Deck 12: Quality Systems in the Pulmonary Function Laboratory

Full screen (f)
exit full mode
Question
A helium analyzer is checked using a 3-L syringe;300 ml of He is added to 2.70 L of air in the syringe.If the dead space of the syringe and tubing is 100 ml,what is the expected concentration of He in the syringe?

A) 11.1%
B) 0.2%
C) 9.7%
D) 8.8%
Use Space or
up arrow
down arrow
to flip the card.
Question
An automated blood gas analyzer displays acceptable calibrations with no electrode drift noted.Multiple-level QC shows values for PCO2 and PO2 that are within 2 SD of the mean,but all pH values are more than 3 SD above the mean.Which of the following is most likely the cause?

A) Buffers are contaminated.
B) Calibration gas is contaminated.
C) Analyzer temperature is not 37° C.
D) Gas-liquid correction factor is set too low.
Question
A 3-L syringe is used to calibrate a flow-sensing spirometer;the spirometer reads a volume of 3.04 L.The software correction factor for this system would be which of the following?

A) 0.95
B) 0.99
C) 1.01
D) 1.04
Question
A biologic control subject has an established FVC of 5.00 L with an SD of 0.20 L.After spirometer maintenance,the following values are obtained from the biologic control: <strong>A biologic control subject has an established FVC of 5.00 L with an SD of 0.20 L.After spirometer maintenance,the following values are obtained from the biologic control:   On the basis of these findings,the pulmonary function technologist should conclude what?</strong> A) Spirometer performance is within acceptable statistical limits. B) Spirometer performance is questionable. C) The control subject's efforts were greater than normal. D) The control subject's efforts were not temperature-corrected to BTPS. <div style=padding-top: 35px> On the basis of these findings,the pulmonary function technologist should conclude what?

A) Spirometer performance is within acceptable statistical limits.
B) Spirometer performance is questionable.
C) The control subject's efforts were greater than normal.
D) The control subject's efforts were not temperature-corrected to BTPS.
Question
Multiple injections from a 3-L syringe into a flow-based spirometer produce an average reading of 2.92 L.This value indicates:

A) The spirometer has a leak.
B) The spirometer meets ATS accuracy requirements.
C) Excessive flows were used for calibration.
D) The spirometer should not be used for pediatric patients.
Question
Tonometry of whole blood may be used to provide QC for which of the following?

A) PO2 and PCO2 electrodes
B) PCO2 and pH electrodes
C) PO2 electrode only
D) pH electrode only
Question
Quality control of an automated DLCO system is performed using a 3-L syringe.The syringe is filled to 1.5 L with room air and is then connected to the system.A simulated DLCOsb maneuver is performed with 1.5 L of test gas drawn into the syringe.Acceptable performance of the gas analyzers should produce a DLCO of approximately:

A) 0 ml CO/min/mm Hg
B) 25 ml CO/min/mm Hg
C) 50 ml CO/min/mm Hg
D) 100 ml CO/min/mm Hg
Question
A spirometer recorder used for plotting flow-volume curves has a volume (x axis)sensitivity of 25 mm/L.The flow axis (y axis)sensitivity should be:

A) 12.5 mm/L/sec
B) 25.0 mm/L/sec
C) 50.0 mm/L/sec
D) 50.0 mm/L
Question
According to ATS recommendations,the acceptable "back pressure" (flow resistance)from a spirometer is less than _____ cm H2O at _____ L/sec.

A) 3.0;7
B) 3.0;14
C) 1.5;7
D) 1.5;14
Question
According to ATS recommendations,the range and accuracy for a diagnostic spirometer should be which of the following?

A) 0 to 14 L with less than 1.5 cm H2O/L/sec resistance
B) 0.5 to 8 L ± 3% of reading or ± 0.05 L,whichever is greater
C) 0.5 to 6 L ± 5% of reading or ± 0.1 L,whichever is greater
D) 0.1 to 5 L ± 3% of reading or ± 0.05 L,whichever is greater
Question
A CO2 analyzer is calibrated using a 5% CO2 mixture with a water vapor absorber in the sample line.If the barometric pressure is 747 mm Hg and the analyzer is set up to read partial pressure,the analyzer should be adjusted to read which of the following?

A) 40 mm Hg
B) 37 mm Hg
C) 35 mm Hg
D) 31 mm Hg
Question
Quality control run on a PO2 electrode shows the following results when plotted on a QC chart: <strong>Quality control run on a PO<sub>2</sub> electrode shows the following results when plotted on a QC chart:   Which of the following best describes the result of run 7?</strong> A) Out-of-control situation B) Shift C) Trend D) Random error <div style=padding-top: 35px> Which of the following best describes the result of run 7?

A) Out-of-control situation
B) Shift
C) Trend
D) Random error
Question
A 50-ml sine-wave pump is used to calibrate a 450-L plethysmograph.While the pump is running,the computer display shows a deflection of 25 cm.What is the box calibration factor for this system?

A) 0.5 L/cm
B) 1.25 L/cm
C) 2 ml/cm
D) 0.5 ml/cm
Question
Quality control of a body plethysmograph is performed using an isothermal lung analog and a biologic control subject with the following results: <strong>Quality control of a body plethysmograph is performed using an isothermal lung analog and a biologic control subject with the following results:   On the basis of these findings,the pulmonary function technologist should conclude what?</strong> A) There is a leak in the door seal. B) Box pressure calibration was performed incorrectly. C) Mouth pressure calibration was performed incorrectly. D) The box is functioning within acceptable limits. <div style=padding-top: 35px> On the basis of these findings,the pulmonary function technologist should conclude what?

