Deck 10: Dynamic Imaging: Fluoroscopy

Full screen (f)
exit full mode
Question
___________ imaging is accomplished with fluoroscopy as opposed to radiography.

A) Static
B) Still
C) Dynamic
D) Colorized
Use Space or
up arrow
down arrow
to flip the card.
Question
If a 30/23/15 cm image intensifier is operated in the 23 cm mode, the fluoroscopic image will be magnified by a factor of:

A) 1.3
B) 1.5
C) 1.7
D) 2
Question
The __________________________ absorbs x-rays and emits light.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Accelerating anode
Question
Which of the following most accurately shows the energy sequence during image-intensified fluoroscopy?

A) Light to x-rays to electrons to light
B) X-rays to light to electrons to light
C) X-rays to electrons to light
D) Electrons to light to electrons
Question
The __________________ absorbs electrons and emits light.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Accelerating anode
Question
Magnification of the fluoroscopic image results in improved:

A) Brightness
B) Spatial resolution
C) Contrast
D) Exposure
Question
The formula to determine the amount of magnification created when in magnification mode is:

A) MF = full size input phosphor / selected input phosphor
B) MF = full size input phosphor2 / selected input phosphor2
C) MF = selected input phosphor / full size input phosphor
D) MF = selected input phosphor2 / full size input phosphor2
Question
Distortion of the fluoroscopic image that appears as unequal magnification is:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Question
Typical fluoroscopic systems have spatial resolution capabilities in the range of:

A) 2 to 3 Lp/mm
B) 4 to 6 Lp/mm
C) 7 to 9 Lp/mm
D) 10 to 12 Lp/mm
Question
This part of the image intensifier absorbs light and produces electrons.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Electrostatic focusing lenses
Question
When operated in magnification mode, the electrostatic lenses have ________ of a negative charge.

A) Less
B) More
C) The same amount as usual
D) One half
Question
In reference to image intensification, the function that adjusts and maintains the overall image brightness and contrast during the fluoroscopic procedure is:

A) DQE
B) APR
C) ABC
D) AEC
Question
The gain related to the decrease in size from the input to the output phosphor is the:

A) Brightness gain
B) Flux
C) Minification gain
D) Conversion factor
Question
The formula for brightness gain is:

A) Brightness gain = minification gain - flux gain
B) Brightness gain = minification gain / flux gain
C) Brightness gain = minification gain ×\times flux gain
D) None of the above
Question
The conversion factor is a(n)___________ term for brightness gain and is typically about __________ of the brightness gain value.

A) Older; 1%
B) Older; 10%
C) Newer; 1%
D) Newer; 10%
Question
In fluoroscopy, shape distortion is caused by:

A) Angling the tube
B) Angling the image intensifier
C) The curved shape of the photocathode
D) The position of the electrostatic focusing lenses
Question
A loss of brightness around the edge of the fluoroscopic image due to the curve of the photocathode is:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Question
The input phosphor is ______________ than the output phosphor.

A) Larger
B) Smaller
C) The same size as
D) Brighter
Question
This is measured in cd/m2/mR/s.

A) Brightness gain
B) Flux
C) Minification gain
D) Conversion factor
Question
A disadvantage of using magnification mode during fluoroscopy is:

A) The image has less resolution.
B) The image brightness is inconsistent.
C) The patient receives additional dose.
D) None of the above
Question
A typical high resolution monitor will have:

A) 825 lines
B) 1024 lines
C) 1872 lines
D) 2034 lines
Question
The purpose of a beam splitting mirror is to:

A) Be able to view the fluoroscopic image on more than one television
B) Reduce the intensity of the image from the output phosphor
C) Allow spot filming to be done during fluoroscopy
D) None of the above
Question
The recording device that "photographs" the image off of the output phosphor is the:

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Question
The electron gun in the monitor moves in the same ___________ pattern as did the electron gun in the tube camera.

A) Sweep
B) Brush
C) Raster
D) Rooster
Question
Vidicon refers to a type of:

A) Television monitor
B) Camera tube
C) CCD
D) Fiber optics
Question
Typical television monitors have:

A) 300 lines
B) 450 lines
C) 525 lines
D) 875 lines
Question
This recording system uses 70 mm or 105 mm film.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Question
This recording system allows the film to be divided into two, four, or more images.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Question
Which of the following require the x-ray beam to be in radiographic mode?

