Deck 8: Exposure Technique Selection

Full screen (f)
exit full mode
Question
Which exposure variable is controlled by the automatic exposure control device?

A) kVp
B) Focal spot size
C) Seconds
D) mA
Use Space or
up arrow
down arrow
to flip the card.
Question
Where is the sensor located in an automatic exposure device with a gas-filled chamber?

A) In front of the patient
B) Behind the Bucky tray
C) In front of the image receptor
D) On the tabletop
Question
Which automatic exposure device works by converting x-ray photons first into light and then into an electronic signal?

A) Phototimer
B) Gas detector
C) Ionization chamber
D) Backup timer
Question
If the detector is directly exposed to the primary beam during a film-screen study,the area of interest will appear:

A) overexposed.
B) underexposed.
C) correctly exposed.
D) to have high contrast.
Question
Which automatic exposure device works by using a gas-filled chamber?

A) Phototimer
B) Diode sensor
C) Ionization chamber
D) Bucky proliferator
Question
When doing a film-screen study using AEC,what will happen to the density in the area of interest when changing from 70 to 90 kVp?

A) Density will increase.
B) Density will decrease.
C) Density will not change.
Question
Why was automatic exposure control (AEC)developed?

A) To decrease the need for technique charts
B) To make the radiographer more of a "button pusher"
C) To determine the radiation exposure required to produce a quality image
D) To increase the productivity and the patient dose
Question
With modern x-ray equipment,the AEC system's minimum response time is typically:

A) 1 ms.
B) 10 ms.
C) 50 ms.
D) 100 ms.
Question
What would the result be if a film-screen radiographic examination required an exposure time shorter than the minimum response time of the AEC sensors?

A) Image density would be optimal.
B) Image density would be excessive.
C) Response time has nothing to do with AEC.
D) Image contrast would be high.
Question
What describes the shortest exposure time required for the AEC device to operate?

A) Backup time
B) Minimum response time
C) Master density time
D) A and B
E) A and C
Question
When using AEC with film-screen,what should the radiographer adjust to manipulate the overall density of the image?

A) kVp
B) Backup time
C) Focal spot size
D) Density control
Question
In order to become familiar with manual technical factors (i.e.,setting mAs and kVp and not using AEC),it is very helpful to pay attention to the:

A) exposure time.
B) heat units.
C) density control.
D) mAs readout.
Question
What is the purpose of setting a backup time/mAs?

A) To prevent excessive exposure of the patient
B) To make certain the contrast is acceptable if the chosen kVp is too low
C) To make certain the density is acceptable if the patient moves
D) To prevent the selection of the wrong detector
Question
When using AEC,what should the radiographer adjust to manipulate the contrast of the image?

A) kVp
B) Detector selection
C) Exposure time
D) Focal spot size
Question
What is the appropriate setting for backup time/mAs?

A) 50% of the expected mAs
B) 100% of the expected mAs
C) 150% of the expected mAs
D) 400% of the expected mAs
Question
Which of the following is a system that allows the radiographer to select a particular button on the control panel that represents an anatomic area and displays a preprogrammed set of exposure factors?

A) mAs readout
B) Anatomically programmed technique
C) Automatic exposure control
D) Phototiming
Question
AEC eliminates the need for the radiographer to select the actual _____ the examination will require.

A) kVp
B) SID
C) exposure time
D) focal spot size
Question
Once a predetermined amount of radiation is transmitted through a patient to the AEC detector,which of the following occurs?

A) The x-ray exposure increases.
B) The x-ray exposure decreases.
C) The x-ray exposure terminates.
D) A and B.
Question
Which of the following terms does not belong in the list?

A) Photodiode
B) Ion chamber
C) Ionization chamber
D) mAs readout
Question
Where is the sensor located in an automatic exposure device that converts light into an electrical signal?

A) In front of the patient
B) Behind the image receptor
C) In front of the image receptor
D) On the tabletop
Question
Where does a radiographer find the established guidelines for selecting exposure factors for a radiographic examination?