A) There is a leak in the door seal.
B) Box pressure calibration was performed incorrectly.
C) Mouth pressure calibration was performed incorrectly.
D) The box is functioning within acceptable limits.
Question
A patient who is short of breath performs eight FVC maneuvers,and these results from the three best efforts are recorded: <strong>A patient who is short of breath performs eight FVC maneuvers,and these results from the three best efforts are recorded:   Which of the following is appropriate to report in the technologist's comments?</strong> A) Spirometry meets all ATS criteria. B) FVC is not repeatable. C) FEV<sub>1</sub> is not repeatable. D) FEF<sub>50% </sub>shows poor patient effort. <div style=padding-top: 35px> Which of the following is appropriate to report in the technologist's comments?

A) "Spirometry meets all ATS criteria."
B) "FVC is not repeatable."
C) "FEV1 is not repeatable."
D) "FEF50% shows poor patient effort."
Question
A CO2 analyzer is calibrated using a 5% CO2 mixture with the sampling pump set at 500 ml/min;a water vapor absorber is then added to the sample line.Which of the following should be done before using this system?

A) Increase the sampling pump flow to at least 750 ml/min.
B) Decrease the sampling pump flow to less than 50 ml/min.
C) Recalibrate the analyzer.
D) Calculate the "corrected" CO2 reading.
Question
A 3-L syringe is used to calibrate a computerized spirometer;at ATPS the spirometer reads a volume of 2.94 L.The software correction factor for this system is:

A) 1.02
B) 1.00
C) 0.98
D) 0.95
Question
The pulmonary function technologist should observe which of the following precautions in drawing an arterial blood sample?

A) Determine whether gloves are needed by checking the patient's history.
B) Carefully resheath used needles using the original protective covering.
C) Dispose of used needles in puncture-resistant containers marked "biohazard."
D) Use protective barriers if there is danger of blood splashing.
Question
An automated blood gas analyzer performs two-point calibrations that match expected values.Subsequent quality control runs show unacceptable values for PO2 and PCO2,but pH is acceptable.Which of the following is the most likely problem?

A) Contaminated buffers
B) Contaminated calibration gas
C) Analyzer temperature not at 37° C
D) Computer error
Question
Multiple injections by a 3-L syringe into a bellows-type spirometer produce the following results: <strong>Multiple injections by a 3-L syringe into a bellows-type spirometer produce the following results:   Which of the following best describes these results?</strong> A) The spirometer shows excessive drift. B) The volume is being inappropriately corrected to BTPS. C) The spirometer bellows are sticking. D) Spirometer performance is acceptable. <div style=padding-top: 35px> Which of the following best describes these results?

A) The spirometer shows excessive drift.
B) The volume is being inappropriately corrected to BTPS.
C) The spirometer bellows are sticking.
D) Spirometer performance is acceptable.
Question
Which of the following control materials would be best suited for quality control of a PO2 electrode used to measure O2 tensions less than 50 mm Hg?

A) Fluorocarbon-based control
B) Aqueous-based control
C) Tonometered buffer
D) Tonometered whole blood
Question
A pulmonary function technologist uses a 3-L syringe and a tank of 10% He to prepare serial dilutions of gas to verify the linearity of a helium analyzer used for closed-circuit measurements of FRC.The following data are recorded: <strong>A pulmonary function technologist uses a 3-L syringe and a tank of 10% He to prepare serial dilutions of gas to verify the linearity of a helium analyzer used for closed-circuit measurements of FRC.The following data are recorded:   On the basis of these results,the technologist should conclude that:</strong> A) The He analyzer is performing acceptably. B) The He analyzer is nonlinear. C) The syringe had a significant leak. D) The He tank was mislabeled. <div style=padding-top: 35px> On the basis of these results,the technologist should conclude that:

A) The He analyzer is performing acceptably.
B) The He analyzer is nonlinear.
C) The syringe had a significant leak.
D) The He tank was mislabeled.
Question
According to ATS-ERS recommendations (2005),a computer display of flow-volume curves should have a scale factor of:

A) 5 mm/L on the volume axis
B) 5 mm/L/sec on the flow axis
C) 20 mm/sec on the time axis
D) A 1:1 aspect ratio with expiratory flow plotted downward
Question
A patient performs spirometry,and the following results are obtained: <strong>A patient performs spirometry,and the following results are obtained:   The pulmonary function technologist should conclude which of the following?</strong> A) Only the FVC is repeatable. B) Only the FEV<sub>1</sub> is repeatable. C) Both the FVC and FEV<sub>1</sub> are repeatable. D) The FEV<sub>1</sub>/FVC is 52%. <div style=padding-top: 35px> The pulmonary function technologist should conclude which of the following?

A) Only the FVC is repeatable.
B) Only the FEV1 is repeatable.
C) Both the FVC and FEV1 are repeatable.
D) The FEV1/FVC is 52%.
Question
A 3-L syringe is used to calibrate a spirometer;injection of the complete volume produces an output of 3.04 L.The correction factor for this system would be:

A) 0.967
B) 0.987
C) 1.013
D) 1.100
Question
A subject performs spirometry,and these results are recorded from three acceptable trials: <strong>A subject performs spirometry,and these results are recorded from three acceptable trials:   The flow-volume curve from which trial should be reported?</strong> A) Trial 1 B) Trial 2 C) Trial 3 D) Either Trial 1 or Trial 3 <div style=padding-top: 35px> The flow-volume curve from which trial should be reported?

A) Trial 1
B) Trial 2
C) Trial 3
D) Either Trial 1 or Trial 3
Question
The greatest risk of cross-contamination in the pulmonary function laboratory is associated with:

A) Failure to use bacteria filters for spirometry
B) Inadequate disinfection of reusable nebulizers
C) Moisture inside dry rolling-seal spirometers
D) Pulse oximetry finger sensors
Question
A biologic control subject performs DLCO measurements on 20 consecutive days,and the following statistics are calculated: <strong>A biologic control subject performs DLCO measurements on 20 consecutive days,and the following statistics are calculated:   The coefficient of variation for the DLCO in this subject:</strong> A) Is approximately 4.5% B) Cannot be calculated until 30 measurements have been made C) Suggests that the patient may have asthma D) Indicates that the DLCO system displays excessive variability <div style=padding-top: 35px> The coefficient of variation for the DLCO in this subject:

A) Is approximately 4.5%
B) Cannot be calculated until 30 measurements have been made
C) Suggests that the patient may have asthma
D) Indicates that the DLCO system displays excessive variability
Question
Quality control runs on a PCO2 electrode have produced a mean of 40 mm Hg ± 2 mm Hg (1 SD).What are the chances of obtaining a value of 45 mm Hg on a control run if the electrode is functioning properly?

