Deck 7: Quality Control of Fluoroscopic Equipment

Full screen (f)
exit full mode
Question
At any combination of tube potential and current, fluoroscopic equipment with an automatic brightness control should not produce an entrance skin exposure exceeding ____ R/min.

A) 2.5
B) 5
C) 10
D) 20
Use Space or
up arrow
down arrow
to flip the card.
Question
A photocathode is a device that emits

A) light photons.
B) x-ray photons.
C) electrons.
D) All of the answer options are correct.
Question
Which type of distortion that can occur with tube-type fluoroscopic image intensifiers is most often caused by the presence of strong magnetic fields?

A) Veiling glare
B) S distortion
C) Pincushion distortion
D) Vignetting
Question
The input phosphor of an image intensifier tube converts a(n) _____ image into a(n) _____ image.

A) x-ray; light
B) light; electron
C) electron; light
D) x-ray; electron
Question
The television camera most commonly used to capture the fluoroscopic image from a tube type image intensifier is the:

A) image orthicon.
B) plumbicon.
C) vidicon.
D) charge-coupled device.
Question
A reduction in brightness at the periphery of the image captured with an image intensifier tube is referred to as

A) pincushion distortion.
B) vignetting.
C) barrel distortion.
D) quantum mottle.
Question
Performance testing of fluoroscopic systems must include all the following except

A) x-ray tube housing.
B) automatic film processor.
C) total filtration.
D) automatic brightness control.
Question
Quality control tests of fluoroscopic systems should be performed

A) daily.
B) weekly.
C) semiannually.
D) annually.
Question
Which type of test tool may be used to demonstrate pincushion distortion?

A) Dosimeter
B) Wire mesh tool
C) Ion chamber
D) Homogenous phantom
Question
In standard broadcast television, one image frame consists of

A) one television field.
B) two television fields.
C) 262.5 scan lines.
D) 17 ms.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 7: Quality Control of Fluoroscopic Equipment
1
At any combination of tube potential and current, fluoroscopic equipment with an automatic brightness control should not produce an entrance skin exposure exceeding ____ R/min.

A) 2.5
B) 5
C) 10
D) 20
C
The intensity of the x-ray beam at tabletop should not exceed 10 R/min (2.6 mC kg-1min-1) for units equipped with ABC.
2
A photocathode is a device that emits

A) light photons.
B) x-ray photons.
C) electrons.
D) All of the answer options are correct.
C
The light photons release electrons from the photocathode through the process of photoemission.
3
Which type of distortion that can occur with tube-type fluoroscopic image intensifiers is most often caused by the presence of strong magnetic fields?

A) Veiling glare
B) S distortion
C) Pincushion distortion
D) Vignetting
B
An S distortion artifact is a warping of the image along an S-shaped axis and is the result of strong magnetic fields changing the trajectory of the electrons moving across the image intensifier tube.
4
The input phosphor of an image intensifier tube converts a(n) _____ image into a(n) _____ image.

A) x-ray; light
B) light; electron
C) electron; light
D) x-ray; electron
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
5
The television camera most commonly used to capture the fluoroscopic image from a tube type image intensifier is the:

A) image orthicon.
B) plumbicon.
C) vidicon.
D) charge-coupled device.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
6
A reduction in brightness at the periphery of the image captured with an image intensifier tube is referred to as

A) pincushion distortion.
B) vignetting.
C) barrel distortion.
D) quantum mottle.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
7
Performance testing of fluoroscopic systems must include all the following except

A) x-ray tube housing.
B) automatic film processor.
C) total filtration.
D) automatic brightness control.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
8
Quality control tests of fluoroscopic systems should be performed

A) daily.
B) weekly.
C) semiannually.
D) annually.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
9
Which type of test tool may be used to demonstrate pincushion distortion?

A) Dosimeter
B) Wire mesh tool
C) Ion chamber
D) Homogenous phantom
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
10
In standard broadcast television, one image frame consists of

A) one television field.
B) two television fields.
C) 262.5 scan lines.
D) 17 ms.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.