A) Technique chart
B) Generator control panel
C) Examination requisition
D) A or B
E) B or C
Question
Exposure time will ______ using AEC when patient thickness decreases.

A) decrease
B) remain the same
C) increase
D) double
Question
Radiographic contrast would be standardized with the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Question
What is the purpose of a technique chart?

A) To decrease the patient exposure by always using the highest kVp required
B) To ensure consistent image quality
C) To reduce the repeat imaging due to technique error and, in turn, reduce the patient exposure
D) A and B
E) B and C
Question
If a part measures 14 cm and requires the use of 75 kVp,how much kVp would a part measuring 18 cm require when using a variable kVp-fixed mAs technique chart?

A) 67 kVp
B) 77 kVp
C) 83 kVp
D) 91 kVp
Question
Which of the following is/are true regarding the development of an effective technique chart?

A) The radiographic equipment must be operating properly.
B) A good-quality control program is in place.
C) The equipment must be from the same manufacturer.
D) A and B.
E) A and C.
Question
Which of the following is not part of a standardized technique chart?

A) SID
B) OID
C) mA
D) kVp
Question
The most common cause for repeating images done with film-screen and AEC is inaccurate:

A) kVp selection.
B) backup mAs selection.
C) centering.
D) alignment of the tube and image receptor.
Question
What are calipers used for?

A) To measure the mA
B) To select the kVp
C) To select focal spot size
D) To measure the part
Question
Accurate patient measurement is most critical for the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Question
When doing a tabletop forearm study with CR,you should:

A) use AEC.
B) set manual technique.
C) A and B.
Question
When using a fixed kVp-variable mAs technique chart,if part thickness increases by 5 cm,what needs to happen to the mAs?

A) It should be halved.
B) It should be doubled.
C) It should be quartered.
D) It should be tripled.
Question
When a digital IR is overexposed due to poor use of AEC:

A) image contrast will be increased.
B) quantum noise may be highly visible.
C) patient exposure is increased.
D) patient exposure is unaffected.
Question
The mAs readout:

A) is of little value because it's not necessary to know mAs when using AEC.
B) reports the actual mAs used for an exposure.
C) reports the patient dose for the entire imaging study.
D) allows the radiographer to better set the mAs for an AEC examination.
Question
If the backup time/mAs is automatically set by the x-ray unit,the exposure should end when a maximum of _________ has been reached.

A) 600 s
B) 600 kVp
C) 600 mA
D) 600 mAs
Question
What is the appropriate change in kVp when using a variable kVp-fixed mAs technique chart for a 1 cm change in tissue thickness?

A) 2 kVp
B) 4 kVp
C) 6 kVp
D) 10 kVp
Question
What kind of chart uses a kVp value that is high enough to adequately penetrate the part but does not diminish radiographic contrast?

A) A variable kVp technique chart
B) A fixed kVp technique chart
C) A fixed mAs technique chart
D) A and B
E) A and C
Question
Departmental standards,such as the SID or whether an examination is done tabletop or Bucky,should be determined:

A) prior to the development of technique charts.
B) after the development of technique charts.
C) at the same time as the development of technique charts.
D) whenever the department can come to an agreement.
Question
With a CR system,when using AEC to image a patient who is having difficulty holding still,increasing the mA (if the unit allows)will:

A) be a bad idea because the patient will be overexposed.
B) be a bad idea because the patient will be underexposed.
C) be a good idea because the exposure time will increase.
D) be a good idea because the exposure time will decrease.
Question
A technique chart should be established for:

A) each radiographer.
B) each radiographic tube.
C) each radiologist.
D) A and B.
E) A and C.
Question
When using film-screen,what happens when an outside detector is selected for an AP thoracic spine?

A) The spine will have insufficient contrast.
B) The spine will be too dark.
C) The spine will have excessive contrast.
D) The spine will have insufficient density.
Question
The technique chart will provide information as to how to recognize and compensate for additive pathology.
Question
A good time to use AEC is when the radiographer is unable to accurately center the x-ray beam.
Question
Generally speaking,patient dose will be decreased with the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Question
Using the concept of comparative anatomy,if a child's elbow measures the same as an adult wrist and all other procedural variables are the same (i.e.,SID,image receptor type,etc.),the child's elbow will require _____ in comparison to the adult wrist.