A) 95%
B) 50%
C) 5%
D) Less than 1%
Question
Quality control is performed on a cartridge-based blood gas analyzer,and the following control chart is recorded for the pH: <strong>Quality control is performed on a cartridge-based blood gas analyzer,and the following control chart is recorded for the pH:   What action should the pulmonary function technologist take,based on these results?</strong> A) Begin analyzing blood gases because the instrument is in control. B) Run one additional Level 2 control to determine instrument status. C) Take the instrument out of service because it is out of control. D) Re-membrane the pH electrode and rerun Level 2. <div style=padding-top: 35px> What action should the pulmonary function technologist take,based on these results?

A) Begin analyzing blood gases because the instrument is in control.
B) Run one additional Level 2 control to determine instrument status.
C) Take the instrument out of service because it is out of control.
D) Re-membrane the pH electrode and rerun Level 2.
Question
According to the current ATS-ERS guidelines,spirometers should be able to measure volumes of at least:

A) 7 L
B) 8 L
C) 12 L
D) 14 L
Question
A 3-L syringe is used to check the calibration of a portable spirometer,and the following results are obtained:  <strong>A 3-L syringe is used to check the calibration of a portable spirometer,and the following results are obtained:   Which of the following best describes these findings?</strong> A) The spirometer temperature is less than 24 \degree  C. B) The BTPS correction factor is being applied. C) The syringe has a small but significant leak. D) The spirometer is functioning acceptably. <div style=padding-top: 35px>  Which of the following best describes these findings?

A) The spirometer temperature is less than 24 °\degree C.
B) The BTPS correction factor is being applied.
C) The syringe has a small but significant leak.
D) The spirometer is functioning acceptably.
Question
Spirometers used to measure FEV1 should be able to measure flows in the range of 0 to 14 L/sec (ATS-ERS,2005).The maximal resistance allowable at all flows is:

A) Less than 6 cmH2O/L/sec
B) Less than 1.5 cmH2O/L/sec
C) Determined by the type of filter used
D) One half of the flow rate in L/sec
Question
A patient whose FVC is 4 L performs two DLCOsb maneuvers with the following results: <strong>A patient whose FVC is 4 L performs two DLCOsb maneuvers with the following results:   The pulmonary function technologist should:</strong> A) Reject Trial 1 because of the V<sub>I.</sub> B) Reject Trial 2 because of the breath-hold time. C) Report the DLCOsb as 22.0 ml CO/min/mm Hg. D) Report the DLCOsb as 21.2 ml CO/min/mm Hg. <div style=padding-top: 35px> The pulmonary function technologist should:

A) Reject Trial 1 because of the VI.
B) Reject Trial 2 because of the breath-hold time.
C) Report the DLCOsb as 22.0 ml CO/min/mm Hg.
D) Report the DLCOsb as 21.2 ml CO/min/mm Hg.
Question
A pulmonary function technologist uses an isothermal lung analog to check the accuracy of a constant volume plethysmograph.The lung analog has a volume of 4.00 L.Simulating the VTG maneuver produces the following volumes: <strong>A pulmonary function technologist uses an isothermal lung analog to check the accuracy of a constant volume plethysmograph.The lung analog has a volume of 4.00 L.Simulating the V<sub>TG</sub> maneuver produces the following volumes:   On the basis of these findings,the technologist should:</strong> A) Check the door seal for leaks or cracks. B) Recalibrate the box pressure transducer. C) Zero the mouth pressure transducer and repeat the measurement. D) Begin patient testing with the body box. <div style=padding-top: 35px> On the basis of these findings,the technologist should:

A) Check the door seal for leaks or cracks.
B) Recalibrate the box pressure transducer.
C) Zero the mouth pressure transducer and repeat the measurement.
D) Begin patient testing with the body box.
Question
According to 2005 ATS-ERS recommendations,calibration syringes should be accurate to within (±):

A) 3.5%
B) 0.5%
C) 20 ml
D) 30 ml
Question
A DLCO simulator using two precision gases is set up to check the accuracy of an automated single-breath diffusion testing system.The following results are obtained using a 4-L inspired volume: <strong>A DLCO simulator using two precision gases is set up to check the accuracy of an automated single-breath diffusion testing system.The following results are obtained using a 4-L inspired volume:   Both of the DLCO values are less than the expected values from the simulator.Which of the following best describes these findings?</strong> A) The IVC measurement is erroneous. B) The gas analyzer measured CO incorrectly. C) The gas analyzer measured the trace gas incorrectly. D) BTPS correction was inadvertently left on during the simulation. <div style=padding-top: 35px> Both of the DLCO values are less than the expected values from the simulator.Which of the following best describes these findings?

A) The IVC measurement is erroneous.
B) The gas analyzer measured CO incorrectly.
C) The gas analyzer measured the trace gas incorrectly.
D) BTPS correction was inadvertently left on during the simulation.
Question
The calibration gas for a blood gas analyzer contains 20.1% O2 and 4.99% CO2.If the barometric pressure is 744 mm Hg,what is the partial pressure of gas used to calibrate the PCO2 electrode?

