Exam 12: Quality Systems in the Pulmonary Function Laboratory
Exam 1: Indications for Pulmonary Function Testing25 Questions
Exam 2: Spirometry32 Questions
Exam 3: Diffusing Capacity Tests33 Questions
Exam 4: Lung Volumes,airway Resistance,and Gas Distribution Tests22 Questions
Exam 5: Ventilation and Ventilatory Control Tests24 Questions
Exam 6: Blood Gases and Related Tests28 Questions
Exam 7: Cardiopulmonary Exercise Testing26 Questions
Exam 8: Pediatric Pulmonary Function Tests27 Questions
Exam 9: Bronchoprovocation Testing18 Questions
Exam 10: Specialized Test Regimens21 Questions
Exam 11: Pulmonary Function Test Equipment38 Questions
Exam 12: Quality Systems in the Pulmonary Function Laboratory53 Questions
Exam 13: Reference Values and Interpretation Strategies20 Questions
Select questions type
Tonometry of whole blood may be used to provide QC for which of the following?
Free
(Multiple Choice)
4.8/5
(44)
Correct Answer:
A
When developing a quality manual,which of the following questions is addressed with a process?
Free
(Multiple Choice)
4.8/5
(29)
Correct Answer:
C
A 3-L syringe is used to calibrate a spirometer;injection of the complete volume produces an output of 3.04 L.The correction factor for this system would be:
Free
(Multiple Choice)
4.9/5
(33)
Correct Answer:
B
A DLCO simulator using two precision gases is set up to check the accuracy of an automated single-breath diffusion testing system.The following results are obtained using a 4-L inspired volume:
Both of the DLCO values are less than the expected values from the simulator.Which of the following best describes these findings?

(Multiple Choice)
4.9/5
(38)
If a volume-measuring spirometer meets the ATS-ERS standards for measuring FVC and is verified using a calibration syringe,how much measurement error is allowed?
(Multiple Choice)
4.7/5
(37)
A spirometer recorder used for plotting flow-volume curves has a volume (x axis)sensitivity of 25 mm/L.The flow axis (y axis)sensitivity should be:
(Multiple Choice)
4.8/5
(43)
Quality control is performed on a cartridge-based blood gas analyzer,and the following control chart is recorded for the pH:
What action should the pulmonary function technologist take,based on these results?

(Multiple Choice)
4.8/5
(32)
An automated blood gas analyzer displays acceptable calibrations with no electrode drift noted.Multiple-level QC shows values for PCO2 and PO2 that are within 2 SD of the mean,but all pH values are more than 3 SD above the mean.Which of the following is most likely the cause?
(Multiple Choice)
4.8/5
(33)
A pulmonary function technologist uses a 3-L syringe and a tank of 10% He to prepare serial dilutions of gas to verify the linearity of a helium analyzer used for closed-circuit measurements of FRC.The following data are recorded:
On the basis of these results,the technologist should conclude that:

(Multiple Choice)
4.8/5
(27)
A biologic control subject performs DLCO measurements on 20 consecutive days,and the following statistics are calculated:
The coefficient of variation for the DLCO in this subject:

(Multiple Choice)
4.9/5
(33)
A CO2 analyzer is calibrated using a 5% CO2 mixture with a water vapor absorber in the sample line.If the barometric pressure is 747 mm Hg and the analyzer is set up to read partial pressure,the analyzer should be adjusted to read which of the following?
(Multiple Choice)
4.8/5
(36)
According to ATS-ERS recommendations,what three flow rates are required to calibrate a flow sensor?
(Multiple Choice)
4.8/5
(30)
A subject performs spirometry,and these results are recorded from three acceptable trials:
The flow-volume curve from which trial should be reported?

(Multiple Choice)
4.8/5
(31)
According to 2005 ATS-ERS recommendations,calibration syringes should be accurate to within (±):
(Multiple Choice)
4.8/5
(38)
Multiple injections by a 3-L syringe into a bellows-type spirometer produce the following results:
Which of the following best describes these results?

(Multiple Choice)
4.8/5
(44)
The BTPS correction adjusts the measured result obtained in the spirometer to the:
(Multiple Choice)
4.9/5
(32)
A patient whose FVC is 4 L performs two DLCOsb maneuvers with the following results:
The pulmonary function technologist should:

(Multiple Choice)
4.8/5
(37)
According to ATS recommendations,the range and accuracy for a diagnostic spirometer should be which of the following?
(Multiple Choice)
4.9/5
(20)
A portable office spirometer is being checked with a 3-L syringe as a simulated patient.The spirometer uses disposable flow sensors that are calibrated during manufacture.These results are recorded:
On the basis of these findings,the pulmonary function technologist should:

(Multiple Choice)
4.9/5
(39)
What should the technologist do in response to the following Levy-Jennings plot of a PO2 electrode? 

(Multiple Choice)
4.8/5
(45)