Exam 18: Analysis and Monitoring of Gas Exchange

arrow
  • Select Tags
search iconSearch Question
flashcardsStudy Flashcards
  • Select Tags

Before attaching a transcutaneous blood gas monitor sensor to a patient, what should you do? I.Provide a specified warm-up time and set the probe temperature. II.Check the membrane and prepare a sensor with an adhesive ring and gel. III.Prepare the monitoring site (remove excess hair and clean the skin).

(Multiple Choice)
4.9/5
(36)

Tissue injuries to be on guard for at the site of transcutaneous blood gas electrode placement include all of the following except which one?

(Multiple Choice)
4.8/5
(46)

All of the following can result in falsely high HbO2 readings with a laboratory hemoximeter except:

(Multiple Choice)
4.9/5
(43)

How often should blood gas calibration verification by control media take place?

(Multiple Choice)
4.7/5
(41)

Which blood gas analyzer electrode uses a separate reference electrode?

(Multiple Choice)
4.9/5
(37)

Before a sample of capillary blood is taken, what should you do to the site? I.warmed to 42° C for 10 minutes II.squeezed lightly until blanched III.cleaned with an antiseptic solution

(Multiple Choice)
4.8/5
(33)

All of the following are common sites for transcutaneous blood gas electrode placement except the:

(Multiple Choice)
4.8/5
(33)

The advantage that point-of-care testing has over traditional laboratory testing is that point-of-care testing:

(Multiple Choice)
4.8/5
(32)

While checking a galvanic fuel cell O2 analyzer, you determine that the device fails to read 100% when exposed to pure O2. Which of the following actions would be the proper first step?

(Multiple Choice)
4.8/5
(31)

Which of the following describes the correct procedure for an Allen test?

(Multiple Choice)
4.7/5
(36)

At the very beginning of exhalation, the PETCO2 normally should be at what level?

(Multiple Choice)
4.8/5
(42)

To avoid transmission of blood-borne diseases when handling a used needle, what should you do? I.Never recap the needle without a safety device (or scoop method). II.Never handle the needle with both hands, or point it toward the body. III.Never bend, break, or remove the needle from the syringe by hand. IV.Always dispose of the syringe or needle in a proper sharps container.

(Multiple Choice)
4.7/5
(26)

Before taking a spot reading with a pulse oximeter, what should you do? I.Allow sufficient response time. II.Set the low alarm to 88% to 90%. III.Confirm adequacy of pulse signal.

(Multiple Choice)
4.9/5
(39)

A blood gas analyzer measures all of the following except:

(Multiple Choice)
4.9/5
(35)

Continuous SpO2 monitoring (versus a spot check) is indicated in all of the following situations except:

(Multiple Choice)
4.8/5
(37)

All of the following are TRUE about capillary blood gas sampling except:

(Multiple Choice)
4.9/5
(31)

To ensure that the output of a blood gas analyzer is both accurate and linear across the range of measured values, a calibration procedure must measure what?

(Multiple Choice)
4.8/5
(42)

What is the normal range for end-tidal CO2 as measured by capnography?

(Multiple Choice)
4.8/5
(36)

Which of the following guidelines should you adhere to when performing pulse oximetry? I.Never mix different sensors among different devices. II.Make sure that the sensors are the correct size and are properly applied. III.Avoid using pulse oximetry to monitor hyperoxia in neonates. IV.Whenever possible, validate the initial SpO2 against the actual SaO2.

(Multiple Choice)
4.9/5
(36)

Which of the following would you expect to occur if too much heparin was used in gathering an ABG sample from a patient breathing room air?

(Multiple Choice)
4.8/5
(30)
Showing 61 - 80 of 125
close modal

Filters

  • Essay(0)
  • Multiple Choice(0)
  • Short Answer(0)
  • True False(0)
  • Matching(0)