A) 149.5 mm Hg
B) 140.0 mm Hg
C) 37.1 mm Hg
D) 34.8 mm Hg
Question
The acceptable range for accuracy of a spirometer checked with a 3-L syringe is:

A) 0.00 to 14.00 L
B) 0.00 to 8.00 L
C) 2.910 to 3.090 L
D) 2.895 to 3.105 L
Question
A patient whose VC is 3.5 L performs two DLCOsb maneuvers with the following results: <strong>A patient whose VC is 3.5 L performs two DLCOsb maneuvers with the following results:   The pulmonary function technologist should do which of the following?</strong> A) Perform at least one additional maneuver. B) Report the average of Trials 1 and 2. C) Report the DLCOsb as 20.3 ml CO/min/mm Hg. D) Report the DLCOsb as 19.0 ml CO/min/mm Hg. <div style=padding-top: 35px> The pulmonary function technologist should do which of the following?

A) Perform at least one additional maneuver.
B) Report the average of Trials 1 and 2.
C) Report the DLCOsb as 20.3 ml CO/min/mm Hg.
D) Report the DLCOsb as 19.0 ml CO/min/mm Hg.
Question
A portable office spirometer is being checked with a 3-L syringe as a simulated patient.The spirometer uses disposable flow sensors that are calibrated during manufacture.These results are recorded: <strong>A portable office spirometer is being checked with a 3-L syringe as a simulated patient.The spirometer uses disposable flow sensors that are calibrated during manufacture.These results are recorded:   On the basis of these findings,the pulmonary function technologist should:</strong> A) Replace the device's temperature sensor. B) Stop using the spirometer for patient testing until it can be repaired. C) Have the calibration syringe checked for volume accuracy. D) Continue using the spirometer for patient testing. <div style=padding-top: 35px> On the basis of these findings,the pulmonary function technologist should:

A) Replace the device's temperature sensor.
B) Stop using the spirometer for patient testing until it can be repaired.
C) Have the calibration syringe checked for volume accuracy.
D) Continue using the spirometer for patient testing.
Question
What would be your next action after reviewing the data from the following level 1 PaO2 QC Levy-Jennings plot? <strong>What would be your next action after reviewing the data from the following level 1 PaO<sub>2</sub> QC Levy-Jennings plot?  </strong> A) Begin analyzing blood gases because the instrument is in-control. B) Run one additional Level 1 control to determine instrument status. C) Take the instrument out of service because it is out-of-control. D) Run the next control level to determine instrument status. <div style=padding-top: 35px>

A) Begin analyzing blood gases because the instrument is in-control.
B) Run one additional Level 1 control to determine instrument status.
C) Take the instrument out of service because it is out-of-control.
D) Run the next control level to determine instrument status.
Question
All of the following statements are important ATS instrument specifications except:

A) Back pressure and resistance must be less than 1.5 cm H2O/L/sec up to an air flow of 14 L/sec.
B) Spirometer must be capable of measuring volumes of at least 8 L (BTPS)with flows between zero and 14 L/sec.
C) The paper must move at a chart speed of at least 32 mm/sec.
D) Spirometer must be capable of measuring expired volumes for at least 15 seconds.
Question
According to ATS-ERS recommendations,what three flow rates are required to calibrate a flow sensor?

A) 0.5 L/sec,4.0 L/sec,6.0 L/sec
B) 1.5 L/sec,3.0 L/sec,6.0 L/sec
C) 0.5 L/sec,3.0 L/sec,12.0 L/sec
D) 1.0 L/sec,4.0 L/sec,12.0 L/sec
Question
When developing a quality manual,which of the following questions is addressed with a process?

A) How do we do this at organization XYZ?
B) How do we perform spirometry?
C) How are reliable lung volumes obtained?
D) How do we obtain an arterial specimen?
Question
While performing a calibration check of a spirometer using a 3.00-L syringe,2.84 L is measured.What should you do next?

A) Proceed with testing subjects.
B) Add 0.16 L to all subject measurements.
C) Check for leaks and recheck calibration.
D) Turn on the BTPS correction and recheck calibration.
Question
A pulmonary laboratory has purchased a new plethysmograph.During the installation process,the technologist verifies the data integrity.Which QSE is the technologist following?

A) Equipment
B) Process management
C) Information management
D) Purchasing and inventory control
Question
If a volume-measuring spirometer meets the ATS-ERS standards for measuring FVC and is verified using a calibration syringe,how much measurement error is allowed?

A) 0%
B) 3.5%
C) 5.3%
D) 10%
Question
A pulmonary function technologist injects 3-L volumes from a calibration syringe into a spirometer that uses a turbine device to sense flow.The following results are observed: <strong>A pulmonary function technologist injects 3-L volumes from a calibration syringe into a spirometer that uses a turbine device to sense flow.The following results are observed:   The pulmonary function technologist should:</strong> A) Turn off BTPS correction and repeat the trial. B) Turn on BTPS correction and repeat the maneuvers. C) Have the syringe recalibrated. D) Have the spirometer serviced. <div style=padding-top: 35px> The pulmonary function technologist should:

A) Turn off BTPS correction and repeat the trial.
B) Turn on BTPS correction and repeat the maneuvers.
C) Have the syringe recalibrated.
D) Have the spirometer serviced.
Question
The technologist is following a procedure for measuring the correct height of a patient.This step is from which portion of the path of workflow?

A) Pre-test process
B) Test performance
C) Post-testing
D) Clinical evaluation
Question
The BTPS correction adjusts the measured result obtained in the spirometer to the:

A) Ambient temperature for the surrounding air
B) Volume that was originally in the lungs
C) Barometric pressure at the altitude where the testing is conducted
D) Relative humidity in the testing room
Question
What should the technologist do in response to the following Levy-Jennings plot of a PO2 electrode? <strong>What should the technologist do in response to the following Levy-Jennings plot of a PO<sub>2</sub> electrode?  </strong> A) Rerun the control. B) Take the analyzer out of service. C) Run another PaO<sub>2</sub> control level. D) Continue to test patient samples and call technical support. <div style=padding-top: 35px>

A) Rerun the control.
B) Take the analyzer out of service.
C) Run another PaO2 control level.
D) Continue to test patient samples and call technical support.
Question
According to the ATS-ERS recommendations,when using a 3.00-L syringe to check spirometer calibration,which of the following represents an acceptable difference between the expected and actual result?

