Exam 12: Quality Control in Computed Tomography

arrow
  • Select Tags
search iconSearch Question
flashcardsStudy Flashcards
  • Select Tags

A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.

Free
(Multiple Choice)
4.8/5
(35)
Correct Answer:
Verified

C

For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.

Free
(Multiple Choice)
4.9/5
(37)
Correct Answer:
Verified

A

Section thickness should not vary by more than ____ mm for sections greater than 5 mm.

Free
(Multiple Choice)
4.7/5
(32)
Correct Answer:
Verified

B

Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.

(Multiple Choice)
4.7/5
(36)

The CT number of water should not deviate by more than ____ CT numbers from 0.

(Multiple Choice)
4.8/5
(34)

For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist

(Multiple Choice)
4.8/5
(28)

The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.

(Multiple Choice)
4.7/5
(36)

The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.

(Multiple Choice)
4.8/5
(30)

The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?

(Multiple Choice)
4.8/5
(30)

The laser light field and radiation field should coincide to within ____ mm.

(Multiple Choice)
4.9/5
(34)