Deck 11: Equipment Design for Radiation Protection

Full screen (f)
exit full mode
Question
Digital radiography images can be accessed:

A)at only one workstation at a time,thereby decreasing convenience for physicians providing patient care.
B)at no more than two workstations at a time,thereby decreasing convenience for physicians providing patient care.
C)at several workstations at the same time,making image viewing very convenient for physicians providing patient care.
D)digital images cannot be accessed at any workstations by physicians.
Use Space or
up arrow
down arrow
to flip the card.
Question
Depending on the area of the body being examined with a fluoroscopic image intensification system,for adult patients,a range of ______ kVp is generally used.

A)30 to 60
B)60 to 75
C)75 to 110
D)110 to 140
Question
Compared with the traditional material in the front of a cassette (aluminum or cardboard),the cassette front containing the carbon fiber absorbs approximately ______ as much radiation,resulting in a __________ dose for the patient because ________ radiographic techniques are required to produce the recorded image.

A)four times,higher,higher
B)three times,higher,higher
C)twice,higher,higher
D)half,lower,lower
Question
Federal government specifications recommend a minimum total filtration of _______________ for stationary (fixed) fluoroscopic x-ray units operating above 70 kVp.

A)1.0-mm aluminum equivalent
B)1.5-mm aluminum equivalent
C)2.0-mm aluminum equivalent
D)2.5-mm aluminum equivalent
Question
During a fluoroscopic examination,a resettable cumulative timing device times the x-ray beam-on time and sounds an audible alarm or temporarily interrupts the exposure after the fluoroscope has been activated for what length of time?

A)1 minute
B)3 minutes
C)5 minutes
D)10 minutes
Question
Computed radiography,because of its higher exposure latitude,makes grid use on the pediatric population:

A)absolutely necessary for all radiographic examinations.
B)less necessary than was previously believed.
C)necessary only for extremity examinations.
D)unnecessary for any examination.
Question
Sharper size restriction of the radiographic beam is achieved when a cone or cylinder is:

A)longer.
B)shorter.
C)made of aluminum.
D)made of durable plastic.
Question
When obtaining a dorsoplantar projection of a foot,which of the following types of filters should be used to provide a greater uniform density of the anatomy?

A)Bilateral wedge filter
B)Trough filter
C)Thoraeus filter
D)Wedge filter
Question
Current federal standards limit entrance skin exposure rates of general-purpose intensified fluoroscopic units to a maximum of ______________ measured at tabletop with the image intensifier entrance surface at a prescribed 30 cm above.

A)100 mGya/min
B)300 mGya/min
C)500 mGya/min
D)600 mGya/min
Question
When the speed of screen-film systems doubles,for example,when changing from a 200-speed to a 400-speed system,patient radiation exposure is ______ by approximately ______.

A)increased,25%
B)increased,50%
C)reduced,25%
D)reduced,50%
Question
CR imaging has greater ____________ flexibility than conventional screen-film radiography.

A)kilovoltage
B)milliamperage
C)time of exposure
D)x-ray beam collimation
Question
In standard image intensification fluoroscopy,an x-ray beam half-value-layer (HVL) of 3 to 4.5 mm aluminum is considered acceptable when peak kilovoltage ranges from:

A)50 to 60.
B)60 to 70.
C)70 to 80.
D)80 to 100.
Question
When an exposed computed radiography imaging plate is ready to be processed,an imaging reading unit is used to scan the photostimulable phosphor imaging plate with a helium-neon laser beam.This process results in the emission of violet light that is changed into an electronic signal by a device called a:

A)charge-coupled-device array.
B)photodiode.
C)photomultiplier tube.
D)scintillator.
Question
A radiographer uses a high-speed,rare-earth screen-film system with optimal technical exposure factors to obtain PA and lateral radiographs of a 2-year-old child's chest.To maximize reduction of radiation exposure to the patient following adequate immobilization,the radiographer should:

A)collimate the x-ray beam so that it is only slightly larger than the margins of the image receptor and shield the child's reproductive organs.
B)collimate the x-ray beam so that it is no larger than the margins of the image receptor and shield the child's reproductive organs.
C)collimate the x-ray beam so that it is smaller than the margins of the image receptor and shield the child's reproductive organs.
D)collimate the x-ray beam so that it is smaller than the margins of the image receptor and not shield the child's reproductive organs.
Question
What is the purpose of radiographic beam filtration?

A)To decrease beam hardness,thereby reducing patient skin dose and the dose to superficial tissues
B)To increase beam hardness,thereby reducing patient skin dose and the dose to superficial tissues
C)To eliminate short wavelength radiation to reduce the radiation received by the patient's skin and superficial tissues
D)To increase beam hardness,thereby increasing patient skin dose with only a slight increase to superficial tissues
Question
Which of the following is not an x-ray beam limitation device?

