Deck 9: Image Evaluation

Full screen (f)
exit full mode
Question
A digital image with decreased spatial resolution may be due to:
1)Increased central ray angulation
2)Decreased grid ratio
3)Decreased SID

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Use Space or
up arrow
down arrow
to flip the card.
Question
A film-screen image with increased recorded detail may be due to:
1)Increased SID
2)Decreased focal spot size
3)Decreased film-screen speed

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Question
Using a Konica digital imaging system,if an image's S number is 600 while the optimal range ends at 300,this means that the IR was ______________ and the exposure to the IR needs to be ___________________.

A) underexposed; doubled
B) under exposed; halved
C) overexposed; doubled
D) overexposed; halved
Question
A film-screen image with decreased distortion may be due to:
1)Decreased patient thickness
2)Increased SID
3)Increased focal spot size

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Question
A digital image with higher contrast may be due to:
1)Decreased kVp
2)Increased grid ratio
3)Decreased patient thickness

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Question
A film-screen image with increased density may be due to:
1)Increased OID
2)Decreased SID
3)Increased film-screen speed

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Question
An area of decreased brightness visualized on a computer monitor has ________density on a film image.

A) increased
B) decreased
C) no
Question
When using a film-screen imaging system,changing from small to large focal spot size would result in _____________ density and ____________ recorded detail.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Question
The amount of luminance (light emission)of a display monitor is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Question
Using an Agfa digital imaging system,if an image's exposure indicator is 2.3 lgM while the optimal range begins at 2.6 lgM,this means that the IR was ______________ and the exposure to the IR needs to be ___________________ .

A) underexposed; increased by 30%
B) underexposed; increased by a factor of two
C) overexposed; decreased by 30%
D) overexposed; decreased by a factor of two
Question
When using a digital imaging system,changing from 8 × 10 collimation to 14 × 17 collimation would result in _____________ brightness and ____________ contrast.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Question
The visibility of anatomic structures in a radiographic image is determined by:

A) spatial resolution and image contrast.
B) brightness/density and image contrast.
C) spatial resolution and brightness/density.
D) all of the above.
Question
A digital image with increased distortion may be due to:
1)Decreased OID
2)Increased CR angulation
3)Decreased SID

A)1 and 2 only.
B)1 and 3 only.
C)2 and 3 only.
D)1,2,and 3.
Question
When using a digital imaging system,changing from 40 in SID to 72 in SID would result in _____________ brightness and ____________ contrast.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Question
A film-screen image with lower contrast may be due to:
1)Increased patient thickness
2)Increased beam restriction
3)Increased kVp

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Question
A film-screen image produced using 16 mAs at 80 kVp has to be repeated because it is too dark.Which of the following exposure factors would be the best choice to correct the problem?

A) 4 mAs at 80 kVp
B) 8 mAs at 80 kVp
C) 8 mAs at 68 kVp
D) 16 mAs at 68 kVp
Question
Which of the following changes would produce a digital image with decreased brightness?

A) Increased mAs
B) Decreased kVp
C) Increased patient thickness
D) None of the above
Question
Using a Carestream digital imaging system,if an image's EI number is 1800 while the optimal range ends at 1200,this means that the IR was ______________ and the exposure to the IR needs to be ___________________.

A) underexposed; doubled
B) underexposed; increased by a factor of four
C) overexposed; halved
D) overexposed; reduced by a factor of four
Question
The amount of overall blackness on the processed image is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Question
The variation in brightness and density on digital or film imaging is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Question
In order to produce a radiographic image with structural lines as accurate as possible,you must maximize the spatial resolution and distortion.
Question
Assuming no other changes are made,which of the following would result in increased quantum noise in a digital image?
1)Decreasing mAs
2)Increasing kVp
3)Imaging a thicker part

A) 1 and 2 only
B) 1 and 3 only
C) 2 and 3 only
D) 1, 2, and 3
Question
Which of the following would reduce quantum noise in a digital image?
1)Decreasing mAs
2)Increasing mAs
3)Changing to a lower ratio grid

A) 1 and 2 only
B) 1 and 3 only
C) 2 and 3 only
D) 1, 2, and 3
Question
Artifacts are detrimental to radiographs because they can make the visibility of anatomy,pathologic conditions,or patient identification information difficult or impossible.
Question
______________ is a universal standard value that reflects the difference between the desired or target exposure to the IR and the actual exposure to the IR.

