Deck 15: Fluoroscopic Imaging

Full screen (f)
exit full mode
Question
The ________________ is (are) found along the entire length of the image intensifier.

A) input phosphor
B) output phosphor
C) photocathode
D) electrostatic focusing lenses
Use Space or
up arrow
down arrow
to flip the card.
Question
This part of the image intensifier is made of cesium and antimony compounds.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Electrostatic focusing lenses.
Question
Before the image intensifier, the operator may have worn _____ goggles to make it easier to see the dim fluoroscopic image.

A) blue
B) green
C) red
D) yellow
Question
The image intensifier improved fluoroscopy by

A) making the image easier to see because it is brighter
B) making the image easier to see because it is bigger
C) allowing indirect viewing of the fluoroscopic image
D) making the image easier to see because it is brighter and allowing indirect viewing of the fluoroscopic image
Question
The __________________________ absorbs x-rays and emits light.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Question
The electrostatic focusing lenses have a ___________ charge.

A) negative
B) positive
C) neutral
D) changing
Question
The commercial fluoroscope was developed by

A) Rutherford
B) Edison
C) Newton
D) Einstein
Question
The ratio of the number of light photons at the output phosphor as compared with the input phosphor is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Question
This part of the image intensifier is made of silver-activated zinc cadmium sulfide.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Accelerating anode.
Question
The image intensifier is located

A) next to the x-ray tube used for fluoroscopy
B) beneath the x-ray table
C) on a shelf just inside the room
D) inside the fluoroscopic tower
Question
The __________________ absorbs electrons and emits light.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Question
This part of the image intensifier is made of cesium iodide.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Accelerating anode.
Question
The image intensifier was introduced in the

A) 1920s
B) 1950s
C) 1970s
D) 1980s
Question
The _______________________ is bonded directly to the input phosphor.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Question
The electronic vacuum tube that converts the remnant x-ray beam to light, then to electrons, then back to light, increasing the light intensity in the process is the

A) x-ray tube
B) image intensifier
C) recording system
D) mirror optics
Question
The _____________________ is designed to set the electron stream in motion at a constant velocity, and is located very close to the output phosphor.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Question
___________ imaging is used to view the human body in real time, accomplished with fluoroscopy as opposed to radiography.

A) Static
B) Still
C) Dynamic
D) Colorized
Question
Photoemission occurs at the

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Question
The original fluoroscopic image was very

A) dim
B) bright
C) clear
D) fuzzy
Question
This part of the image intensifier absorbs light and produces electrons.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Electrostatic focusing lenses.
Question
The function of the fluoroscopic unit that maintains the overall appearance of the fluoroscopic image (contrast and brightness) by automatically adjusting the kilovoltage peak (kVp), milliamperage (mA), or both is

A) DQE
B) APR
C) ABC
D) AEC
Question
____________ is measured in cd/m2/mR/s.

A) Brightness gain
B) Flux gain
C) Minification gain
D) Conversion factor
Question
The formula to determine the amount of magnification created when in magnification mode, the formula is

A) MF = full size input phosphor / selected input phosphor
B) MF = full size input phosphor2 / selected input phosphor2
C) MF = selected input phosphor / full size input phosphor
D) MF = selected input phosphor2 / full size input phosphor2
Question
In fluoroscopy, shape distortion is caused by

A) angling the x-ray tube
B) angling the image intensifier in the fluoroscopic tower
C) inaccurate control or focusing of the electrons released at the periphery of the photocathode, and the curved shape of the photocathode
D) inaccurate control or focusing of the electrons released at the periphery of the photocathode, and the flat shape of the photocathode
Question
If a 30/23/15 cm image intensifier is operated in the 23-cm mode, the fluoroscopic image will be magnified by a factor of

A) 1.3
B) 1.5
C) 1.7
D) 2
Question
If the input phosphor measures 30 cm and the output phosphor measures 2.5 cm, what is the minification gain?