A)Aperture diaphragm
B)Collimator
C)Filter
D)Extension cylinder
Question
When using digital fluoroscopy systems,making use of the last-image-hold feature can:

A)be an effective dose reduction technique.
B)eliminate the need to collimate the primary beam.
C)eliminate increased involuntary patient motion.
D)eliminate increased voluntary patient motion.
Question
When a fluoroscopic image is electronically amplified by the use of an image intensification system,which of the following benefits results?
1)Increased image brightness
2)Saving of time for the radiologist
3)Patient dose reduction

A)1 and 2 only
B)1 and 3 only
C)2 and 3 only
D)1,2,and 3
Question
When performing a mobile fluoroscopic procedure,to reduce the radiation exposure to the patient,the radiographer must use a minimal source-skin distance of:

A)15 cm.
B)30 cm.
C)38 cm.
D)45 cm.
Question
To visualize smaller and lower contrast objects during interventional procedures,high-level-control fluoroscopy uses exposure rates that are ____________ those normally used in routine fluoroscopy.

A)higher than
B)lower than
C)the same as
D)none of the above because smaller and lower contrast objects cannot be visualized by fluoroscopy during interventional procedures
Question
Luminance is determined by measuring the concentration of light:

A)at only one particular point of the field of view.
B)over a particular field of view.
C)at only one particular point of the field of view and then squaring the number.
D)over a particular field of view and then squaring the number.
Question
Of the following procedures,which involve extended fluoroscopic time?
1)Biliary drainage
2)Stent and filter placement
3)Urinary or biliary stone removal

A)1 and 2 only
B)1 and 3 only
C)2 and 3 only
D)1,2,and 3
Question
Which of the following materials is commonly used in the tabletop of a radiographic examination table to make the tabletop as radiolucent as possible so that it will absorb only a minimal amount of radiation,thereby reducing the patient's radiation dose?

A)Carbon fiber material
B)Formica
C)Granite
D)Slate
Question
The control panel,where technical exposure factors are selected and seen on indicators by the equipment operator,must be located:

A)at the foot end of the radiographic examining table.
B)at the head end of the radiographic examining table.
C)behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure.
D)behind any regular wall that has a regular glass window that permits observation of the patient during any procedure.
Question
When settings are changed from one mA to a neighboring mA station,the most that linearity can vary is:

A)0%.
B)10%.
C)20%.
D)50%.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/25
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 11: Equipment Design for Radiation Protection
1
Digital radiography images can be accessed:

A)at only one workstation at a time,thereby decreasing convenience for physicians providing patient care.
B)at no more than two workstations at a time,thereby decreasing convenience for physicians providing patient care.
C)at several workstations at the same time,making image viewing very convenient for physicians providing patient care.
D)digital images cannot be accessed at any workstations by physicians.
at several workstations at the same time,making image viewing very convenient for physicians providing patient care.
2
Depending on the area of the body being examined with a fluoroscopic image intensification system,for adult patients,a range of ______ kVp is generally used.

A)30 to 60
B)60 to 75
C)75 to 110
D)110 to 140
75 to 110
3
Compared with the traditional material in the front of a cassette (aluminum or cardboard),the cassette front containing the carbon fiber absorbs approximately ______ as much radiation,resulting in a __________ dose for the patient because ________ radiographic techniques are required to produce the recorded image.

A)four times,higher,higher
B)three times,higher,higher
C)twice,higher,higher
D)half,lower,lower
half,lower,lower
4
Federal government specifications recommend a minimum total filtration of _______________ for stationary (fixed) fluoroscopic x-ray units operating above 70 kVp.

A)1.0-mm aluminum equivalent
B)1.5-mm aluminum equivalent
C)2.0-mm aluminum equivalent
D)2.5-mm aluminum equivalent
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
5
During a fluoroscopic examination,a resettable cumulative timing device times the x-ray beam-on time and sounds an audible alarm or temporarily interrupts the exposure after the fluoroscope has been activated for what length of time?

A)1 minute
B)3 minutes
C)5 minutes
D)10 minutes
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
6
Computed radiography,because of its higher exposure latitude,makes grid use on the pediatric population:

A)absolutely necessary for all radiographic examinations.
B)less necessary than was previously believed.
C)necessary only for extremity examinations.
D)unnecessary for any examination.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
7
Sharper size restriction of the radiographic beam is achieved when a cone or cylinder is:

A)longer.
B)shorter.
C)made of aluminum.
D)made of durable plastic.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
8
When obtaining a dorsoplantar projection of a foot,which of the following types of filters should be used to provide a greater uniform density of the anatomy?

A)Bilateral wedge filter
B)Trough filter
C)Thoraeus filter
D)Wedge filter
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
9
Current federal standards limit entrance skin exposure rates of general-purpose intensified fluoroscopic units to a maximum of ______________ measured at tabletop with the image intensifier entrance surface at a prescribed 30 cm above.