A) 100 ml
B) 110 ml
C) 120 ml
D) 130 ml
Unlock Deck
Sign up to unlock the cards in this deck!
Unlock Deck
Unlock Deck
1/53
auto play flashcards
Play
simple tutorial
Full screen (f)
exit full mode
Deck 12: Quality Systems in the Pulmonary Function Laboratory
1
A helium analyzer is checked using a 3-L syringe;300 ml of He is added to 2.70 L of air in the syringe.If the dead space of the syringe and tubing is 100 ml,what is the expected concentration of He in the syringe?

A) 11.1%
B) 0.2%
C) 9.7%
D) 8.8%
C
2
An automated blood gas analyzer displays acceptable calibrations with no electrode drift noted.Multiple-level QC shows values for PCO2 and PO2 that are within 2 SD of the mean,but all pH values are more than 3 SD above the mean.Which of the following is most likely the cause?

A) Buffers are contaminated.
B) Calibration gas is contaminated.
C) Analyzer temperature is not 37° C.
D) Gas-liquid correction factor is set too low.
A
3
A 3-L syringe is used to calibrate a flow-sensing spirometer;the spirometer reads a volume of 3.04 L.The software correction factor for this system would be which of the following?

A) 0.95
B) 0.99
C) 1.01
D) 1.04
B
4
A biologic control subject has an established FVC of 5.00 L with an SD of 0.20 L.After spirometer maintenance,the following values are obtained from the biologic control: <strong>A biologic control subject has an established FVC of 5.00 L with an SD of 0.20 L.After spirometer maintenance,the following values are obtained from the biologic control:   On the basis of these findings,the pulmonary function technologist should conclude what?</strong> A) Spirometer performance is within acceptable statistical limits. B) Spirometer performance is questionable. C) The control subject's efforts were greater than normal. D) The control subject's efforts were not temperature-corrected to BTPS. On the basis of these findings,the pulmonary function technologist should conclude what?

A) Spirometer performance is within acceptable statistical limits.
B) Spirometer performance is questionable.
C) The control subject's efforts were greater than normal.
D) The control subject's efforts were not temperature-corrected to BTPS.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
5
Multiple injections from a 3-L syringe into a flow-based spirometer produce an average reading of 2.92 L.This value indicates:

A) The spirometer has a leak.
B) The spirometer meets ATS accuracy requirements.
C) Excessive flows were used for calibration.
D) The spirometer should not be used for pediatric patients.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
6
Tonometry of whole blood may be used to provide QC for which of the following?

A) PO2 and PCO2 electrodes
B) PCO2 and pH electrodes
C) PO2 electrode only
D) pH electrode only
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
7
Quality control of an automated DLCO system is performed using a 3-L syringe.The syringe is filled to 1.5 L with room air and is then connected to the system.A simulated DLCOsb maneuver is performed with 1.5 L of test gas drawn into the syringe.Acceptable performance of the gas analyzers should produce a DLCO of approximately:

A) 0 ml CO/min/mm Hg
B) 25 ml CO/min/mm Hg
C) 50 ml CO/min/mm Hg
D) 100 ml CO/min/mm Hg
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
8
A spirometer recorder used for plotting flow-volume curves has a volume (x axis)sensitivity of 25 mm/L.The flow axis (y axis)sensitivity should be:

A) 12.5 mm/L/sec
B) 25.0 mm/L/sec
C) 50.0 mm/L/sec
D) 50.0 mm/L
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
9
According to ATS recommendations,the acceptable "back pressure" (flow resistance)from a spirometer is less than _____ cm H2O at _____ L/sec.

A) 3.0;7
B) 3.0;14
C) 1.5;7
D) 1.5;14
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
10
According to ATS recommendations,the range and accuracy for a diagnostic spirometer should be which of the following?

A) 0 to 14 L with less than 1.5 cm H2O/L/sec resistance
B) 0.5 to 8 L ± 3% of reading or ± 0.05 L,whichever is greater
C) 0.5 to 6 L ± 5% of reading or ± 0.1 L,whichever is greater
D) 0.1 to 5 L ± 3% of reading or ± 0.05 L,whichever is greater
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
11
A CO2 analyzer is calibrated using a 5% CO2 mixture with a water vapor absorber in the sample line.If the barometric pressure is 747 mm Hg and the analyzer is set up to read partial pressure,the analyzer should be adjusted to read which of the following?

A) 40 mm Hg
B) 37 mm Hg
C) 35 mm Hg
D) 31 mm Hg
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
12
Quality control run on a PO2 electrode shows the following results when plotted on a QC chart: <strong>Quality control run on a PO<sub>2</sub> electrode shows the following results when plotted on a QC chart:   Which of the following best describes the result of run 7?</strong> A) Out-of-control situation B) Shift C) Trend D) Random error Which of the following best describes the result of run 7?

A) Out-of-control situation
B) Shift
C) Trend
D) Random error
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
13
A 50-ml sine-wave pump is used to calibrate a 450-L plethysmograph.While the pump is running,the computer display shows a deflection of 25 cm.What is the box calibration factor for this system?

A) 0.5 L/cm
B) 1.25 L/cm
C) 2 ml/cm
D) 0.5 ml/cm
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
14
Quality control of a body plethysmograph is performed using an isothermal lung analog and a biologic control subject with the following results: <strong>Quality control of a body plethysmograph is performed using an isothermal lung analog and a biologic control subject with the following results:   On the basis of these findings,the pulmonary function technologist should conclude what?</strong> A) There is a leak in the door seal. B) Box pressure calibration was performed incorrectly. C) Mouth pressure calibration was performed incorrectly. D) The box is functioning within acceptable limits. On the basis of these findings,the pulmonary function technologist should conclude what?

