Deck 13: Minimizing Patient Dose

Full screen (f)
exit full mode
Question
The approximate entrance skin exposure (ESE) for a radiographic study taken at 90 kVp and 23 mAs is

A)57.5 mR.
B)67 mR.
C)103.5 mR.
D)2.7 R.
Use Space or
up arrow
down arrow
to flip the card.
Question
An effective method of reducing patient exposure is to use

A)a slow speed image receptor.
B)a medium speed image receptor.
C)a grid.
D)the highest speed possible image receptor.
Question
In an effort to reduce dose, the recommended SID for routine radiography is

A)36-38 inches.
B)40-42 inches.
C)44-48 inches.
D)56 inches.
Question
When the primary-beam field size ____, the patient exposure ____.

A)increases; decreases
B)decreases; decreases
C)decreases; increases
D)increases; increases
Question
In an effort to minimize patient dose,

A)the small focal spot should be used.
B)the large focal spot should be used.
C)beam filtration should be minimum.
D)none of the above.
Question
The use of a grid will

A)decrease patient exposure.
B)increase patient exposure.
C)have no effect on patient exposure.
D)increase kVp.
Question
When patient SID or SOD increases,

A)patient ESE decreases.
B)patient ESE increases.
C)patient ESE increases by the square of the distance.
D)there is no relationship between ESE and distance.
Question
When comparing ESE and actual dose to an area of interest,

A)actual dose is higher than ESE.
B)ESE and actual are equal.
C)the actual dose and ESE are inverse.
D)actual dose is lower than ESE.
Question
Entrance skin exposure is a/an ____ exposure.

A)maximum
B)minimum
C)average
D)theoretical
Question
To reduce patient exposure, the __________ consistent with image quality should be selected.

A)highest mAs
B)highest kVp
C)shortest distance
D)highest mAs and lowest kVp
Question
The radiation intensity from a diagnostic x-ray unit will vary in a/an

A)direct relationship with mAs.
B)direct relationship with the square of the kVp.
C)inverse relationship with the square of the distance.
D)all of the above
Question
The Food and Drug Administration (FDA) sets an entrance skin exposure limit of ____ R/min for fluoroscopic systems in the United States.

A)11.5
B)20.2
C)35.1
D)42.7
Question
The conscientious radiographer can reduce the patient dose by at least ____ in most examinations by choosing appropriate exposure factors.

A)100 percent
B)75 percent
C)50 percent
D)25 percent
Question
Skull radiography routinely includes a posterior-anterior (PA) projection.This is intended to

A)reduce dose to the lens of the eye.
B)reduce dose to the cerebellum.
C)reduce dose to the paranasal sinuses.
D)reduce dose to the thyroid.
Question
Which of the following radiographic procedures delivers the lowest entrance skin exposure?

A)hip
B)pelvis
C)lumbar spine
D)elbow
Question
Each of the following is a type of gonad shield EXCEPT

A)flat contact.
B)shadow.
C)collimator.
D)shaped contact.
Question
Patient entrance skin exposure decreases as

A)source-to-image receptor distance (SID) decreases.
B)SID increases.
C)source-to-object distance (SOD) decreases.
D)object-to-image receptor distance (OID) increases.
Question
Of the choices below, the combination of exposure factors that would minimize patient dose would be

A)24 mAs, 80 kVp, 8"x10" field size, 44" SID
B)12 mAs, 92 kVp, 8"x10" field size, 40" SID
C)6 mAs, 106 kVp, 8"x10" field size, 44" SID
D)24 mAs, 80 kVp, 10"x12" field size, 40" SID
Question
The maximum exposure received by the patient is at

A)the skin exit to the body.
B)the area of interest.
C)the gonads.
D)the skin entrance to the body.
Question
Which x-ray generator will substantially decrease the patient entrance skin exposure?

A)single-phase, unrectified
B)single phase, rectified
C)three-phase, 6-pulse
D)three-phase, 12-pulse
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
1.7 mR/mAs

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Question
A radiographic examination of the small bowel for a 25-cm abdomen was performed doing a 15-min PA; 30-min PA; 60-min PA; and 90-min PA radiograph.What is the total approximate entrance skin exposure if each radiograph was produced at 100 kVp, 25 mA, and 40 in (100 cm)? According to calculations, 100 kVp = 5.6 mR/mA.
Question
What is the approximate entrance skin exposure in mR for the fluoroscopic examination of the stomach performed for 4.5 minutes at 110 kVp at 1.7 R/min?
Question
What is the approximate entrance skin exposure for a 15-cm AP abdomen produced at 70 kVp, 20 mA, and a SID of 40 in (100 cm), if 70 kVp = 2.5 mR/mA?
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
has no impact upon patient dose

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
male gonadal shielding

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
dose relationship is inverse.