A) less mAs
B) more mAs
C) the same mAs
D) higher kVp
Question
AEC can be used with tabletop examinations.
Question
Optimal kVp values for each anatomic area have been established by law.
Question
A good reason to use AEC density control is to increase the exposure to the IR to compensate for improper centering.
Question
With AEC it is easy to use multiple types of image receptors,because the detectors can sense image receptor type and make appropriate adjustments.
Question
Newer x-ray units may keep you from making an exposure if you've selected the table Bucky on the control panel but there's no IR in the table Bucky.
Question
A patient measuring 26 cm requires 20 mAs at 75 kVp based on a fixed kVp-variable mAs technique chart.What technique should be used for the next patient,who measures 34 cm?

A) 40 mAs at 75 kVp
B) 80 mAs at 75 kVp
C) 20 mAs at 91 kVp
D) 40 mAs at 91 kVp
Question
A variable kVp-fixed mAs chart may be most effective with:

A) geriatric patients.
B) trauma patients.
C) patients imaged with mobile equipment.
D) pediatric patients.
Question
When exposure factors are selected using anatomically programmed technique,the variables should not be adjusted.
Question
Technique charts are still needed when using AEC.
Question
Two identical rooms having the same generator and tube and installed at the same time do not require development of separate technique charts.The chart for one room can accurately be used in the other.
Question
If a technique chart is being used,and the IR is overexposed,the radiographer should assume the chart is inaccurate and come up with his or her own exposure factors.
Question
When the radiographer is considering whether to repeat an image using AEC,it is important to determine the reason the first radiograph failed.
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/57
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 8: Exposure Technique Selection
1
Which exposure variable is controlled by the automatic exposure control device?

A) kVp
B) Focal spot size
C) Seconds
D) mA
Seconds
2
Where is the sensor located in an automatic exposure device with a gas-filled chamber?

A) In front of the patient
B) Behind the Bucky tray
C) In front of the image receptor
D) On the tabletop
In front of the image receptor
3
Which automatic exposure device works by converting x-ray photons first into light and then into an electronic signal?

A) Phototimer
B) Gas detector
C) Ionization chamber
D) Backup timer
Phototimer
4
If the detector is directly exposed to the primary beam during a film-screen study,the area of interest will appear:

A) overexposed.
B) underexposed.
C) correctly exposed.
D) to have high contrast.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
5
Which automatic exposure device works by using a gas-filled chamber?

A) Phototimer
B) Diode sensor
C) Ionization chamber
D) Bucky proliferator
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
6
When doing a film-screen study using AEC,what will happen to the density in the area of interest when changing from 70 to 90 kVp?

A) Density will increase.
B) Density will decrease.
C) Density will not change.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
7
Why was automatic exposure control (AEC)developed?

A) To decrease the need for technique charts
B) To make the radiographer more of a "button pusher"
C) To determine the radiation exposure required to produce a quality image
D) To increase the productivity and the patient dose
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
8
With modern x-ray equipment,the AEC system's minimum response time is typically:

A) 1 ms.
B) 10 ms.
C) 50 ms.
D) 100 ms.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
9
What would the result be if a film-screen radiographic examination required an exposure time shorter than the minimum response time of the AEC sensors?

A) Image density would be optimal.
B) Image density would be excessive.
C) Response time has nothing to do with AEC.
D) Image contrast would be high.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
10
What describes the shortest exposure time required for the AEC device to operate?

A) Backup time
B) Minimum response time
C) Master density time
D) A and B
E) A and C
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
11
When using AEC with film-screen,what should the radiographer adjust to manipulate the overall density of the image?