A)100 mGya/min
B)300 mGya/min
C)500 mGya/min
D)600 mGya/min
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
10
When the speed of screen-film systems doubles,for example,when changing from a 200-speed to a 400-speed system,patient radiation exposure is ______ by approximately ______.

A)increased,25%
B)increased,50%
C)reduced,25%
D)reduced,50%
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
11
CR imaging has greater ____________ flexibility than conventional screen-film radiography.

A)kilovoltage
B)milliamperage
C)time of exposure
D)x-ray beam collimation
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
12
In standard image intensification fluoroscopy,an x-ray beam half-value-layer (HVL) of 3 to 4.5 mm aluminum is considered acceptable when peak kilovoltage ranges from:

A)50 to 60.
B)60 to 70.
C)70 to 80.
D)80 to 100.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
13
When an exposed computed radiography imaging plate is ready to be processed,an imaging reading unit is used to scan the photostimulable phosphor imaging plate with a helium-neon laser beam.This process results in the emission of violet light that is changed into an electronic signal by a device called a:

A)charge-coupled-device array.
B)photodiode.
C)photomultiplier tube.
D)scintillator.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
14
A radiographer uses a high-speed,rare-earth screen-film system with optimal technical exposure factors to obtain PA and lateral radiographs of a 2-year-old child's chest.To maximize reduction of radiation exposure to the patient following adequate immobilization,the radiographer should:

A)collimate the x-ray beam so that it is only slightly larger than the margins of the image receptor and shield the child's reproductive organs.
B)collimate the x-ray beam so that it is no larger than the margins of the image receptor and shield the child's reproductive organs.
C)collimate the x-ray beam so that it is smaller than the margins of the image receptor and shield the child's reproductive organs.
D)collimate the x-ray beam so that it is smaller than the margins of the image receptor and not shield the child's reproductive organs.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
15
What is the purpose of radiographic beam filtration?

A)To decrease beam hardness,thereby reducing patient skin dose and the dose to superficial tissues
B)To increase beam hardness,thereby reducing patient skin dose and the dose to superficial tissues
C)To eliminate short wavelength radiation to reduce the radiation received by the patient's skin and superficial tissues
D)To increase beam hardness,thereby increasing patient skin dose with only a slight increase to superficial tissues
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
16
Which of the following is not an x-ray beam limitation device?

A)Aperture diaphragm
B)Collimator
C)Filter
D)Extension cylinder
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
17
When using digital fluoroscopy systems,making use of the last-image-hold feature can:

A)be an effective dose reduction technique.
B)eliminate the need to collimate the primary beam.
C)eliminate increased involuntary patient motion.
D)eliminate increased voluntary patient motion.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
18
When a fluoroscopic image is electronically amplified by the use of an image intensification system,which of the following benefits results?
1)Increased image brightness
2)Saving of time for the radiologist
3)Patient dose reduction

A)1 and 2 only
B)1 and 3 only
C)2 and 3 only
D)1,2,and 3
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
19
When performing a mobile fluoroscopic procedure,to reduce the radiation exposure to the patient,the radiographer must use a minimal source-skin distance of:

A)15 cm.
B)30 cm.
C)38 cm.
D)45 cm.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
20
To visualize smaller and lower contrast objects during interventional procedures,high-level-control fluoroscopy uses exposure rates that are ____________ those normally used in routine fluoroscopy.

A)higher than
B)lower than
C)the same as
D)none of the above because smaller and lower contrast objects cannot be visualized by fluoroscopy during interventional procedures
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
21
Luminance is determined by measuring the concentration of light:

A)at only one particular point of the field of view.
B)over a particular field of view.
C)at only one particular point of the field of view and then squaring the number.
D)over a particular field of view and then squaring the number.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
22
Of the following procedures,which involve extended fluoroscopic time?
1)Biliary drainage
2)Stent and filter placement
3)Urinary or biliary stone removal

A)1 and 2 only
B)1 and 3 only
C)2 and 3 only
D)1,2,and 3
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
23
Which of the following materials is commonly used in the tabletop of a radiographic examination table to make the tabletop as radiolucent as possible so that it will absorb only a minimal amount of radiation,thereby reducing the patient's radiation dose?

A)Carbon fiber material
B)Formica
C)Granite
D)Slate
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
24
The control panel,where technical exposure factors are selected and seen on indicators by the equipment operator,must be located:

A)at the foot end of the radiographic examining table.
B)at the head end of the radiographic examining table.
C)behind a suitable protective barrier that has a radiation-absorbent window that permits observation of the patient during any procedure.
D)behind any regular wall that has a regular glass window that permits observation of the patient during any procedure.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
25
When settings are changed from one mA to a neighboring mA station,the most that linearity can vary is:

A)0%.
B)10%.
C)20%.
D)50%.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 25 flashcards in this deck.