A) 8.3
B) 12
C) 69.2
D) 144
Question
A loss of brightness around the edge of the fluoroscopic image caused by the curve of the photocathode is

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Question
The formula for brightness gain is

A) brightness gain = minification gain - flux gain
B) brightness gain = minification gain / flux gain
C) brightness gain = minification gain * flux gain
D) None of these
Question
Typical fluoroscopic systems have spatial resolution capabilities in the range of

A) 2-3 Lp/mm
B) 4-6 Lp/mm
C) 7-9 Lp/mm
D) 10-12 Lp/mm
Question
Distortion of the fluoroscopic image that appears as unequal magnification is

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Question
The diameter of the input phosphor typically measures between

A) 5 and 10 cm
B) 15 and 30 cm
C) 40 and 65 cm
D) 75 and 100 cm
Question
A disadvantage of using magnification mode during fluoroscopy is

A) the image has less resolution
B) the image brightness is inconsistent
C) the patient receives additional dose
D) the image has increased noise
Question
Increasing the mA is the way to correct a fluoroscopic image that has

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Question
The diameter of the output phosphor is approximately

A) 1 cm
B) 2.5 cm
C) 5 cm
D) 25 cm
Question
The term that means an expression of the degree to which the image is made smaller from input phosphor to output phosphor. is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Question
The formula for minification gain is

A) minification gain = input phosphor diameter / output phosphor diameter
B) minification gain = input phosphor diameter2 / output phosphor diameter2
C) minification gain = output phosphor diameter / input phosphor diameter
D) minification gain = output phosphor diameter2 / input phosphor diameter2
Question
When operated in magnification mode, the electrostatic lenses have ________ of a negative charge.

A) less
B) more
C) the same amount as usual
D) one half
Question
Magnification of the fluoroscopic image results in improved

A) brightness
B) spatial resolution
C) contrast
D) exposure
Question
Fluoroscopic imaging systems operate at: ________

A) 2-5 mA
B) 15-20 mA
C) 200-500 mA
D) Unlimited mA
Question
An expression of the ability of an image intensifier tube to convert x-ray energy into light energy and increase the brightness of the image in the process is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Question
The _____________ is a diode tube contained in a glass envelope to maintain a vacuum, and is connected to the output phosphor of the image intensifier by either a fiberoptic bundle or an optical lens system.

A) television monitor
B) Vidicon camera tube
C) charge coupled device (CCD)
D) liquid crystal display (LCD)
Question
Typical television monitors are

A) 300 line systems
B) 450 line systems
C) 525 line systems
D) 875 line systems
Question
When the electron beam is sweeping the anode, if the electron beam and light from the output phosphor are incident on the same place at the same time, _____________ through the target to the signal plate.

A) x-rays are produced
B) the image is magnified
C) electrons are transmitted
D) light photons are transmitted
Question
A light-sensitive semiconducting device that generates an electrical charge when stimulated by light and stores this charge in a capacitor is the

A) anode of the camera tube
B) charge coupled device (CCD)
C) cathode of the camera tube
D) x-ray tube
Question
In image-intensified fluoroscopy units the method for viewing the fluoroscopic image is by

A) using a handheld device
B) looking directly at the output phosphor
C) using a television monitor
D) using a mirror-optics system
Question
The target layer is a photoconductive layer made of antimony trisulfide in ______ tubes.

A) x-ray
B) CCD
C) vidicon
D) plumbicon
Question
This recording system uses 70-mm or 105-mm film.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of these
Question
An electron gun can be found in the

A) television monitor
B) CCD
C) vidicon camera
D) television monitor and vidicon camera
Question
In terms of resolution, the weakest part of the fluoroscopic system is the

A) image intensifier
B) television monitor
C) camera tube
D) CCD
Question
Semiconductor capacitors made of metal oxide are components of the

A) camera tube
B) CCD
C) image intensifier
D) x-ray tube
Question
The CCDs electric charge from the capacitors is sent to the

A) output phosphor
B) television monitor
C) x-ray tube
D) image intensifier
Question
The purpose of the is to convert the electronic signal from the camera tube or CCD into a visible image.

A) image intensifier
B) fiberoptic bundle
C) television monitor
D) anode of the vidicon tube
Question
Each capacitor in a CCD represents a(n)

A) line
B) image
C) pixel
D) electric charge
Question
The fiberoptic bundle or optical lens system is used to

A) deliver the image from the camera to the television monitor
B) record the fluoroscopic image
C) link the output phosphor and camera or CCD
D) none of these
Question
The television is capable of resolving approximately ___________.

