Deck 7: Quality Control of X-Ray Generators and Ancillary Radiographic Equipment

Full screen (f)
exit full mode
Question
The kilovolts peak (kVp) must be accurate to

A)±2%.
B)±5%.
C)±10%.
D)±15%.
Use Space or
up arrow
down arrow
to flip the card.
Question
What is the acceptance criterion for a timer circuit with an exposure time shorter than 10 ms?

A)5%
B)10%
C)15%
D)20%
Question
An exposure of 600 mA and 16 ms yields an exposure output of 240 mR.What is the mR/mAs?

A)2.5
B)25
C)250
D)2500
Question
Which of the following tests can be used to determine focal spot size?

A)Safelight test
B)Wire mesh test
C)Wisconsin test cassette
D)Star pattern test
Question
What is the NEMA allowance parameter for an x-ray tube that has a stated focal spot size of 1.5 mm?

A)50%
B)40%
C)30%
D)20%
Question
The light field and radiation field of a radiographic unit must be congruent to within ____ of the SID.

A)±1%
B)±2%
C)±3%
D)±5%
Question
The minimum filtration required for general-purpose radiographic systems is _____ mm Al.

A)2.0
B)2.1
C)2.5
D)3.0
Question
The x-ray tube must be mounted inside its housing so that the central ray exits the housing to within ____ degree(s) of perpendicular.

A)1
B)2
C)3
D)4
Question
During testing for grid uniformity, films should be uniform within an optical density of

A)±0.4.
B)±0.3.
C)±0.2.
D)±0.1.
Question
The tomographic exposure angle indicated on the equipment and the actual tomographic angle must coincide within ____ degrees for units operating at angles greater than 30 degrees.

A)2
B)5
C)10
D)20
Question
Conventional tomographic systems should produce images of tissue that lie within ____ mm of the indicated tomographic plane.

A)5
B)10
C)15
D)20
Question
Which of the following devices can be used as the sensor in AEC units?

A)Ion chamber
B)Photocell
C)Solid-state detector
D)All of the above
Question
What is the accepted half-value layer (HVL) for an x-ray unit operating at 80 kVp?

A)2.1 mm Al
B)2.3 mm Al
C)2.5 mm Al
D)2.7 mm Al
Question
X-ray system mAs reciprocity has an acceptance parameter of

A)±5%.
B)±10%.
C)±15%.
D)±20%.
Question
Path closure in pluridirectional tomographic units should be within ____ of the path length.

A)2%
B)5%
C)10%
D)15%
Question
Collimation systems equipped with positive beam limitation must be accurate to within ____ of the SID.

A)±5%
B)±4%
C)±3%
D)±2%
Question
Two common quality control tests that should be performed on grids are grid _____ and grid _____ tests.

A)ratio; frequency
B)alignment; uniformity
C)ratio; alignment
D)uniformity; ratio
Question
During AEC evaluation with varying kVp, the AEC system should maintain a constant film density to within ____ optical density units.

A)±0.05
B)±0.1
C)±0.2
D)±0.3
Question
The maximum variation coefficient for reproducibility of x-ray generators is

A)0.05.
B)0.07.
C)0.10.
D)0.20.
Question
A synchronous spinning top is used on a three-phase unit with a time of 17 ms.Approximately how large of an arc should appear if the timer circuit is accurate?

A)3 degrees
B)6 degrees
C)17 degrees
D)36 degrees
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/20
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 7: Quality Control of X-Ray Generators and Ancillary Radiographic Equipment
1
The kilovolts peak (kVp) must be accurate to

A)±2%.
B)±5%.
C)±10%.
D)±15%.
B
2
What is the acceptance criterion for a timer circuit with an exposure time shorter than 10 ms?

A)5%
B)10%
C)15%
D)20%
D
3
An exposure of 600 mA and 16 ms yields an exposure output of 240 mR.What is the mR/mAs?

A)2.5
B)25
C)250
D)2500
B
4
Which of the following tests can be used to determine focal spot size?

A)Safelight test
B)Wire mesh test
C)Wisconsin test cassette
D)Star pattern test
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
5
What is the NEMA allowance parameter for an x-ray tube that has a stated focal spot size of 1.5 mm?

A)50%
B)40%
C)30%
D)20%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
6
The light field and radiation field of a radiographic unit must be congruent to within ____ of the SID.

A)±1%
B)±2%
C)±3%
D)±5%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
7
The minimum filtration required for general-purpose radiographic systems is _____ mm Al.

A)2.0
B)2.1
C)2.5
D)3.0
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
8
The x-ray tube must be mounted inside its housing so that the central ray exits the housing to within ____ degree(s) of perpendicular.

A)1
B)2
C)3
D)4
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
9
During testing for grid uniformity, films should be uniform within an optical density of

A)±0.4.
B)±0.3.
C)±0.2.
D)±0.1.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
10
The tomographic exposure angle indicated on the equipment and the actual tomographic angle must coincide within ____ degrees for units operating at angles greater than 30 degrees.

A)2
B)5
C)10
D)20
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
11
Conventional tomographic systems should produce images of tissue that lie within ____ mm of the indicated tomographic plane.

A)5
B)10
C)15
D)20
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
12
Which of the following devices can be used as the sensor in AEC units?

A)Ion chamber
B)Photocell
C)Solid-state detector
D)All of the above
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
13
What is the accepted half-value layer (HVL) for an x-ray unit operating at 80 kVp?

A)2.1 mm Al
B)2.3 mm Al
C)2.5 mm Al
D)2.7 mm Al
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
14
X-ray system mAs reciprocity has an acceptance parameter of

A)±5%.
B)±10%.
C)±15%.
D)±20%.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
15
Path closure in pluridirectional tomographic units should be within ____ of the path length.

A)2%
B)5%
C)10%
D)15%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
16
Collimation systems equipped with positive beam limitation must be accurate to within ____ of the SID.

A)±5%
B)±4%
C)±3%
D)±2%
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
17
Two common quality control tests that should be performed on grids are grid _____ and grid _____ tests.

A)ratio; frequency
B)alignment; uniformity
C)ratio; alignment
D)uniformity; ratio
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
18
During AEC evaluation with varying kVp, the AEC system should maintain a constant film density to within ____ optical density units.

A)±0.05
B)±0.1
C)±0.2
D)±0.3
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
19
The maximum variation coefficient for reproducibility of x-ray generators is

A)0.05.
B)0.07.
C)0.10.
D)0.20.
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
20
A synchronous spinning top is used on a three-phase unit with a time of 17 ms.Approximately how large of an arc should appear if the timer circuit is accurate?

A)3 degrees
B)6 degrees
C)17 degrees
D)36 degrees
Unlock Deck
Unlock for access to all 20 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 20 flashcards in this deck.