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
lowers dose to the kidneys

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Question
What is the approximate entrance skin exposure for a 20-cm AP pelvis produced at 80 kVp, 20 mA, and a SID of 40 in (100 cm), if 70 kVp = 2.5 mR/mA?
Question
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
45 mR ESE

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/30
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 13: Minimizing Patient Dose
1
The approximate entrance skin exposure (ESE) for a radiographic study taken at 90 kVp and 23 mAs is

A)57.5 mR.
B)67 mR.
C)103.5 mR.
D)2.7 R.
103.5 mR.
2
An effective method of reducing patient exposure is to use

A)a slow speed image receptor.
B)a medium speed image receptor.
C)a grid.
D)the highest speed possible image receptor.
the highest speed possible image receptor.
3
In an effort to reduce dose, the recommended SID for routine radiography is

A)36-38 inches.
B)40-42 inches.
C)44-48 inches.
D)56 inches.
44-48 inches.
4
When the primary-beam field size ____, the patient exposure ____.

A)increases; decreases
B)decreases; decreases
C)decreases; increases
D)increases; increases
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
5
In an effort to minimize patient dose,

A)the small focal spot should be used.
B)the large focal spot should be used.
C)beam filtration should be minimum.
D)none of the above.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
6
The use of a grid will

A)decrease patient exposure.
B)increase patient exposure.
C)have no effect on patient exposure.
D)increase kVp.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
7
When patient SID or SOD increases,

A)patient ESE decreases.
B)patient ESE increases.
C)patient ESE increases by the square of the distance.
D)there is no relationship between ESE and distance.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
8
When comparing ESE and actual dose to an area of interest,

A)actual dose is higher than ESE.
B)ESE and actual are equal.
C)the actual dose and ESE are inverse.
D)actual dose is lower than ESE.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
9
Entrance skin exposure is a/an ____ exposure.

A)maximum
B)minimum
C)average
D)theoretical
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
10
To reduce patient exposure, the __________ consistent with image quality should be selected.

A)highest mAs
B)highest kVp
C)shortest distance
D)highest mAs and lowest kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
11
The radiation intensity from a diagnostic x-ray unit will vary in a/an

A)direct relationship with mAs.
B)direct relationship with the square of the kVp.
C)inverse relationship with the square of the distance.
D)all of the above
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
12
The Food and Drug Administration (FDA) sets an entrance skin exposure limit of ____ R/min for fluoroscopic systems in the United States.

A)11.5
B)20.2
C)35.1
D)42.7
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
13
The conscientious radiographer can reduce the patient dose by at least ____ in most examinations by choosing appropriate exposure factors.

A)100 percent
B)75 percent
C)50 percent
D)25 percent
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
14
Skull radiography routinely includes a posterior-anterior (PA) projection.This is intended to

A)reduce dose to the lens of the eye.
B)reduce dose to the cerebellum.
C)reduce dose to the paranasal sinuses.
D)reduce dose to the thyroid.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
15
Which of the following radiographic procedures delivers the lowest entrance skin exposure?

A)hip
B)pelvis
C)lumbar spine
D)elbow
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
16
Each of the following is a type of gonad shield EXCEPT

A)flat contact.
B)shadow.
C)collimator.
D)shaped contact.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
17
Patient entrance skin exposure decreases as

A)source-to-image receptor distance (SID) decreases.
B)SID increases.
C)source-to-object distance (SOD) decreases.
D)object-to-image receptor distance (OID) increases.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
18
Of the choices below, the combination of exposure factors that would minimize patient dose would be

A)24 mAs, 80 kVp, 8"x10" field size, 44" SID
B)12 mAs, 92 kVp, 8"x10" field size, 40" SID
C)6 mAs, 106 kVp, 8"x10" field size, 44" SID
D)24 mAs, 80 kVp, 10"x12" field size, 40" SID
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
19
The maximum exposure received by the patient is at

A)the skin exit to the body.
B)the area of interest.
C)the gonads.
D)the skin entrance to the body.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
20
Which x-ray generator will substantially decrease the patient entrance skin exposure?

A)single-phase, unrectified
B)single phase, rectified
C)three-phase, 6-pulse
D)three-phase, 12-pulse
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
21
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
1.7 mR/mAs

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
22
A radiographic examination of the small bowel for a 25-cm abdomen was performed doing a 15-min PA; 30-min PA; 60-min PA; and 90-min PA radiograph.What is the total approximate entrance skin exposure if each radiograph was produced at 100 kVp, 25 mA, and 40 in (100 cm)? According to calculations, 100 kVp = 5.6 mR/mA.
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
23
What is the approximate entrance skin exposure in mR for the fluoroscopic examination of the stomach performed for 4.5 minutes at 110 kVp at 1.7 R/min?
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
24
What is the approximate entrance skin exposure for a 15-cm AP abdomen produced at 70 kVp, 20 mA, and a SID of 40 in (100 cm), if 70 kVp = 2.5 mR/mA?
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
25
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
has no impact upon patient dose

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
26
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
male gonadal shielding

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
27
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
dose relationship is inverse.

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
28
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
lowers dose to the kidneys

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
29
What is the approximate entrance skin exposure for a 20-cm AP pelvis produced at 80 kVp, 20 mA, and a SID of 40 in (100 cm), if 70 kVp = 2.5 mR/mA?
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
30
Match the following choices with the correct answer as each relates to minimizing patient radiation dose.
45 mR ESE

A)60 kVp
B)shaped contact
C)focal spot size
D)distance (SID)
E)44-48 inches
F)AP supine abdomen position
G)oblique projection of the abdomen
H)18 mAs @ 70 kVp
Unlock Deck
Unlock for access to all 30 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 30 flashcards in this deck.