Deck 12: Quality Control in Computed Tomography

Full screen (f)
exit full mode
Question
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
Use Space or
up arrow
down arrow
to flip the card.
Question
The laser light field and radiation field should coincide to within ____ mm.

A)0.5
B)1.0
C)2.0
D)5.0
Question
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A)quarterly.
B)monthly.
C)weekly.
D)daily.
Question
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A)daily.
B)weekly.
C)monthly.
D)semi-annually.
Question
The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.

A)20
B)40
C)30
D)80
Question
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A)2
B)3
C)5
D)10
Question
The CT number of water should not deviate by more than ____ CT numbers from 0.

A)2
B)3
C)5
D)10
Question
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A)±0.5
B)±1.0
C)±2.0
D)±5.0
Question
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A)0.5
B)1.0
C)2.0
D)3.0
Question
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A)±2%
B)±5%
C)±10%
D)±20%
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/10
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 12: Quality Control in Computed Tomography
1
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
C
2
The laser light field and radiation field should coincide to within ____ mm.

A)0.5
B)1.0
C)2.0
D)5.0
C
3
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

A)quarterly.
B)monthly.
C)weekly.
D)daily.
D
4
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

A)daily.
B)weekly.
C)monthly.
D)semi-annually.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
5
The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.

A)20
B)40
C)30
D)80
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
6
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

A)2
B)3
C)5
D)10
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
7
The CT number of water should not deviate by more than ____ CT numbers from 0.

A)2
B)3
C)5
D)10
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
8
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

A)±0.5
B)±1.0
C)±2.0
D)±5.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
9
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

A)0.5
B)1.0
C)2.0
D)3.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
10
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

A)±2%
B)±5%
C)±10%
D)±20%
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 10 flashcards in this deck.