A) 0.5-.75 Lp/mm
B) 1-2 Lp/mm
C) 3 Lp/mm
D) 5 Lp/mm
Question
When the electron beam sweeps the anode and travels back and forth across, from top to bottom, it is moving in a ___________ pattern.

A) sweep
B) brush
C) raster
D) rooster
Question
A face plate, signal plate, and target are all components of the

A) anode of the camera tube
B) charge coupled device (CCD)
C) cathode of the camera tube
D) x-ray tube
Question
A typical high-resolution monitor has

A) 825 line systems
B) 1024 line systems
C) 525 line systems
D) 2034 line systems
Question
The purpose of a beam-splitting mirror is to

A) view the fluoroscopic image on more than one television
B) reduce the intensity of the image from the output phosphor
C) allow spot filming to be done during fluoroscopy
D) all of these
Question
The recording device that "photographs" the image off of the output phosphor is the

A) cassette spot filming
B) film cameras for spot filming
C) videotape
D) flat panel detector
Question
Which of the following is true about the CCD used in digital fluoroscopy? I. The CCD is more light sensitive with a higher DQE value. II. The CCD exhibits less noise and no spatial distortion. III. The CCD has higher spatial resolution. IV. The CCD requires more radiation to create an image.

A) the CCD is more light sensitive with a higher DQE value and CCD exhibits less noise and no spatial distortion only.
B) the CCD is more light sensitive with a higher DQE value and the CCD has higher spatial resolution only.
C) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, and the CCD has higher spatial resolution only.
D) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, and the CCD requires more radiation to create an image only.
E) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, the CCD has higher spatial resolution, and the CCD requires more radiation to create an image.
Question
What fluoroscopic feature allows the position of the collimator plates to be viewed on the monitor without exposing the patient to radiation?

A) last image hold
B) frame averaging
C) digital subtraction
D) electronic magnification
Question
A large fill factor refers to

A) the entire detector being filled with x-rays
B) the entire detector needing a large amount of x-rays to produce an image
C) a large area of the pixel being sensitive to x-rays
D) a small area of the pixel being sensitive to x-rays
Question
Digital fluoroscopy is improved by using

A) a vidicon camera
B) videotape
C) a beam-splitting mirror
D) a charge-coupling device (CCD)
Question
This allows the film to be divided into 2, 4, or more images.

A) Cassette spot filming with masking shutters
B) Film cameras for spot filming
C) Videotape
D) Flat panel detector
Question
More advanced fluoroscopy systems use a flat-panel detector

A) to record the spot images during fluoroscopy
B) to record the overhead images following fluoroscopy
C) in place of the image intensifier
D) in addition to the image intensifier
Question
Which of the following recording systems require the x-ray-beam to be in radiographic mode?

A) Cassette spot filming.
B) Film cameras for spot filming.
C) Cine filming.
D) Flat panel detector.
Question
For detectors to produce high-quality fluoroscopic images, they must be able to

A) respond very quickly
B) maintain a large fill factor
C) have application-specific integrated circuits for noise reduction
D) all of these
Question
A fluoroscopic feature that automatically turns the x-ray beam on and off rapidly during operations is known as:

A) Pulsed fluoroscopy
B) Frame averaging
C) Intermittent fluoroscopy
D) Pulse rate
Question
Flat-panel detectors for fluoroscopic imaging can be the

A) cesium iodide amorphous silicon indirect-capture detector
B) amorphous selenium direct-capture detector
C) cesium and antimony direct-capture detector
D) cesium iodide amorphous silicon indirect-capture detector and amorphous selenium direct-capture detector
E) cesium iodide amorphous silicon indirect-capture detector and cesium and antimony direct-capture detector
Question
Frame averaging is

A) an operation that reduces overall patient dose
B) an operation that increases overall patient dose
C) an operation that reduces image noise by averaging multiple image frames together
D) an operation that reduces overall patient dose and an operation that reduces image noise by averaging multiple image frames together
Question
Which of the following is required in the indirect-capture detector?

A) Amorphous silicon photodetector.
B) Thin film transistor (TFT) array.
C) A scintillator that uses cesium iodide or gadolinium oxysulfide as the phosphor.
D) All of these.
Question
Which of the following has a rectangular field of view and a wider dynamic range?