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Question
Early versions of digital fluoroscopy used the standard system and added:

A) A different camera
B) An ADC
C) Videotape
D) None of the above
Question
Increasing the mA is the way to correct a fluoroscopic image that has:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Question
In order to record the dynamic image you will need to use:

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Question
The conventional or current method for viewing the fluoroscopic image is by:

A) Using a hand-held device
B) Looking directly at the output phosphor
C) Using a TV monitor
D) Using a mirror-optics system
Question
The fiberoptic bundle or optical lens system is used to:

A) Deliver the image from the camera to the television monitor
B) Record the fluoroscopic image
C) Link the output phosphor and camera or CCD
D) None of the above
Question
In terms of resolution, the weakest part of the fluoroscopic system is the:

A) Image intensifier
B) Television monitor
C) Camera tube
D) None of the above
Question
The CCD's electrical charge from the capacitors is sent to the:

A) Output phosphor
B) Television monitor
C) X-ray tube
D) Image intensifier
Question
When the electron beam in the camera travels back and forth across, from the top to bottom, it is moving in a ___________ pattern.

A) Sweep
B) Brush
C) Raster
D) Rooster
Question
Semiconductor capacitors are components of the:

A) Camera tube
B) CCD
C) Image intensifier
D) X-ray tube
Question
The image intensifier is capable of resolving about ___________, while the monitor can resolve about _______________.

A) 1 to 2 Lp/mm; 1 to 2 Lp/mm
B) 1 to 2 Lp/mm; 5 Lp/mm
C) 5 Lp/mm; 5 Lp/mm
D) 5 Lp/mm; 1 to 2 Lp/mm
Question
Each capacitor in a CCD represents a(n):

A) Line
B) Image
C) Pixel
D) Electric charge
Question
Fluoroscopic equipment should be inspected every:

A) Week
B) Month
C) 6 months
D) Year
Question
Which of the following produces an image with more noise?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Question
Which of the following requires more radiation to produce quality fluoroscopic images?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Question
The newest fluoroscopy systems use a flat-panel detector:

A) To display the spot images during fluoroscopy
B) To record the overhead images following fluoroscopy
C) In place of the image intensifier
D) In addition to the image intensifier
Question
Which of the following has a rectangular field of view and a wider dynamic range?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Question
Digital fluoroscopy is improved by using:

A) A vidicon camera
B) Videotape
C) A beam splitting mirror
D) A CCD
Question
Flat-panel detectors for fluoroscopic imaging can be the:

A) Amorphous silicon indirect capture type
B) Photostimulable phosphor plate type
C) Amorphous selenium direct capture type
D) A and C
Question
The analog to digital converter:

A) Makes the electrical signal understandable to the computer
B) Determines the contrast resolution of the system
C) Determines the image matrix
D) All of the above
Question
The radiographer may be responsible for the ________________ inspection of the fluoroscopic equipment for QC purposes.

A) Performance
B) Operational
C) A and B
D) None of the above
Question
Which of the following does not require any additional devices for spot filming during the fluoroscopic procedure?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Question
Use of which of the following systems will result in less radiation dose to the patient?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Question
The x-ray beam used with a digital fluoroscopic unit uses ___________ mA and a ______________ beam.

A) Low; constant
B) High; constant
C) Low; pulsed
D) High; pulsed
Question
As compared to the tube-type camera, the CCD:

A) Is smaller
B) Is larger
C) Uses an electron gun to produce the video signal
D) B and C
Question
Which of the following is more light-sensitive?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Question
The SSD should be no less than ___________ for a fixed or stationary fluoro unit and no less than ____________ for a mobile fluoro unit.

A) 38 cm; 30 cm
B) 30 cm; 38 cm
C) 38 inches; 30 inches
D) 30 inches; 38 inches
Question
Inspecting the protective curtain, Bucky slot cover, and fluoroscopic timer are part of the:

A) Performance QC testing
B) Operational QC testing
C) A and B
D) None of the above
Question
Which of the following has the possibility of demonstrating veiling glare and pincushion distortion?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Question
In comparison, the vidicon camera tube has ______________ spatial resolution, and the CCD requires ________________ radiation to produce a good fluoroscopic image.

A) Less; more
B) Less; less
C) Better; more
D) Better; less
Question
Which of the following has more spatial distortion?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Question
Which of the following is easier to move around and allows better access to the patient during a procedure?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Question
The beam splitting mirror is never used with fiber optics.
Question
The purpose of a camera tube or CCD is to record the fluoroscopic image.
Question
The electron gun in the television monitor sends out a steady stream of electrons to activate the fluorescent screen.
Question
When using magnification mode, the fluoroscopist needs to manually set the exposure factors higher to compensate for the reduced diameter input phosphor.
Question
Each electron that interacts with the output phosphor will result in significantly fewer light photons being produced in comparison to the number of light photons it took to release that electron.
Question
When a CCD is exposed to light, it generates and briefly stores light energy.
Question
The ABC can be slow in responding, seen as a brief delay in the brightness adjustment.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/67
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 10: Dynamic Imaging: Fluoroscopy
1
___________ imaging is accomplished with fluoroscopy as opposed to radiography.

