Deck 10: Quality Control in Computed Tomography

Full screen (f)
exit full mode
Question
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 20%
Use Space or
up arrow
down arrow
to flip the card.
Question
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A) 2
B) 3
C) 5
D) 10
Question
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A) ±\pm 0.5
B) ±\pm 1.0
C) ±\pm 2.0
D) ±\pm 5.0
Question
The laser light field and radiation field should coincide to within ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 5.0
Question
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
Question
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A) Low-contrast resolution
B) Uniformity
C) High-contrast resolution
D) Image noise
Question
The CT number of water should not deviate by more than ____ CT numbers from 0.

A) 2
B) 3
C) 5
D) 10
Question
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 3.0
Question
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A) quarterly.
B) monthly.
C) weekly.
D) daily.
Question
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.

A) 20
B) 40
C) 30
D) 80
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 10: Quality Control in Computed Tomography
1
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A) ±\pm 2%
B) ±\pm 5%
C) ±\pm 10%
D) ±\pm 20%
±\pm 10%
2
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A) 2
B) 3
C) 5
D) 10
C
Edge-to-center mean CT number difference must be less than 5 HU for all four edge positions.
3
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A) ±\pm 0.5
B) ±\pm 1.0
C) ±\pm 2.0
D) ±\pm 5.0
±\pm 1.0
4
The laser light field and radiation field should coincide to within ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 5.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
5
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A) daily.
B) weekly.
C) monthly.
D) semi-annually.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
6
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A) Low-contrast resolution
B) Uniformity
C) High-contrast resolution
D) Image noise
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
7
The CT number of water should not deviate by more than ____ CT numbers from 0.

A) 2
B) 3
C) 5
D) 10
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
8
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A) 0.5
B) 1.0
C) 2.0
D) 3.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
9
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A) quarterly.
B) monthly.
C) weekly.
D) daily.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
10
The CTDIvol pass/fail criterion for an adult abdomen is ____ mGy.

A) 20
B) 40
C) 30
D) 80
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 10 flashcards in this deck.