A) Conventional image intensifier.
B) Flat-panel detector in place of image intensifier.
C) There is no difference between the two.
Question
The difference between geometric magnification and electronic magnification is

A) geometric magnification is where the patient is placed closer to the x-ray tube to create a larger OID and electronic magnification involves the selection of a smaller field of view
B) electronic magnification is where the patient is placed closer to the x-ray tube to create a larger OID and geometric magnification involves the selection of a smaller field of view
C) there is no difference between geometric magnification and electronic magnification
Question
Which of the following is an advantage of flat panel detectors used in fluoroscopy?

A) Much lighter and more compact.
B) Produce a digital signal directly.
C) Less electronic noise.
D) 60 times larger operational dynamic range because they do not exhibit veiling glare.
E) All of these.
Question
The ability to adjust the size of the field of view without exposing the patient to radiation is known as:

A) frame averaging
B) last image hold
C) digital subtraction
D) virtual collimation
Question
Flat panel detector fluoroscopy units operate at what mA?

A) 2- 5 mA
B) 10-30 mA
C) 50-1200 mA
D) 1500-2000 mA
Question
What feature automatically adjusts the tube current (mA), voltage (kVp), filtration, and pulse width to maintain radiation exposure to the flat panel detector?

A) Pulsed fluoroscopy
B) Last image hold
C) Frame averaging
D) None of these
Question
The analog-to-digital converter

A) takes the video (analog) signal and divides it into binary language that the computer "understands."
B) determines the contrast resolution of the system
C) is necessary for the computer to process and display the image
D) all of these
Question
Early versions of digital fluoroscopy used the conventional fluoroscopic chain and added

A) a different camera
B) an analog-to-digital converter (ADC)
C) videotape
D) a digital-to-analog converter (DAC)
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/100
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 15: Fluoroscopic Imaging
1
The ________________ is (are) found along the entire length of the image intensifier.

A) input phosphor
B) output phosphor
C) photocathode
D) electrostatic focusing lenses
electrostatic focusing lenses
2
This part of the image intensifier is made of cesium and antimony compounds.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Electrostatic focusing lenses.
Photocathode.
3
Before the image intensifier, the operator may have worn _____ goggles to make it easier to see the dim fluoroscopic image.

A) blue
B) green
C) red
D) yellow
red
4
The image intensifier improved fluoroscopy by

A) making the image easier to see because it is brighter
B) making the image easier to see because it is bigger
C) allowing indirect viewing of the fluoroscopic image
D) making the image easier to see because it is brighter and allowing indirect viewing of the fluoroscopic image
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
5
The __________________________ absorbs x-rays and emits light.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
6
The electrostatic focusing lenses have a ___________ charge.

A) negative
B) positive
C) neutral
D) changing
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
7
The commercial fluoroscope was developed by

A) Rutherford
B) Edison
C) Newton
D) Einstein
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
8
The ratio of the number of light photons at the output phosphor as compared with the input phosphor is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
9
This part of the image intensifier is made of silver-activated zinc cadmium sulfide.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Accelerating anode.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
10
The image intensifier is located

A) next to the x-ray tube used for fluoroscopy
B) beneath the x-ray table
C) on a shelf just inside the room
D) inside the fluoroscopic tower
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
11
The __________________ absorbs electrons and emits light.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
12
This part of the image intensifier is made of cesium iodide.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Accelerating anode.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
13
The image intensifier was introduced in the

A) 1920s
B) 1950s
C) 1970s
D) 1980s
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
14
The _______________________ is bonded directly to the input phosphor.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
15
The electronic vacuum tube that converts the remnant x-ray beam to light, then to electrons, then back to light, increasing the light intensity in the process is the

A) x-ray tube
B) image intensifier
C) recording system
D) mirror optics
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
16
The _____________________ is designed to set the electron stream in motion at a constant velocity, and is located very close to the output phosphor.

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
17
___________ imaging is used to view the human body in real time, accomplished with fluoroscopy as opposed to radiography.

