Deck 5: Quality Control of X-Ray Generators and Ancillary Radiographic Equipment

Full screen (f)
exit full mode
Question
What is the NEMA allowance parameter for an x-ray tube that has a stated focal spot size of 1.5 mm?

A) 50%
B) 40%
C) 30%
D) 20%
Use Space or
up arrow
down arrow
to flip the card.
Question
Conventional tomographic systems should produce images of tissue that lie within ____ mm of the indicated tomographic plane.

A) 5
B) 10
C) 15
D) 20
Question
During AEC evaluation with varying kVp, the AEC system should maintain a constant film density to within ____ optical density units.

A) ±\pm 0.05
B) ±\pm 0.1
C) ±\pm 0.2
D) ±\pm 0.3
Question
Which of the following devices can be used as the sensor in AEC units?

A) Ion chamber
B) Photocell
C) Solid-state detector
D) All of the answer options are correct.
Question
A synchronous spinning top is used on a three-phase unit with a time of 17 ms. Approximately how large of an arc should appear if the timer circuit is accurate?

A) 3 degrees
B) 6 degrees
C) 17 degrees
D) 36 degrees
Question
The x-ray tube must be mounted inside its housing so that the central ray exits the housing to within ____ degree(s) of perpendicular.

A) 1
B) 2
C) 3
D) 4
Question
Path closure in pluridirectional tomographic units should be within ____ of the path length.

A) 2%
B) 5%
C) 10%
D) 15%
Question
Two common quality control tests that should be performed on grids are grid _____ and grid _____ tests.

A) ratio; frequency
B) alignment; uniformity
C) ratio; alignment
D) uniformity; ratio
Question
X-ray system mAs reciprocity has an acceptance parameter of

A) ±\pm 5%
B) ±\pm 10%
C) ±\pm 15%
D) ±\pm 20%
Question
What is the acceptance criterion for a timer circuit with an exposure time shorter than 10 ms?

A) 5%
B) 10%
C) 15%
D) 20%
Question
Collimation systems equipped with positive beam limitation must be accurate to within ____ of the SID.

A) ±\pm 5%
B) ±\pm 4%
C) ±\pm 3%
D) ±\pm 2%
Question
The maximum variation coefficient for reproducibility of x-ray generators is

A) 0.05.
B) 0.07.
C) 0.10.
D) 0.20.
Question
An exposure of 600 mA and 16 ms yields an exposure output of 240 mR. What is the mR/mAs?

A) 2.5
B) 25
C) 250
D) 2500
Question
During testing for grid uniformity, films should be uniform within an optical density of

A) ±\pm 0.4.
B) ±\pm 0.3.
C) ±\pm 0.2.
D) ±\pm 0.1.
Question
The light field and radiation field of a radiographic unit must be congruent to within ____ of the SID.

A) ±\pm 1%
B) ±\pm 2%
C) ±\pm 3%
D) ±\pm 5%
Question
What is the accepted half-value layer (HVL) for an x-ray unit operating at 80 kVp?

A) 2.1 mm Al
B) 2.3 mm Al
C) 2.5 mm Al
D) 2.7 mm Al
Question
The minimum filtration required for general-purpose radiographic systems is _____ mm Al.

A) 2.0
B) 2.1
C) 2.5
D) 3.0
Question
The kilovolts peak (kVp) must be accurate to

A) ±2%.
B) ±5%.
C) ±10%.
D) ±15%.
Question
The tomographic exposure angle indicated on the equipment and the actual tomographic angle must coincide within ____ degrees for units operating at angles greater than 30 degrees.

A) 2
B) 5
C) 10
D) 20
Question
Which of the following tests can be used to determine focal spot size?

A) Safelight test
B) Wire mesh test
C) Wisconsin test cassette
D) Star pattern test
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/20
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 5: Quality Control of X-Ray Generators and Ancillary Radiographic Equipment
1
What is the NEMA allowance parameter for an x-ray tube that has a stated focal spot size of 1.5 mm?

A) 50%
B) 40%
C) 30%
D) 20%
40%
2
Conventional tomographic systems should produce images of tissue that lie within ____ mm of the indicated tomographic plane.

A) 5
B) 10
C) 15
D) 20
A
The level of the tomographic section (fulcrum level) that is indicated on the equipment and the actual level of the tomographic section that is imaged above the tabletop should be the same or within 5 mm.
3
During AEC evaluation with varying kVp, the AEC system should maintain a constant film density to within ____ optical density units.

A) ±\pm 0.05
B) ±\pm 0.1
C) ±\pm 0.2
D) ±\pm 0.3
±\pm 0.3
4
Which of the following devices can be used as the sensor in AEC units?

A) Ion chamber
B) Photocell
C) Solid-state detector
D) All of the answer options are correct.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
5
A synchronous spinning top is used on a three-phase unit with a time of 17 ms. Approximately how large of an arc should appear if the timer circuit is accurate?

A) 3 degrees
B) 6 degrees
C) 17 degrees
D) 36 degrees
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
6
The x-ray tube must be mounted inside its housing so that the central ray exits the housing to within ____ degree(s) of perpendicular.

A) 1
B) 2
C) 3
D) 4
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
7
Path closure in pluridirectional tomographic units should be within ____ of the path length.

A) 2%
B) 5%
C) 10%
D) 15%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
8
Two common quality control tests that should be performed on grids are grid _____ and grid _____ tests.

A) ratio; frequency
B) alignment; uniformity
C) ratio; alignment
D) uniformity; ratio
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
9
X-ray system mAs reciprocity has an acceptance parameter of

A) ±\pm 5%
B) ±\pm 10%
C) ±\pm 15%
D) ±\pm 20%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
10
What is the acceptance criterion for a timer circuit with an exposure time shorter than 10 ms?

A) 5%
B) 10%
C) 15%
D) 20%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
11
Collimation systems equipped with positive beam limitation must be accurate to within ____ of the SID.

A) ±\pm 5%
B) ±\pm 4%
C) ±\pm 3%
D) ±\pm 2%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
12
The maximum variation coefficient for reproducibility of x-ray generators is

A) 0.05.
B) 0.07.
C) 0.10.
D) 0.20.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
13
An exposure of 600 mA and 16 ms yields an exposure output of 240 mR. What is the mR/mAs?

A) 2.5
B) 25
C) 250
D) 2500
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
14
During testing for grid uniformity, films should be uniform within an optical density of

A) ±\pm 0.4.
B) ±\pm 0.3.
C) ±\pm 0.2.
D) ±\pm 0.1.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
15
The light field and radiation field of a radiographic unit must be congruent to within ____ of the SID.

A) ±\pm 1%
B) ±\pm 2%
C) ±\pm 3%
D) ±\pm 5%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
16
What is the accepted half-value layer (HVL) for an x-ray unit operating at 80 kVp?

A) 2.1 mm Al
B) 2.3 mm Al
C) 2.5 mm Al
D) 2.7 mm Al
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
17
The minimum filtration required for general-purpose radiographic systems is _____ mm Al.

A) 2.0
B) 2.1
C) 2.5
D) 3.0
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
18
The kilovolts peak (kVp) must be accurate to

A) ±2%.
B) ±5%.
C) ±10%.
D) ±15%.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
19
The tomographic exposure angle indicated on the equipment and the actual tomographic angle must coincide within ____ degrees for units operating at angles greater than 30 degrees.

A) 2
B) 5
C) 10
D) 20
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
20
Which of the following tests can be used to determine focal spot size?

A) Safelight test
B) Wire mesh test
C) Wisconsin test cassette
D) Star pattern test