A) There is a leak in the door seal.
B) Box pressure calibration was performed incorrectly.
C) Mouth pressure calibration was performed incorrectly.
D) The box is functioning within acceptable limits.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
15
A patient who is short of breath performs eight FVC maneuvers,and these results from the three best efforts are recorded: <strong>A patient who is short of breath performs eight FVC maneuvers,and these results from the three best efforts are recorded:   Which of the following is appropriate to report in the technologist's comments?</strong> A) Spirometry meets all ATS criteria. B) FVC is not repeatable. C) FEV<sub>1</sub> is not repeatable. D) FEF<sub>50% </sub>shows poor patient effort. Which of the following is appropriate to report in the technologist's comments?

A) "Spirometry meets all ATS criteria."
B) "FVC is not repeatable."
C) "FEV1 is not repeatable."
D) "FEF50% shows poor patient effort."
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
16
A CO2 analyzer is calibrated using a 5% CO2 mixture with the sampling pump set at 500 ml/min;a water vapor absorber is then added to the sample line.Which of the following should be done before using this system?

A) Increase the sampling pump flow to at least 750 ml/min.
B) Decrease the sampling pump flow to less than 50 ml/min.
C) Recalibrate the analyzer.
D) Calculate the "corrected" CO2 reading.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
17
A 3-L syringe is used to calibrate a computerized spirometer;at ATPS the spirometer reads a volume of 2.94 L.The software correction factor for this system is:

A) 1.02
B) 1.00
C) 0.98
D) 0.95
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
18
The pulmonary function technologist should observe which of the following precautions in drawing an arterial blood sample?

A) Determine whether gloves are needed by checking the patient's history.
B) Carefully resheath used needles using the original protective covering.
C) Dispose of used needles in puncture-resistant containers marked "biohazard."
D) Use protective barriers if there is danger of blood splashing.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
19
An automated blood gas analyzer performs two-point calibrations that match expected values.Subsequent quality control runs show unacceptable values for PO2 and PCO2,but pH is acceptable.Which of the following is the most likely problem?

A) Contaminated buffers
B) Contaminated calibration gas
C) Analyzer temperature not at 37° C
D) Computer error
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
20
Multiple injections by a 3-L syringe into a bellows-type spirometer produce the following results: <strong>Multiple injections by a 3-L syringe into a bellows-type spirometer produce the following results:   Which of the following best describes these results?</strong> A) The spirometer shows excessive drift. B) The volume is being inappropriately corrected to BTPS. C) The spirometer bellows are sticking. D) Spirometer performance is acceptable. Which of the following best describes these results?

A) The spirometer shows excessive drift.
B) The volume is being inappropriately corrected to BTPS.
C) The spirometer bellows are sticking.
D) Spirometer performance is acceptable.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
21
Which of the following control materials would be best suited for quality control of a PO2 electrode used to measure O2 tensions less than 50 mm Hg?

A) Fluorocarbon-based control
B) Aqueous-based control
C) Tonometered buffer
D) Tonometered whole blood
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
22
A pulmonary function technologist uses a 3-L syringe and a tank of 10% He to prepare serial dilutions of gas to verify the linearity of a helium analyzer used for closed-circuit measurements of FRC.The following data are recorded: <strong>A pulmonary function technologist uses a 3-L syringe and a tank of 10% He to prepare serial dilutions of gas to verify the linearity of a helium analyzer used for closed-circuit measurements of FRC.The following data are recorded:   On the basis of these results,the technologist should conclude that:</strong> A) The He analyzer is performing acceptably. B) The He analyzer is nonlinear. C) The syringe had a significant leak. D) The He tank was mislabeled. On the basis of these results,the technologist should conclude that:

A) The He analyzer is performing acceptably.
B) The He analyzer is nonlinear.
C) The syringe had a significant leak.
D) The He tank was mislabeled.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
23
According to ATS-ERS recommendations (2005),a computer display of flow-volume curves should have a scale factor of:

A) 5 mm/L on the volume axis
B) 5 mm/L/sec on the flow axis
C) 20 mm/sec on the time axis
D) A 1:1 aspect ratio with expiratory flow plotted downward
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
24
A patient performs spirometry,and the following results are obtained: <strong>A patient performs spirometry,and the following results are obtained:   The pulmonary function technologist should conclude which of the following?</strong> A) Only the FVC is repeatable. B) Only the FEV<sub>1</sub> is repeatable. C) Both the FVC and FEV<sub>1</sub> are repeatable. D) The FEV<sub>1</sub>/FVC is 52%. The pulmonary function technologist should conclude which of the following?

A) Only the FVC is repeatable.
B) Only the FEV1 is repeatable.
C) Both the FVC and FEV1 are repeatable.
D) The FEV1/FVC is 52%.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
25
A 3-L syringe is used to calibrate a spirometer;injection of the complete volume produces an output of 3.04 L.The correction factor for this system would be:

A) 0.967
B) 0.987
C) 1.013
D) 1.100
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
26
A subject performs spirometry,and these results are recorded from three acceptable trials: <strong>A subject performs spirometry,and these results are recorded from three acceptable trials:   The flow-volume curve from which trial should be reported?</strong> A) Trial 1 B) Trial 2 C) Trial 3 D) Either Trial 1 or Trial 3 The flow-volume curve from which trial should be reported?

A) Trial 1
B) Trial 2
C) Trial 3
D) Either Trial 1 or Trial 3
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
27
The greatest risk of cross-contamination in the pulmonary function laboratory is associated with:

A) Failure to use bacteria filters for spirometry
B) Inadequate disinfection of reusable nebulizers
C) Moisture inside dry rolling-seal spirometers
D) Pulse oximetry finger sensors
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
28
A biologic control subject performs DLCO measurements on 20 consecutive days,and the following statistics are calculated: <strong>A biologic control subject performs DLCO measurements on 20 consecutive days,and the following statistics are calculated:   The coefficient of variation for the DLCO in this subject:</strong> A) Is approximately 4.5% B) Cannot be calculated until 30 measurements have been made C) Suggests that the patient may have asthma D) Indicates that the DLCO system displays excessive variability The coefficient of variation for the DLCO in this subject:

A) Is approximately 4.5%
B) Cannot be calculated until 30 measurements have been made
C) Suggests that the patient may have asthma
D) Indicates that the DLCO system displays excessive variability
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
29
Quality control runs on a PCO2 electrode have produced a mean of 40 mm Hg ± 2 mm Hg (1 SD).What are the chances of obtaining a value of 45 mm Hg on a control run if the electrode is functioning properly?

