edge information enhancement.
C) smoothing of edge information.
D) enhancing noise.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
25
In SPECT/CT imaging, it is best if the patient does shallow breathing throughout the scan.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
26
Body contouring or elliptical orbits in SPECT imaging have an advantage over circular orbits because they

A) improve the accuracy of gantry location.
B) reduce the camera-to-patient distance.
C) maximize resolution.
D) minimize noise.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
27
In multihole low-energy collimators, longer and smaller holes produce

A) better uniformity.
B) better sensitivity.
C) better intrinsic resolution.
D) better spatial resolution.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
28
When the same nonuniformity is back-projected at each of the angles of acquisition, the result is

A) blurring.
B) a star artifact.
C) a ring artifact.
D) streak artifact.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
29
The center-of-rotation measurement

A) determines the number of projections.
B) corresponds to the pixel size at the center of the orbit.
C) corresponds to the radius.
D) determines the offset between the camera center and projections.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
30
In the reconstruction of SPECT images, segmenting refers to

A) volume rendering.
B) determining the number of counts per pixel.
C) separating the organ from the background or nearby structures.
D) dividing an organ into different components.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
31
A low-pass filter

A) allows low frequencies to pass and blocks high frequencies.
B) blocks low frequencies and allows high frequencies to pass.
C) smoothes images by reducing low frequencies.
D) reduces noise by blocking low frequencies.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
32
Compared to a low-energy collimator, a high-energy collimator will have all of the following EXCEPT

A) thicker septa.
B) longer holes.
C) lower resolution.
D) higher sensitivity.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
33
When referring to image quality, contrast

A) refers to how well objects can be separated in space.
B) determines whether two objects have different activity concentrations.
C) refers to the ability to see the edge of a structure.
D) occurs because of the random nature of counting statistics.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
34
According to the Nuclear Regulatory Commission (NRC), survey meters must be calibrated:

A) daily.
B) monthly.
C) quarterly.
D) annually.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
35
Digitization of the signal of each photomultiplier tube (PMT)

A) increases sensitivity.
B) decreases scan time.
C) increases contrast.
D) increases processing flexibility.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
36
The 360-degree orbit is most often used in SPECT studies EXCEPT for scans of

A) the brain.
B) the heart.
C) the liver.
D) bone.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
37
The devices that operate in the saturation range include all of the following EXCEPT

A) dose calibrator.
B) ionization chamber.
C) personnel dosimeters.
D) Geiger-Mueller counter.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
38
If a detector can produce an electronic signal for each individual interaction whose amplitude reflects the amount of energy absorbed, it is said to have spatial resolution.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
39
Quench gas is used in a Geiger-Mueller counter to

A) enhance ionizations.
B) stop the discharge.
C) alter the voltage.
D) improve the counting of alphas.
Unlock Deck
Unlock for access to all 39 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 39 flashcards in this deck.