Deck 20: Quality Control for Computed Tomography Scanners

Full screen (f)
exit full mode
Question
Often a quality control program can result in reduced down time.
Use Space or
up arrow
down arrow
to flip the card.
Question
All of the following are possible causes of failure for the standard deviation of the computed tomography number except:

A)electronic noise from amplifiers
B)decreased dose
C)electronic noise from the analog-to-digital converters
D)miscalibration of the algorithm
Question
The following quality control tests should be done annually except:

A)computed tomography number versus patient size
B)hard copy output
C)computed tomography number versus slice width
D)standard deviation of computed tomography number in water
Question
Quality control tests that are effective must be objective, quantitative, and easy and can be done very quickly.
Question
Quality control test results do not have to be interpreted as soon as the tests are completed.
Question
In quality control, steps are taken to correct problems.
Question
Record keeping is not an important consideration in a quality control program.
Question
The tolerance limit for a computed tomography number calibration quality control test is:

A)±3 computed tomography numbers
B)±5 computed tomography numbers
C)±8 computed tomography numbers
D)no tolerance limit has been established
Question
Quality control tests can be performed by all of the following except:

A)technologists
B)medical physicists
C)quality control technicians
D)department aides
Question
Which tests the performance of a computed tomography scanner on a number of parameters and compares the performance with a standard?

A)quality administration
B)quality assurance
C)quality control
D)quality maintenance
Question
Computed tomography scanner manufacturers do not offer any information regarding quality control testing.
Question
The average CT number of water QC test should be done:

A)daily
B)weekly
C)monthly
D)annually
Question
All quality control tests should be conducted on a monthly basis.
Question
Quality control tests, test results, and interpretation of test results should be:

A)recorded in a quality control logbook
B)recorded on pieces of paper
C)recorded on the bulletin board beside the computed tomography scanner
D)recorded in the positioning notebook of the technologist
Question
Water phantoms are not used for quality control tests.
Question
Regulatory agencies do not require any standard of quality.
Question
The following are all basic tenets of a quality control program for computed tomography scanners except:

A)tests should be performed on a regular period basis
B)prompt interpretation of the results
C)faithful record keeping
D)phantoms for quality control measurements must be stored in a controlled environment
Question
Quality control tests must be performed on a routine basis.
Question
A measurement of a computed tomography scanner's performance to ensure the scanner is operating at an acceptable level describes:

A)quality administration
B)quality assurance
C)quality control
D)quality maintenance
Question
A possible cause of failure for the high-contrast resolution quality control test is that of an enlarged focal spot in the x-ray tube.
Question
The following quality control tests should be done biannually except:

A)high-contrast resolution
B)low-contrast resolution
C)light field accuracy
D)flatness of computed tomography number
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/21
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 20: Quality Control for Computed Tomography Scanners
1
Often a quality control program can result in reduced down time.
True
2
All of the following are possible causes of failure for the standard deviation of the computed tomography number except:

A)electronic noise from amplifiers
B)decreased dose
C)electronic noise from the analog-to-digital converters
D)miscalibration of the algorithm
miscalibration of the algorithm
3
The following quality control tests should be done annually except:

A)computed tomography number versus patient size
B)hard copy output
C)computed tomography number versus slice width
D)standard deviation of computed tomography number in water
standard deviation of computed tomography number in water
4
Quality control tests that are effective must be objective, quantitative, and easy and can be done very quickly.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
5
Quality control test results do not have to be interpreted as soon as the tests are completed.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
6
In quality control, steps are taken to correct problems.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
7
Record keeping is not an important consideration in a quality control program.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
8
The tolerance limit for a computed tomography number calibration quality control test is:

A)±3 computed tomography numbers
B)±5 computed tomography numbers
C)±8 computed tomography numbers
D)no tolerance limit has been established
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
9
Quality control tests can be performed by all of the following except:

A)technologists
B)medical physicists
C)quality control technicians
D)department aides
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
10
Which tests the performance of a computed tomography scanner on a number of parameters and compares the performance with a standard?

A)quality administration
B)quality assurance
C)quality control
D)quality maintenance
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
11
Computed tomography scanner manufacturers do not offer any information regarding quality control testing.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
12
The average CT number of water QC test should be done:

A)daily
B)weekly
C)monthly
D)annually
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
13
All quality control tests should be conducted on a monthly basis.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
14
Quality control tests, test results, and interpretation of test results should be:

A)recorded in a quality control logbook
B)recorded on pieces of paper
C)recorded on the bulletin board beside the computed tomography scanner
D)recorded in the positioning notebook of the technologist
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
15
Water phantoms are not used for quality control tests.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
16
Regulatory agencies do not require any standard of quality.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
17
The following are all basic tenets of a quality control program for computed tomography scanners except:

A)tests should be performed on a regular period basis
B)prompt interpretation of the results
C)faithful record keeping
D)phantoms for quality control measurements must be stored in a controlled environment
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
18
Quality control tests must be performed on a routine basis.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
19
A measurement of a computed tomography scanner's performance to ensure the scanner is operating at an acceptable level describes:

A)quality administration
B)quality assurance
C)quality control
D)quality maintenance
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
20
A possible cause of failure for the high-contrast resolution quality control test is that of an enlarged focal spot in the x-ray tube.
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
21
The following quality control tests should be done biannually except:

A)high-contrast resolution
B)low-contrast resolution
C)light field accuracy
D)flatness of computed tomography number
Unlock Deck
Unlock for access to all 21 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 21 flashcards in this deck.