A) 95%
B) 50%
C) 5%
D) Less than 1%
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
30
Quality control is performed on a cartridge-based blood gas analyzer,and the following control chart is recorded for the pH: <strong>Quality control is performed on a cartridge-based blood gas analyzer,and the following control chart is recorded for the pH:   What action should the pulmonary function technologist take,based on these results?</strong> A) Begin analyzing blood gases because the instrument is in control. B) Run one additional Level 2 control to determine instrument status. C) Take the instrument out of service because it is out of control. D) Re-membrane the pH electrode and rerun Level 2. What action should the pulmonary function technologist take,based on these results?

A) Begin analyzing blood gases because the instrument is in control.
B) Run one additional Level 2 control to determine instrument status.
C) Take the instrument out of service because it is out of control.
D) Re-membrane the pH electrode and rerun Level 2.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
31
According to the current ATS-ERS guidelines,spirometers should be able to measure volumes of at least:

A) 7 L
B) 8 L
C) 12 L
D) 14 L
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
32
A 3-L syringe is used to check the calibration of a portable spirometer,and the following results are obtained:  <strong>A 3-L syringe is used to check the calibration of a portable spirometer,and the following results are obtained:   Which of the following best describes these findings?</strong> A) The spirometer temperature is less than 24 \degree  C. B) The BTPS correction factor is being applied. C) The syringe has a small but significant leak. D) The spirometer is functioning acceptably.  Which of the following best describes these findings?

A) The spirometer temperature is less than 24 °\degree C.
B) The BTPS correction factor is being applied.
C) The syringe has a small but significant leak.
D) The spirometer is functioning acceptably.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
33
Spirometers used to measure FEV1 should be able to measure flows in the range of 0 to 14 L/sec (ATS-ERS,2005).The maximal resistance allowable at all flows is:

A) Less than 6 cmH2O/L/sec
B) Less than 1.5 cmH2O/L/sec
C) Determined by the type of filter used
D) One half of the flow rate in L/sec
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
34
A patient whose FVC is 4 L performs two DLCOsb maneuvers with the following results: <strong>A patient whose FVC is 4 L performs two DLCOsb maneuvers with the following results:   The pulmonary function technologist should:</strong> A) Reject Trial 1 because of the V<sub>I.</sub> B) Reject Trial 2 because of the breath-hold time. C) Report the DLCOsb as 22.0 ml CO/min/mm Hg. D) Report the DLCOsb as 21.2 ml CO/min/mm Hg. The pulmonary function technologist should:

A) Reject Trial 1 because of the VI.
B) Reject Trial 2 because of the breath-hold time.
C) Report the DLCOsb as 22.0 ml CO/min/mm Hg.
D) Report the DLCOsb as 21.2 ml CO/min/mm Hg.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
35
A pulmonary function technologist uses an isothermal lung analog to check the accuracy of a constant volume plethysmograph.The lung analog has a volume of 4.00 L.Simulating the VTG maneuver produces the following volumes: <strong>A pulmonary function technologist uses an isothermal lung analog to check the accuracy of a constant volume plethysmograph.The lung analog has a volume of 4.00 L.Simulating the V<sub>TG</sub> maneuver produces the following volumes:   On the basis of these findings,the technologist should:</strong> A) Check the door seal for leaks or cracks. B) Recalibrate the box pressure transducer. C) Zero the mouth pressure transducer and repeat the measurement. D) Begin patient testing with the body box. On the basis of these findings,the technologist should:

A) Check the door seal for leaks or cracks.
B) Recalibrate the box pressure transducer.
C) Zero the mouth pressure transducer and repeat the measurement.
D) Begin patient testing with the body box.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
36
According to 2005 ATS-ERS recommendations,calibration syringes should be accurate to within (±):

A) 3.5%
B) 0.5%
C) 20 ml
D) 30 ml
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
37
A DLCO simulator using two precision gases is set up to check the accuracy of an automated single-breath diffusion testing system.The following results are obtained using a 4-L inspired volume: <strong>A DLCO simulator using two precision gases is set up to check the accuracy of an automated single-breath diffusion testing system.The following results are obtained using a 4-L inspired volume:   Both of the DLCO values are less than the expected values from the simulator.Which of the following best describes these findings?</strong> A) The IVC measurement is erroneous. B) The gas analyzer measured CO incorrectly. C) The gas analyzer measured the trace gas incorrectly. D) BTPS correction was inadvertently left on during the simulation. Both of the DLCO values are less than the expected values from the simulator.Which of the following best describes these findings?

A) The IVC measurement is erroneous.
B) The gas analyzer measured CO incorrectly.
C) The gas analyzer measured the trace gas incorrectly.
D) BTPS correction was inadvertently left on during the simulation.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
38
The calibration gas for a blood gas analyzer contains 20.1% O2 and 4.99% CO2.If the barometric pressure is 744 mm Hg,what is the partial pressure of gas used to calibrate the PCO2 electrode?

A) 149.5 mm Hg
B) 140.0 mm Hg
C) 37.1 mm Hg
D) 34.8 mm Hg
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
39
The acceptable range for accuracy of a spirometer checked with a 3-L syringe is:

A) 0.00 to 14.00 L
B) 0.00 to 8.00 L
C) 2.910 to 3.090 L
D) 2.895 to 3.105 L
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
40
A patient whose VC is 3.5 L performs two DLCOsb maneuvers with the following results: <strong>A patient whose VC is 3.5 L performs two DLCOsb maneuvers with the following results:   The pulmonary function technologist should do which of the following?</strong> A) Perform at least one additional maneuver. B) Report the average of Trials 1 and 2. C) Report the DLCOsb as 20.3 ml CO/min/mm Hg. D) Report the DLCOsb as 19.0 ml CO/min/mm Hg. The pulmonary function technologist should do which of the following?