A) Static
B) Still
C) Dynamic
D) Colorized
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
18
Photoemission occurs at the

A) input phosphor
B) output phosphor
C) photocathode
D) accelerating anode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
19
The original fluoroscopic image was very

A) dim
B) bright
C) clear
D) fuzzy
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
20
This part of the image intensifier absorbs light and produces electrons.

A) Input phosphor.
B) Output phosphor.
C) Photocathode.
D) Electrostatic focusing lenses.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
21
The function of the fluoroscopic unit that maintains the overall appearance of the fluoroscopic image (contrast and brightness) by automatically adjusting the kilovoltage peak (kVp), milliamperage (mA), or both is

A) DQE
B) APR
C) ABC
D) AEC
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
22
____________ is measured in cd/m2/mR/s.

A) Brightness gain
B) Flux gain
C) Minification gain
D) Conversion factor
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
23
The formula to determine the amount of magnification created when in magnification mode, the formula is

A) MF = full size input phosphor / selected input phosphor
B) MF = full size input phosphor2 / selected input phosphor2
C) MF = selected input phosphor / full size input phosphor
D) MF = selected input phosphor2 / full size input phosphor2
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
24
In fluoroscopy, shape distortion is caused by

A) angling the x-ray tube
B) angling the image intensifier in the fluoroscopic tower
C) inaccurate control or focusing of the electrons released at the periphery of the photocathode, and the curved shape of the photocathode
D) inaccurate control or focusing of the electrons released at the periphery of the photocathode, and the flat shape of the photocathode
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
25
If a 30/23/15 cm image intensifier is operated in the 23-cm mode, the fluoroscopic image will be magnified by a factor of

A) 1.3
B) 1.5
C) 1.7
D) 2
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
26
If the input phosphor measures 30 cm and the output phosphor measures 2.5 cm, what is the minification gain?

A) 8.3
B) 12
C) 69.2
D) 144
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
27
A loss of brightness around the edge of the fluoroscopic image caused by the curve of the photocathode is

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
28
The formula for brightness gain is

A) brightness gain = minification gain - flux gain
B) brightness gain = minification gain / flux gain
C) brightness gain = minification gain * flux gain
D) None of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
29
Typical fluoroscopic systems have spatial resolution capabilities in the range of

A) 2-3 Lp/mm
B) 4-6 Lp/mm
C) 7-9 Lp/mm
D) 10-12 Lp/mm
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
30
Distortion of the fluoroscopic image that appears as unequal magnification is

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
31
The diameter of the input phosphor typically measures between

A) 5 and 10 cm
B) 15 and 30 cm
C) 40 and 65 cm
D) 75 and 100 cm
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
32
A disadvantage of using magnification mode during fluoroscopy is

A) the image has less resolution
B) the image brightness is inconsistent
C) the patient receives additional dose
D) the image has increased noise
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
33
Increasing the mA is the way to correct a fluoroscopic image that has

A) noise
B) pincushion appearance
C) vignetting
D) magnification
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
34
The diameter of the output phosphor is approximately

A) 1 cm
B) 2.5 cm
C) 5 cm
D) 25 cm
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
35
The term that means an expression of the degree to which the image is made smaller from input phosphor to output phosphor. is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
36
The formula for minification gain is

A) minification gain = input phosphor diameter / output phosphor diameter
B) minification gain = input phosphor diameter2 / output phosphor diameter2
C) minification gain = output phosphor diameter / input phosphor diameter
D) minification gain = output phosphor diameter2 / input phosphor diameter2
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
37
When operated in magnification mode, the electrostatic lenses have ________ of a negative charge.

A) less
B) more
C) the same amount as usual
D) one half
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
38
Magnification of the fluoroscopic image results in improved

A) brightness
B) spatial resolution
C) contrast
D) exposure
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
39
Fluoroscopic imaging systems operate at: ________

A) 2-5 mA
B) 15-20 mA
C) 200-500 mA
D) Unlimited mA
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
40
An expression of the ability of an image intensifier tube to convert x-ray energy into light energy and increase the brightness of the image in the process is the

A) brightness gain
B) flux gain
C) minification gain
D) conversion factor
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
41
The _____________ is a diode tube contained in a glass envelope to maintain a vacuum, and is connected to the output phosphor of the image intensifier by either a fiberoptic bundle or an optical lens system.