A) kVp
B) Backup time
C) Focal spot size
D) Density control
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
12
In order to become familiar with manual technical factors (i.e.,setting mAs and kVp and not using AEC),it is very helpful to pay attention to the:

A) exposure time.
B) heat units.
C) density control.
D) mAs readout.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
13
What is the purpose of setting a backup time/mAs?

A) To prevent excessive exposure of the patient
B) To make certain the contrast is acceptable if the chosen kVp is too low
C) To make certain the density is acceptable if the patient moves
D) To prevent the selection of the wrong detector
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
14
When using AEC,what should the radiographer adjust to manipulate the contrast of the image?

A) kVp
B) Detector selection
C) Exposure time
D) Focal spot size
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
15
What is the appropriate setting for backup time/mAs?

A) 50% of the expected mAs
B) 100% of the expected mAs
C) 150% of the expected mAs
D) 400% of the expected mAs
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
16
Which of the following is a system that allows the radiographer to select a particular button on the control panel that represents an anatomic area and displays a preprogrammed set of exposure factors?

A) mAs readout
B) Anatomically programmed technique
C) Automatic exposure control
D) Phototiming
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
17
AEC eliminates the need for the radiographer to select the actual _____ the examination will require.

A) kVp
B) SID
C) exposure time
D) focal spot size
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
18
Once a predetermined amount of radiation is transmitted through a patient to the AEC detector,which of the following occurs?

A) The x-ray exposure increases.
B) The x-ray exposure decreases.
C) The x-ray exposure terminates.
D) A and B.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
19
Which of the following terms does not belong in the list?

A) Photodiode
B) Ion chamber
C) Ionization chamber
D) mAs readout
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
20
Where is the sensor located in an automatic exposure device that converts light into an electrical signal?

A) In front of the patient
B) Behind the image receptor
C) In front of the image receptor
D) On the tabletop
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
21
Where does a radiographer find the established guidelines for selecting exposure factors for a radiographic examination?

A) Technique chart
B) Generator control panel
C) Examination requisition
D) A or B
E) B or C
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
22
Exposure time will ______ using AEC when patient thickness decreases.

A) decrease
B) remain the same
C) increase
D) double
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
23
Radiographic contrast would be standardized with the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
24
What is the purpose of a technique chart?

A) To decrease the patient exposure by always using the highest kVp required
B) To ensure consistent image quality
C) To reduce the repeat imaging due to technique error and, in turn, reduce the patient exposure
D) A and B
E) B and C
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
25
If a part measures 14 cm and requires the use of 75 kVp,how much kVp would a part measuring 18 cm require when using a variable kVp-fixed mAs technique chart?

A) 67 kVp
B) 77 kVp
C) 83 kVp
D) 91 kVp
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
26
Which of the following is/are true regarding the development of an effective technique chart?

A) The radiographic equipment must be operating properly.
B) A good-quality control program is in place.
C) The equipment must be from the same manufacturer.
D) A and B.
E) A and C.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
27
Which of the following is not part of a standardized technique chart?

A) SID
B) OID
C) mA
D) kVp
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
28
The most common cause for repeating images done with film-screen and AEC is inaccurate:

A) kVp selection.
B) backup mAs selection.
C) centering.
D) alignment of the tube and image receptor.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
29
What are calipers used for?

A) To measure the mA
B) To select the kVp
C) To select focal spot size
D) To measure the part
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
30
Accurate patient measurement is most critical for the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
31
When doing a tabletop forearm study with CR,you should:

A) use AEC.
B) set manual technique.
C) A and B.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
32
When using a fixed kVp-variable mAs technique chart,if part thickness increases by 5 cm,what needs to happen to the mAs?

A) It should be halved.
B) It should be doubled.
C) It should be quartered.
D) It should be tripled.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
33
When a digital IR is overexposed due to poor use of AEC:

A) image contrast will be increased.
B) quantum noise may be highly visible.
C) patient exposure is increased.
D) patient exposure is unaffected.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
34
The mAs readout:

A) is of little value because it's not necessary to know mAs when using AEC.
B) reports the actual mAs used for an exposure.
C) reports the patient dose for the entire imaging study.
D) allows the radiographer to better set the mAs for an AEC examination.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
35
If the backup time/mAs is automatically set by the x-ray unit,the exposure should end when a maximum of _________ has been reached.

