Deck 12: Quality Control in Computed Tomography
1
The minimum distance between two objects that allows them to be seen as separate and distinct would define what CT imaging parameter?
A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
A)Low-contrast resolution
B)Uniformity
C)High-contrast resolution
D)Image noise
C
2
The laser light field and radiation field should coincide to within ____ mm.
A)0.5
B)1.0
C)2.0
D)5.0
A)0.5
B)1.0
C)2.0
D)5.0
C
3
For ACR CT accreditation, the Water CT Number and Standard Deviation (Noise) should be evaluated by the CT technologist
A)quarterly.
B)monthly.
C)weekly.
D)daily.
A)quarterly.
B)monthly.
C)weekly.
D)daily.
D
4
A visual inspection of a CT unit should be evaluated at least _____ for ACR accreditation programs.
A)daily.
B)weekly.
C)monthly.
D)semi-annually.
A)daily.
B)weekly.
C)monthly.
D)semi-annually.
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
5
The CTDIᵥₒₗ pass/fail criterion for an adult abdomen is ____ mGy.
A)20
B)40
C)30
D)80
A)20
B)40
C)30
D)80
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
6
The uniformity of the CT number of water should not vary by more than ____ HU from the center of the phantom to the periphery.
A)2
B)3
C)5
D)10
A)2
B)3
C)5
D)10
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
7
The CT number of water should not deviate by more than ____ CT numbers from 0.
A)2
B)3
C)5
D)10
A)2
B)3
C)5
D)10
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
8
Section thickness should not vary by more than ____ mm for sections greater than 5 mm.
A)±0.5
B)±1.0
C)±2.0
D)±5.0
A)±0.5
B)±1.0
C)±2.0
D)±5.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
9
For a slice thickness of 5 mm or less, the slice thickness should not vary by more than ____ mm.
A)0.5
B)1.0
C)2.0
D)3.0
A)0.5
B)1.0
C)2.0
D)3.0
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck
10
Patient dose values for CT scans should remain within ____ of the manufacturer's specifications.
A)±2%
B)±5%
C)±10%
D)±20%
A)±2%
B)±5%
C)±10%
D)±20%
Unlock Deck
Unlock for access to all 10 flashcards in this deck.
Unlock Deck
k this deck