A) Static
B) Still
C) Dynamic
D) Colorized
Dynamic
2
If a 30/23/15 cm image intensifier is operated in the 23 cm mode, the fluoroscopic image will be magnified by a factor of:

A) 1.3
B) 1.5
C) 1.7
D) 2
1.3
3
The __________________________ absorbs x-rays and emits light.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Accelerating anode
Input phosphor
4
Which of the following most accurately shows the energy sequence during image-intensified fluoroscopy?

A) Light to x-rays to electrons to light
B) X-rays to light to electrons to light
C) X-rays to electrons to light
D) Electrons to light to electrons
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
5
The __________________ absorbs electrons and emits light.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Accelerating anode
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
6
Magnification of the fluoroscopic image results in improved:

A) Brightness
B) Spatial resolution
C) Contrast
D) Exposure
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
7
The formula to determine the amount of magnification created when in magnification mode is:

A) MF = full size input phosphor / selected input phosphor
B) MF = full size input phosphor2 / selected input phosphor2
C) MF = selected input phosphor / full size input phosphor
D) MF = selected input phosphor2 / full size input phosphor2
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
8
Distortion of the fluoroscopic image that appears as unequal magnification is:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
9
Typical fluoroscopic systems have spatial resolution capabilities in the range of:

A) 2 to 3 Lp/mm
B) 4 to 6 Lp/mm
C) 7 to 9 Lp/mm
D) 10 to 12 Lp/mm
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
10
This part of the image intensifier absorbs light and produces electrons.

A) Input phosphor
B) Output phosphor
C) Photocathode
D) Electrostatic focusing lenses
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
11
When operated in magnification mode, the electrostatic lenses have ________ of a negative charge.

A) Less
B) More
C) The same amount as usual
D) One half
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
12
In reference to image intensification, the function that adjusts and maintains the overall image brightness and contrast during the fluoroscopic procedure is:

A) DQE
B) APR
C) ABC
D) AEC
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
13
The gain related to the decrease in size from the input to the output phosphor is the:

A) Brightness gain
B) Flux
C) Minification gain
D) Conversion factor
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
14
The formula for brightness gain is:

A) Brightness gain = minification gain - flux gain
B) Brightness gain = minification gain / flux gain
C) Brightness gain = minification gain ×\times flux gain
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
15
The conversion factor is a(n)___________ term for brightness gain and is typically about __________ of the brightness gain value.

A) Older; 1%
B) Older; 10%
C) Newer; 1%
D) Newer; 10%
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
16
In fluoroscopy, shape distortion is caused by:

A) Angling the tube
B) Angling the image intensifier
C) The curved shape of the photocathode
D) The position of the electrostatic focusing lenses
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
17
A loss of brightness around the edge of the fluoroscopic image due to the curve of the photocathode is:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
18
The input phosphor is ______________ than the output phosphor.

A) Larger
B) Smaller
C) The same size as
D) Brighter
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
19
This is measured in cd/m2/mR/s.

A) Brightness gain
B) Flux
C) Minification gain
D) Conversion factor
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
20
A disadvantage of using magnification mode during fluoroscopy is:

A) The image has less resolution.
B) The image brightness is inconsistent.
C) The patient receives additional dose.
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
21
A typical high resolution monitor will have:

A) 825 lines
B) 1024 lines
C) 1872 lines
D) 2034 lines
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
22
The purpose of a beam splitting mirror is to:

A) Be able to view the fluoroscopic image on more than one television
B) Reduce the intensity of the image from the output phosphor
C) Allow spot filming to be done during fluoroscopy
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
23
The recording device that "photographs" the image off of the output phosphor is the:

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
24
The electron gun in the monitor moves in the same ___________ pattern as did the electron gun in the tube camera.

A) Sweep
B) Brush
C) Raster
D) Rooster
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
25
Vidicon refers to a type of:

A) Television monitor
B) Camera tube
C) CCD
D) Fiber optics
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
26
Typical television monitors have:

A) 300 lines
B) 450 lines
C) 525 lines
D) 875 lines
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
27
This recording system uses 70 mm or 105 mm film.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
28
This recording system allows the film to be divided into two, four, or more images.

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
29
Which of the following require the x-ray beam to be in radiographic mode?

