Deck 11: Quality Control for Magnetic Resonance Imaging Equipment

Full screen (f)
exit full mode
Question
The ability of the MR system to differentiate between adjacent tissues having minimal differences in signal intensities is known as

A) signal-to-noise ratio.
B) high-contrast resolution.
C) low-contrast resolution.
D) spatial linearity.
Use Space or
up arrow
down arrow
to flip the card.
Question
The amount of geometric distortion in the image is the

A) signal-to-noise ratio.
B) image uniformity.
C) spatial resolution.
D) spatial linearity.
Question
An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm2.

A) 0.78
B) 1.56
C) 2.08
D) none of the above
Question
Nonuniformity in the magnetic field can be caused by

A) a homogenous main magnetic field.
B) uniformity in the RF coil.
C) the presence of ferromagnetic material within the bore of the magnet.
D) All of the answer options are correct.
Question
Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.

A) 10%
B) 20%
C) 30%
D) 50%
Question
The ACR slice position deviation in MR scanners should not exceed ____ mm.

A) 1
B) 2
C) 3
D) 5
Question
What is the maximum luminance for soft-copy display devices as suggested by the ACR?

A) Less than 1.2 nit
B) Greater than 50 nit
C) Greater than 90 nit
D) Less than 150 nit
Question
What is the resonant frequency in a unit with a 1.0-T magnet and a gyromagnetic ratio of 42.57 MHz/T?

A) 12.77 MHz
B) 21.29 MHz
C) 42.57 MHz
D) 63.86 MHz
Question
The ability of an MR scanner to resolve small objects is known as

A) signal-to-noise ratio.
B) high-contrast resolution.
C) spatial resolution.
D) spatial linearity.
Question
A change in the strength of the main magnetic field can be caused by

A) cryogen boiloff.
B) external ferromagnetic materials.
C) shim coil failure.
D) All of answer options are correct.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 11: Quality Control for Magnetic Resonance Imaging Equipment
1
The ability of the MR system to differentiate between adjacent tissues having minimal differences in signal intensities is known as

A) signal-to-noise ratio.
B) high-contrast resolution.
C) low-contrast resolution.
D) spatial linearity.
C
Low-contrast resolution is a measure of the MR system's ability to differentiate between adjacent tissues having minimal differences in signal intensities.
2
The amount of geometric distortion in the image is the

A) signal-to-noise ratio.
B) image uniformity.
C) spatial resolution.
D) spatial linearity.
D
Geometric accuracy, also referred to as spatial linearity, refers to the amount of geometric distortion in the image due to displacement or improper scaling of the distance between points being displayed on the image.
3
An acquisition with a 200-mm field of view and 256 x and y encodes yields a pixel size of ____ mm2.

A) 0.78
B) 1.56
C) 2.08
D) none of the above
A
200 mm FOV ÷ 256 frequency encodings = 0.78 mm.
4
Nonuniformity in the magnetic field can be caused by

A) a homogenous main magnetic field.
B) uniformity in the RF coil.
C) the presence of ferromagnetic material within the bore of the magnet.
D) All of the answer options are correct.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
5
Luminance uniformity of soft-copy displays used in MRI systems should be within ____ of the maximum brightness measured in the center of the screen.

A) 10%
B) 20%
C) 30%
D) 50%
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
6
The ACR slice position deviation in MR scanners should not exceed ____ mm.

A) 1
B) 2
C) 3
D) 5
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
7
What is the maximum luminance for soft-copy display devices as suggested by the ACR?

A) Less than 1.2 nit
B) Greater than 50 nit
C) Greater than 90 nit
D) Less than 150 nit
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
8
What is the resonant frequency in a unit with a 1.0-T magnet and a gyromagnetic ratio of 42.57 MHz/T?

A) 12.77 MHz
B) 21.29 MHz
C) 42.57 MHz
D) 63.86 MHz
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
9
The ability of an MR scanner to resolve small objects is known as

A) signal-to-noise ratio.
B) high-contrast resolution.
C) spatial resolution.
D) spatial linearity.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
10
A change in the strength of the main magnetic field can be caused by

A) cryogen boiloff.
B) external ferromagnetic materials.
C) shim coil failure.
D) All of answer options are correct.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 10 flashcards in this deck.