A) Exposure index
B) Sensitivity number
C) Deviation index
D) Log median number
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/25
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 9: Image Evaluation
1
A digital image with decreased spatial resolution may be due to:
1)Increased central ray angulation
2)Decreased grid ratio
3)Decreased SID

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
1 and 3 only.
2
A film-screen image with increased recorded detail may be due to:
1)Increased SID
2)Decreased focal spot size
3)Decreased film-screen speed

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
1, 2, and 3.
3
Using a Konica digital imaging system,if an image's S number is 600 while the optimal range ends at 300,this means that the IR was ______________ and the exposure to the IR needs to be ___________________.

A) underexposed; doubled
B) under exposed; halved
C) overexposed; doubled
D) overexposed; halved
underexposed; doubled
4
A film-screen image with decreased distortion may be due to:
1)Decreased patient thickness
2)Increased SID
3)Increased focal spot size

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
5
A digital image with higher contrast may be due to:
1)Decreased kVp
2)Increased grid ratio
3)Decreased patient thickness

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
6
A film-screen image with increased density may be due to:
1)Increased OID
2)Decreased SID
3)Increased film-screen speed

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
7
An area of decreased brightness visualized on a computer monitor has ________density on a film image.

A) increased
B) decreased
C) no
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
8
When using a film-screen imaging system,changing from small to large focal spot size would result in _____________ density and ____________ recorded detail.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
9
The amount of luminance (light emission)of a display monitor is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
10
Using an Agfa digital imaging system,if an image's exposure indicator is 2.3 lgM while the optimal range begins at 2.6 lgM,this means that the IR was ______________ and the exposure to the IR needs to be ___________________ .

A) underexposed; increased by 30%
B) underexposed; increased by a factor of two
C) overexposed; decreased by 30%
D) overexposed; decreased by a factor of two
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
11
When using a digital imaging system,changing from 8 × 10 collimation to 14 × 17 collimation would result in _____________ brightness and ____________ contrast.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
12
The visibility of anatomic structures in a radiographic image is determined by:

A) spatial resolution and image contrast.
B) brightness/density and image contrast.
C) spatial resolution and brightness/density.
D) all of the above.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
13
A digital image with increased distortion may be due to:
1)Decreased OID
2)Increased CR angulation
3)Decreased SID

A)1 and 2 only.
B)1 and 3 only.
C)2 and 3 only.
D)1,2,and 3.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
14
When using a digital imaging system,changing from 40 in SID to 72 in SID would result in _____________ brightness and ____________ contrast.

A) no change in; no change in
B) decreased; increased
C) no change in; decreased
D) decreased; decreased
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
15
A film-screen image with lower contrast may be due to:
1)Increased patient thickness
2)Increased beam restriction
3)Increased kVp

A) 1 and 2 only.
B) 1 and 3 only.
C) 2 and 3 only.
D) 1, 2, and 3.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
16
A film-screen image produced using 16 mAs at 80 kVp has to be repeated because it is too dark.Which of the following exposure factors would be the best choice to correct the problem?

A) 4 mAs at 80 kVp
B) 8 mAs at 80 kVp
C) 8 mAs at 68 kVp
D) 16 mAs at 68 kVp
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
17
Which of the following changes would produce a digital image with decreased brightness?

A) Increased mAs
B) Decreased kVp
C) Increased patient thickness
D) None of the above
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
18
Using a Carestream digital imaging system,if an image's EI number is 1800 while the optimal range ends at 1200,this means that the IR was ______________ and the exposure to the IR needs to be ___________________.

A) underexposed; doubled
B) underexposed; increased by a factor of four
C) overexposed; halved
D) overexposed; reduced by a factor of four
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
19
The amount of overall blackness on the processed image is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
20
The variation in brightness and density on digital or film imaging is the:

A) density.
B) contrast.
C) SID.
D) brightness.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
21
In order to produce a radiographic image with structural lines as accurate as possible,you must maximize the spatial resolution and distortion.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
22
Assuming no other changes are made,which of the following would result in increased quantum noise in a digital image?
1)Decreasing mAs
2)Increasing kVp
3)Imaging a thicker part

A) 1 and 2 only
B) 1 and 3 only
C) 2 and 3 only
D) 1, 2, and 3
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
23
Which of the following would reduce quantum noise in a digital image?
1)Decreasing mAs
2)Increasing mAs
3)Changing to a lower ratio grid

A) 1 and 2 only
B) 1 and 3 only
C) 2 and 3 only
D) 1, 2, and 3
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
24
Artifacts are detrimental to radiographs because they can make the visibility of anatomy,pathologic conditions,or patient identification information difficult or impossible.
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
25
______________ is a universal standard value that reflects the difference between the desired or target exposure to the IR and the actual exposure to the IR.

A) Exposure index
B) Sensitivity number
C) Deviation index
D) Log median number
Unlock Deck
Unlock for access to all 25 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 25 flashcards in this deck.