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
30
Early versions of digital fluoroscopy used the standard system and added:

A) A different camera
B) An ADC
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
31
Increasing the mA is the way to correct a fluoroscopic image that has:

A) Noise
B) Pincushion appearance
C) Vignetting
D) Magnification
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
32
In order to record the dynamic image you will need to use:

A) Cassette spot filming
B) Film cameras for spot filming
C) Videotape
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
33
The conventional or current method for viewing the fluoroscopic image is by:

A) Using a hand-held device
B) Looking directly at the output phosphor
C) Using a TV monitor
D) Using a mirror-optics system
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
34
The fiberoptic bundle or optical lens system is used to:

A) Deliver the image from the camera to the television monitor
B) Record the fluoroscopic image
C) Link the output phosphor and camera or CCD
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
35
In terms of resolution, the weakest part of the fluoroscopic system is the:

A) Image intensifier
B) Television monitor
C) Camera tube
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
36
The CCD's electrical charge from the capacitors is sent to the:

A) Output phosphor
B) Television monitor
C) X-ray tube
D) Image intensifier
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
37
When the electron beam in the camera travels back and forth across, from the top to bottom, it is moving in a ___________ pattern.

A) Sweep
B) Brush
C) Raster
D) Rooster
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
38
Semiconductor capacitors are components of the:

A) Camera tube
B) CCD
C) Image intensifier
D) X-ray tube
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
39
The image intensifier is capable of resolving about ___________, while the monitor can resolve about _______________.

A) 1 to 2 Lp/mm; 1 to 2 Lp/mm
B) 1 to 2 Lp/mm; 5 Lp/mm
C) 5 Lp/mm; 5 Lp/mm
D) 5 Lp/mm; 1 to 2 Lp/mm
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
40
Each capacitor in a CCD represents a(n):

A) Line
B) Image
C) Pixel
D) Electric charge
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
41
Fluoroscopic equipment should be inspected every:

A) Week
B) Month
C) 6 months
D) Year
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
42
Which of the following produces an image with more noise?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
43
Which of the following requires more radiation to produce quality fluoroscopic images?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
44
The newest fluoroscopy systems use a flat-panel detector:

A) To display the spot images during fluoroscopy
B) To record the overhead images following fluoroscopy
C) In place of the image intensifier
D) In addition to the image intensifier
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
45
Which of the following has a rectangular field of view and a wider dynamic range?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
46
Digital fluoroscopy is improved by using:

A) A vidicon camera
B) Videotape
C) A beam splitting mirror
D) A CCD
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
47
Flat-panel detectors for fluoroscopic imaging can be the:

A) Amorphous silicon indirect capture type
B) Photostimulable phosphor plate type
C) Amorphous selenium direct capture type
D) A and C
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
48
The analog to digital converter:

A) Makes the electrical signal understandable to the computer
B) Determines the contrast resolution of the system
C) Determines the image matrix
D) All of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
49
The radiographer may be responsible for the ________________ inspection of the fluoroscopic equipment for QC purposes.

A) Performance
B) Operational
C) A and B
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
50
Which of the following does not require any additional devices for spot filming during the fluoroscopic procedure?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
51
Use of which of the following systems will result in less radiation dose to the patient?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
52
The x-ray beam used with a digital fluoroscopic unit uses ___________ mA and a ______________ beam.

A) Low; constant
B) High; constant
C) Low; pulsed
D) High; pulsed
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
53
As compared to the tube-type camera, the CCD:

A) Is smaller
B) Is larger
C) Uses an electron gun to produce the video signal
D) B and C
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
54
Which of the following is more light-sensitive?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
55
The SSD should be no less than ___________ for a fixed or stationary fluoro unit and no less than ____________ for a mobile fluoro unit.

A) 38 cm; 30 cm
B) 30 cm; 38 cm
C) 38 inches; 30 inches
D) 30 inches; 38 inches
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
56
Inspecting the protective curtain, Bucky slot cover, and fluoroscopic timer are part of the:

A) Performance QC testing
B) Operational QC testing
C) A and B
D) None of the above
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
57
Which of the following has the possibility of demonstrating veiling glare and pincushion distortion?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
58
In comparison, the vidicon camera tube has ______________ spatial resolution, and the CCD requires ________________ radiation to produce a good fluoroscopic image.

A) Less; more
B) Less; less
C) Better; more
D) Better; less
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
59
Which of the following has more spatial distortion?

A) Tube-type camera
B) CCD
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
60
Which of the following is easier to move around and allows better access to the patient during a procedure?

A) Conventional image intensifier
B) Flat-panel detector in place of image intensifier
C) There is no difference between the two.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
61
The beam splitting mirror is never used with fiber optics.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
62
The purpose of a camera tube or CCD is to record the fluoroscopic image.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
63
The electron gun in the television monitor sends out a steady stream of electrons to activate the fluorescent screen.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
64
When using magnification mode, the fluoroscopist needs to manually set the exposure factors higher to compensate for the reduced diameter input phosphor.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
65
Each electron that interacts with the output phosphor will result in significantly fewer light photons being produced in comparison to the number of light photons it took to release that electron.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
66
When a CCD is exposed to light, it generates and briefly stores light energy.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
67
The ABC can be slow in responding, seen as a brief delay in the brightness adjustment.
Unlock Deck
Unlock for access to all 67 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 67 flashcards in this deck.