A) 600 s
B) 600 kVp
C) 600 mA
D) 600 mAs
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
36
What is the appropriate change in kVp when using a variable kVp-fixed mAs technique chart for a 1 cm change in tissue thickness?

A) 2 kVp
B) 4 kVp
C) 6 kVp
D) 10 kVp
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
37
What kind of chart uses a kVp value that is high enough to adequately penetrate the part but does not diminish radiographic contrast?

A) A variable kVp technique chart
B) A fixed kVp technique chart
C) A fixed mAs technique chart
D) A and B
E) A and C
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
38
Departmental standards,such as the SID or whether an examination is done tabletop or Bucky,should be determined:

A) prior to the development of technique charts.
B) after the development of technique charts.
C) at the same time as the development of technique charts.
D) whenever the department can come to an agreement.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
39
With a CR system,when using AEC to image a patient who is having difficulty holding still,increasing the mA (if the unit allows)will:

A) be a bad idea because the patient will be overexposed.
B) be a bad idea because the patient will be underexposed.
C) be a good idea because the exposure time will increase.
D) be a good idea because the exposure time will decrease.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
40
A technique chart should be established for:

A) each radiographer.
B) each radiographic tube.
C) each radiologist.
D) A and B.
E) A and C.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
41
When using film-screen,what happens when an outside detector is selected for an AP thoracic spine?

A) The spine will have insufficient contrast.
B) The spine will be too dark.
C) The spine will have excessive contrast.
D) The spine will have insufficient density.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
42
The technique chart will provide information as to how to recognize and compensate for additive pathology.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
43
A good time to use AEC is when the radiographer is unable to accurately center the x-ray beam.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
44
Generally speaking,patient dose will be decreased with the:

A) variable kVp-fixed mAs technique chart design.
B) variable kVp-variable mAs technique chart design.
C) fixed kVp-variable mAs technique chart design.
D) fixed kVp-fixed mAs technique chart design.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
45
Using the concept of comparative anatomy,if a child's elbow measures the same as an adult wrist and all other procedural variables are the same (i.e.,SID,image receptor type,etc.),the child's elbow will require _____ in comparison to the adult wrist.

A) less mAs
B) more mAs
C) the same mAs
D) higher kVp
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
46
AEC can be used with tabletop examinations.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
47
Optimal kVp values for each anatomic area have been established by law.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
48
A good reason to use AEC density control is to increase the exposure to the IR to compensate for improper centering.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
49
With AEC it is easy to use multiple types of image receptors,because the detectors can sense image receptor type and make appropriate adjustments.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
50
Newer x-ray units may keep you from making an exposure if you've selected the table Bucky on the control panel but there's no IR in the table Bucky.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
51
A patient measuring 26 cm requires 20 mAs at 75 kVp based on a fixed kVp-variable mAs technique chart.What technique should be used for the next patient,who measures 34 cm?

A) 40 mAs at 75 kVp
B) 80 mAs at 75 kVp
C) 20 mAs at 91 kVp
D) 40 mAs at 91 kVp
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
52
A variable kVp-fixed mAs chart may be most effective with:

A) geriatric patients.
B) trauma patients.
C) patients imaged with mobile equipment.
D) pediatric patients.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
53
When exposure factors are selected using anatomically programmed technique,the variables should not be adjusted.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
54
Technique charts are still needed when using AEC.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
55
Two identical rooms having the same generator and tube and installed at the same time do not require development of separate technique charts.The chart for one room can accurately be used in the other.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
56
If a technique chart is being used,and the IR is overexposed,the radiographer should assume the chart is inaccurate and come up with his or her own exposure factors.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
57
When the radiographer is considering whether to repeat an image using AEC,it is important to determine the reason the first radiograph failed.
Unlock Deck
Unlock for access to all 57 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 57 flashcards in this deck.