A) television monitor
B) Vidicon camera tube
C) charge coupled device (CCD)
D) liquid crystal display (LCD)
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
42
Typical television monitors are

A) 300 line systems
B) 450 line systems
C) 525 line systems
D) 875 line systems
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
43
When the electron beam is sweeping the anode, if the electron beam and light from the output phosphor are incident on the same place at the same time, _____________ through the target to the signal plate.

A) x-rays are produced
B) the image is magnified
C) electrons are transmitted
D) light photons are transmitted
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
44
A light-sensitive semiconducting device that generates an electrical charge when stimulated by light and stores this charge in a capacitor is the

A) anode of the camera tube
B) charge coupled device (CCD)
C) cathode of the camera tube
D) x-ray tube
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
45
In image-intensified fluoroscopy units the method for viewing the fluoroscopic image is by

A) using a handheld device
B) looking directly at the output phosphor
C) using a television monitor
D) using a mirror-optics system
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
46
The target layer is a photoconductive layer made of antimony trisulfide in ______ tubes.

A) x-ray
B) CCD
C) vidicon
D) plumbicon
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
47
This recording system uses 70-mm or 105-mm film.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
48
An electron gun can be found in the

A) television monitor
B) CCD
C) vidicon camera
D) television monitor and vidicon camera
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
49
In terms of resolution, the weakest part of the fluoroscopic system is the

A) image intensifier
B) television monitor
C) camera tube
D) CCD
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
50
Semiconductor capacitors made of metal oxide are components of the

A) camera tube
B) CCD
C) image intensifier
D) x-ray tube
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
51
The CCDs electric charge from the capacitors is sent to the

A) output phosphor
B) television monitor
C) x-ray tube
D) image intensifier
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
52
The purpose of the is to convert the electronic signal from the camera tube or CCD into a visible image.

A) image intensifier
B) fiberoptic bundle
C) television monitor
D) anode of the vidicon tube
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
53
Each capacitor in a CCD represents a(n)

A) line
B) image
C) pixel
D) electric charge
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
54
The fiberoptic bundle or optical lens system is used to

A) deliver the image from the camera to the television monitor
B) record the fluoroscopic image
C) link the output phosphor and camera or CCD
D) none of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
55
The television is capable of resolving approximately ___________.

A) 0.5-.75 Lp/mm
B) 1-2 Lp/mm
C) 3 Lp/mm
D) 5 Lp/mm
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
56
When the electron beam sweeps the anode and travels back and forth across, from top to bottom, it is moving in a ___________ pattern.

A) sweep
B) brush
C) raster
D) rooster
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
57
A face plate, signal plate, and target are all components of the

A) anode of the camera tube
B) charge coupled device (CCD)
C) cathode of the camera tube
D) x-ray tube
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
58
A typical high-resolution monitor has

A) 825 line systems
B) 1024 line systems
C) 525 line systems
D) 2034 line systems
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
59
The purpose of a beam-splitting mirror is to

A) view the fluoroscopic image on more than one television
B) reduce the intensity of the image from the output phosphor
C) allow spot filming to be done during fluoroscopy
D) all of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
60
The recording device that "photographs" the image off of the output phosphor is the

A) cassette spot filming
B) film cameras for spot filming
C) videotape
D) flat panel detector
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
61
Which of the following is true about the CCD used in digital fluoroscopy? I. The CCD is more light sensitive with a higher DQE value. II. The CCD exhibits less noise and no spatial distortion. III. The CCD has higher spatial resolution. IV. The CCD requires more radiation to create an image.

A) the CCD is more light sensitive with a higher DQE value and CCD exhibits less noise and no spatial distortion only.
B) the CCD is more light sensitive with a higher DQE value and the CCD has higher spatial resolution only.
C) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, and the CCD has higher spatial resolution only.
D) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, and the CCD requires more radiation to create an image only.
E) the CCD is more light sensitive with a higher DQE value, CCD exhibits less noise and no spatial distortion, the CCD has higher spatial resolution, and the CCD requires more radiation to create an image.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
62
What fluoroscopic feature allows the position of the collimator plates to be viewed on the monitor without exposing the patient to radiation?