A) Perform at least one additional maneuver.
B) Report the average of Trials 1 and 2.
C) Report the DLCOsb as 20.3 ml CO/min/mm Hg.
D) Report the DLCOsb as 19.0 ml CO/min/mm Hg.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
41
A portable office spirometer is being checked with a 3-L syringe as a simulated patient.The spirometer uses disposable flow sensors that are calibrated during manufacture.These results are recorded: <strong>A portable office spirometer is being checked with a 3-L syringe as a simulated patient.The spirometer uses disposable flow sensors that are calibrated during manufacture.These results are recorded:   On the basis of these findings,the pulmonary function technologist should:</strong> A) Replace the device's temperature sensor. B) Stop using the spirometer for patient testing until it can be repaired. C) Have the calibration syringe checked for volume accuracy. D) Continue using the spirometer for patient testing. On the basis of these findings,the pulmonary function technologist should:

A) Replace the device's temperature sensor.
B) Stop using the spirometer for patient testing until it can be repaired.
C) Have the calibration syringe checked for volume accuracy.
D) Continue using the spirometer for patient testing.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
42
What would be your next action after reviewing the data from the following level 1 PaO2 QC Levy-Jennings plot? <strong>What would be your next action after reviewing the data from the following level 1 PaO<sub>2</sub> QC Levy-Jennings plot?  </strong> A) Begin analyzing blood gases because the instrument is in-control. B) Run one additional Level 1 control to determine instrument status. C) Take the instrument out of service because it is out-of-control. D) Run the next control level to determine instrument status.

A) Begin analyzing blood gases because the instrument is in-control.
B) Run one additional Level 1 control to determine instrument status.
C) Take the instrument out of service because it is out-of-control.
D) Run the next control level to determine instrument status.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
43
All of the following statements are important ATS instrument specifications except:

A) Back pressure and resistance must be less than 1.5 cm H2O/L/sec up to an air flow of 14 L/sec.
B) Spirometer must be capable of measuring volumes of at least 8 L (BTPS)with flows between zero and 14 L/sec.
C) The paper must move at a chart speed of at least 32 mm/sec.
D) Spirometer must be capable of measuring expired volumes for at least 15 seconds.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
44
According to ATS-ERS recommendations,what three flow rates are required to calibrate a flow sensor?

A) 0.5 L/sec,4.0 L/sec,6.0 L/sec
B) 1.5 L/sec,3.0 L/sec,6.0 L/sec
C) 0.5 L/sec,3.0 L/sec,12.0 L/sec
D) 1.0 L/sec,4.0 L/sec,12.0 L/sec
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
45
When developing a quality manual,which of the following questions is addressed with a process?

A) How do we do this at organization XYZ?
B) How do we perform spirometry?
C) How are reliable lung volumes obtained?
D) How do we obtain an arterial specimen?
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
46
While performing a calibration check of a spirometer using a 3.00-L syringe,2.84 L is measured.What should you do next?

A) Proceed with testing subjects.
B) Add 0.16 L to all subject measurements.
C) Check for leaks and recheck calibration.
D) Turn on the BTPS correction and recheck calibration.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
47
A pulmonary laboratory has purchased a new plethysmograph.During the installation process,the technologist verifies the data integrity.Which QSE is the technologist following?

A) Equipment
B) Process management
C) Information management
D) Purchasing and inventory control
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
48
If a volume-measuring spirometer meets the ATS-ERS standards for measuring FVC and is verified using a calibration syringe,how much measurement error is allowed?

A) 0%
B) 3.5%
C) 5.3%
D) 10%
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
49
A pulmonary function technologist injects 3-L volumes from a calibration syringe into a spirometer that uses a turbine device to sense flow.The following results are observed: <strong>A pulmonary function technologist injects 3-L volumes from a calibration syringe into a spirometer that uses a turbine device to sense flow.The following results are observed:   The pulmonary function technologist should:</strong> A) Turn off BTPS correction and repeat the trial. B) Turn on BTPS correction and repeat the maneuvers. C) Have the syringe recalibrated. D) Have the spirometer serviced. The pulmonary function technologist should:

A) Turn off BTPS correction and repeat the trial.
B) Turn on BTPS correction and repeat the maneuvers.
C) Have the syringe recalibrated.
D) Have the spirometer serviced.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
50
The technologist is following a procedure for measuring the correct height of a patient.This step is from which portion of the path of workflow?

A) Pre-test process
B) Test performance
C) Post-testing
D) Clinical evaluation
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
51
The BTPS correction adjusts the measured result obtained in the spirometer to the:

A) Ambient temperature for the surrounding air
B) Volume that was originally in the lungs
C) Barometric pressure at the altitude where the testing is conducted
D) Relative humidity in the testing room
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
52
What should the technologist do in response to the following Levy-Jennings plot of a PO2 electrode? <strong>What should the technologist do in response to the following Levy-Jennings plot of a PO<sub>2</sub> electrode?  </strong> A) Rerun the control. B) Take the analyzer out of service. C) Run another PaO<sub>2</sub> control level. D) Continue to test patient samples and call technical support.

A) Rerun the control.
B) Take the analyzer out of service.
C) Run another PaO2 control level.
D) Continue to test patient samples and call technical support.
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
53
According to the ATS-ERS recommendations,when using a 3.00-L syringe to check spirometer calibration,which of the following represents an acceptable difference between the expected and actual result?

A) 100 ml
B) 110 ml
C) 120 ml
D) 130 ml
Unlock Deck
Unlock for access to all 53 flashcards in this deck.
Unlock Deck
k this deck
locked card icon
Unlock Deck
Unlock for access to all 53 flashcards in this deck.