A) last image hold
B) frame averaging
C) digital subtraction
D) electronic magnification
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
63
A large fill factor refers to

A) the entire detector being filled with x-rays
B) the entire detector needing a large amount of x-rays to produce an image
C) a large area of the pixel being sensitive to x-rays
D) a small area of the pixel being sensitive to x-rays
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
64
Digital fluoroscopy is improved by using

A) a vidicon camera
B) videotape
C) a beam-splitting mirror
D) a charge-coupling device (CCD)
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
65
This allows the film to be divided into 2, 4, or more images.

A) Cassette spot filming with masking shutters
B) Film cameras for spot filming
C) Videotape
D) Flat panel detector
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
66
More advanced fluoroscopy systems use a flat-panel detector

A) to record the spot images during fluoroscopy
B) to record the overhead images following fluoroscopy
C) in place of the image intensifier
D) in addition to the image intensifier
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
67
Which of the following recording systems require the x-ray-beam to be in radiographic mode?

A) Cassette spot filming.
B) Film cameras for spot filming.
C) Cine filming.
D) Flat panel detector.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
68
For detectors to produce high-quality fluoroscopic images, they must be able to

A) respond very quickly
B) maintain a large fill factor
C) have application-specific integrated circuits for noise reduction
D) all of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
69
A fluoroscopic feature that automatically turns the x-ray beam on and off rapidly during operations is known as:

A) Pulsed fluoroscopy
B) Frame averaging
C) Intermittent fluoroscopy
D) Pulse rate
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
70
Flat-panel detectors for fluoroscopic imaging can be the

A) cesium iodide amorphous silicon indirect-capture detector
B) amorphous selenium direct-capture detector
C) cesium and antimony direct-capture detector
D) cesium iodide amorphous silicon indirect-capture detector and amorphous selenium direct-capture detector
E) cesium iodide amorphous silicon indirect-capture detector and cesium and antimony direct-capture detector
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
71
Frame averaging is

A) an operation that reduces overall patient dose
B) an operation that increases overall patient dose
C) an operation that reduces image noise by averaging multiple image frames together
D) an operation that reduces overall patient dose and an operation that reduces image noise by averaging multiple image frames together
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
72
Which of the following is required in the indirect-capture detector?

A) Amorphous silicon photodetector.
B) Thin film transistor (TFT) array.
C) A scintillator that uses cesium iodide or gadolinium oxysulfide as the phosphor.
D) All of these.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
73
Which of the following has a rectangular field of view and a wider dynamic range?

A) Conventional image intensifier.
B) Flat-panel detector in place of image intensifier.
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
74
The difference between geometric magnification and electronic magnification is

A) geometric magnification is where the patient is placed closer to the x-ray tube to create a larger OID and electronic magnification involves the selection of a smaller field of view
B) electronic magnification is where the patient is placed closer to the x-ray tube to create a larger OID and geometric magnification involves the selection of a smaller field of view
C) there is no difference between geometric magnification and electronic magnification
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
75
Which of the following is an advantage of flat panel detectors used in fluoroscopy?

A) Much lighter and more compact.
B) Produce a digital signal directly.
C) Less electronic noise.
D) 60 times larger operational dynamic range because they do not exhibit veiling glare.
E) All of these.
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
76
The ability to adjust the size of the field of view without exposing the patient to radiation is known as:

A) frame averaging
B) last image hold
C) digital subtraction
D) virtual collimation
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
77
Flat panel detector fluoroscopy units operate at what mA?

A) 2- 5 mA
B) 10-30 mA
C) 50-1200 mA
D) 1500-2000 mA
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
78
What feature automatically adjusts the tube current (mA), voltage (kVp), filtration, and pulse width to maintain radiation exposure to the flat panel detector?

A) Pulsed fluoroscopy
B) Last image hold
C) Frame averaging
D) None of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
79
The analog-to-digital converter

A) takes the video (analog) signal and divides it into binary language that the computer "understands."
B) determines the contrast resolution of the system
C) is necessary for the computer to process and display the image
D) all of these
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
80
Early versions of digital fluoroscopy used the conventional fluoroscopic chain and added

A) a different camera
B) an analog-to-digital converter (ADC)
C) videotape
D) a digital-to-analog converter (DAC)
Unlock Deck
Unlock for access to all 100 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 100 flashcards in this deck.